Skip to main content

Advertisement

Log in

Clinical and radiographic evaluation of total hip arthroplasties using porous tantalum modular acetabular components: 5-year follow-up of clinical trial

  • Original Article
  • Published:
Modern Rheumatology

Abstract

Objectives

Porous tantalum is a biomaterial newly applied for artificial joints. We present here 5-years follow-up report of a multicenter clinical trial of total hip arthroplasties (THA) with porous tantalum modular acetabular component (modular PTC).

Methods

Study participants received 82 hips in 79 cases, with 61.2 months follow-up on average. Age at operation was 60.9 years. Clinical results were evaluated using Merle d’Aubigne Postel score. Presence of implant loosening, periacetabular radiolucency, osteolysis, and gap filling were examined for radiographic results.

Results

Merle d’Aubigne Postel score improved from 10.0 to 16.4 points. All PTC were radiographically stable, with no evidence of progressive radiolucencies. Average polyethylene wear rate was 0.004 mm/year, with no periacetabular osteolysis. Fifteen hips (18.3%) showed a gap >1 mm; however, all showed bone filling within 12 months. PTC with oversized reaming was significantly less likely to have a gap. No implant failure was noted related to modularity. Resulting survival rate of modular PTC was 100% at 5 years.

Conclusions

Modular PTC showed excellent results at 5-years of follow-up. Some hips showed periacetabular gaps, which were filled with bone within 1 year. Further follow-up was needed to determine long-term efficacy.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5

Similar content being viewed by others

References

  1. Latimer HA, Lachiewicz PF. Porous-coated acetabular components with screw fixation. Five to ten-year results. J Bone Joint Surg Am. 1996;78:975–81.

    PubMed  CAS  Google Scholar 

  2. Archibeck MJ, Berger RA, Jacobs JJ, Quigley LR, Gitelis S, Rosenberg AG, et al. Second-generation cementless total hip arthroplasty: eight to eleven-year results. J Bone Joint Surg Am. 2001;83:1666–73.

    PubMed  Google Scholar 

  3. Fisher DA, Mallory TH, Kraus TJ, Mitchell MB. Evaluation of the design and clinical performance of cementless acetabular components. Semin Arthroplasty. 1990;1:35–44.

    PubMed  CAS  Google Scholar 

  4. Schmalzried TP, Wessinger SJ, Hill GE, Harris WH. The Harris-Galante porous acetabular component press-fit without screw fixation: five-year radiographic analysis of primary cases. J Arthroplasty. 1994;9:235–42.

    Article  PubMed  CAS  Google Scholar 

  5. Bobyn JD, Engh CA, Glassman AH. Radiography and histology of a threaded acetabular implant. One case studied at two years. J Bone Joint Surg Br. 1988;70(2):302–4.

    PubMed  CAS  Google Scholar 

  6. Bono JV, Sanford L, Toussaint JT. Severe polyethylene wear in total hip arthroplasty: observations from retrieved AML PLUS hip implants with an ACS polyethylene liner. J Arthroplasty. 1994;9:119–25.

    Article  PubMed  CAS  Google Scholar 

  7. Schmalzried TP, Jasty M, Harris WH. Periprosthetic bone loss in total hip arthroplasty: polyethylene wear debris and the concept of the effective joint space. J Bone Joint Surg Am. 1992;74:849–63.

    PubMed  CAS  Google Scholar 

  8. Kienapfel H, Sprey C, Wilke A, Griss P. Implant fixation by bone ingrowth. J Arthroplasty. 1999;14(3):355–68.

    Article  PubMed  CAS  Google Scholar 

  9. Nakashima Y, Hayashi K, Inadome T, Uenoyama K, Hara T, Kanemaru T, et al. Hydroxyapatite-coating on titanium arc sprayed titanium implants. J Biomed Mater Res. 1997;35(3):287–98.

    Article  PubMed  CAS  Google Scholar 

  10. D’Antonio JA, Capello WN, Manley MT, Feinberg J. Hydroxyapatite coated implants: total hip arthroplasty in the young patient and patients with avascular necrosis. Clin Orthop Relat Res. 1997;344:124–38.

    Article  PubMed  Google Scholar 

  11. Jafari SM, Bender B, Coyle C, Parvizi J, Sharkey PF, Hozack WJ. Do tantalum and titanium cups show similar results in revision hip arthroplasty? Clin Orthop Relat Res. 2010;468(2):459–65.

    Article  PubMed  Google Scholar 

  12. Bobyn JD, Stackpool GJ, Hacking SA, Tanzer M, Krygier JJ. Characteristics of bone ingrowth and interface mechanics of a new porous tantalum biomaterial. J Bone Joint Surg Br. 1999;81:907–14.

    Article  PubMed  CAS  Google Scholar 

  13. Bobyn JD, Toh KK, Hacking SA, Tanzer M, Krygier JJ. Tissue response to porous tantalum acetabular cups: a canine model. J Arthroplasty. 1999;14:347–54.

    Article  PubMed  CAS  Google Scholar 

  14. Deglurkar M, Davy DT, Stewart M, Goldberg VM, Welter JF. Evaluation of machining methods for trabecular metal implants in a rabbit intramedullary osseointegration model. J Biomed Mater Res B Appl Biomater. 2007;80:528–40.

    PubMed  Google Scholar 

  15. Rahbek O, Kold S, Zippor B, Overgaard S, Soballe K. Particle migration and gap healing around trabecular metal implants. Int Orthop. 2005;29:368–74.

    Article  PubMed  CAS  Google Scholar 

  16. Levine B, Della Valle CJ, Jacobs JJ. Applications of porous tantalum in total hip arthroplasty. J Am Acad Orthop Surg. 2006;14(12):646–55.

    PubMed  Google Scholar 

  17. Baad-Hansen T, Kold S, Nielsen PT, Laursen MB, Christensen PH, Soballe K. Comparison of trabecular metal cups and titanium fiber-mesh cups in primary hip arthroplasty: a randomized RSA and bone mineral densitometry study of 50 hips. Acta Orthop. 2011;82(2):155–60.

    Article  PubMed  Google Scholar 

  18. Periasamy K, Watson WS, Mohammed A, Murray H, Walker B, Patil S, et al. A randomised study of peri-prosthetic bone density after cemented versus trabecular fixation of a polyethylene acetabular component. J Bone Joint Surg Br. 2011;93(8):1033–44.

    Article  PubMed  CAS  Google Scholar 

  19. Shiraza-Adl A, Dammak M, Paiement D. Experimental determination of friction characteristics at the trabecular bone/porous-coated metal interface in cementless implants. J Biomed Mater Res. 2003;27:167–75.

    Article  Google Scholar 

  20. Macheras GA, Papagelopoulos PJ, Kateros K, Kostakos AT, Baltas D, Karachalios TS. Radiological evaluation of the metal–bone interface of a porous tantalum monoblock acetabular component. J Bone Joint Surg Br. 2006;88:304–9.

    Article  PubMed  CAS  Google Scholar 

  21. Malizos KN, Bargiotas K, Papatheodorou L, Hantes M, Karachalios T. Survivorship of monoblock trabecular metal cups in primary THA: midterm results. Clin Orthop Relat Res. 2008;466(1):159–66.

    Article  PubMed  Google Scholar 

  22. Simon JP, Bellemans J. Clinical and radiological evaluation of modular trabecular metal acetabular cups. Short-term results in 64 hips. Acta Orthop Belg. 2009;75(5):623–30.

    PubMed  Google Scholar 

  23. Crowe JF, Mani VJ, Ranawat CS. Total hip replacement in congenital dislocation and dysplasia of the hip. J Bone Joint Surg Am. 1979;61(1):15–23.

    PubMed  CAS  Google Scholar 

  24. Kerboul M, Thomine J, Postel M, Merle d’Aubigné R. The conservative surgical treatment of idiopathic aseptic necrosis of the femoral head. J Bone Joint Surg Br. 1974;56(2):291–6.

    PubMed  CAS  Google Scholar 

  25. Schmalzried TP, Harris WH. The Harris-Galante porous-coated acetabular component with screw fixation. Radiographic analysis of eighty-three primary hip replacements at a minimum of five years. J Bone Joint Surg Am. 1992;74(8):1130–9.

    PubMed  CAS  Google Scholar 

  26. Ebramzadeh E, Sangiorgio SN, Lattuada F, Kang JS, Chiesa R, McKellop HA, et al. Accuracy of measurement of polyethylene wear with use of radiographs of total hip replacements. J Bone Joint Surg Am. 2003;85(12):2378–84.

    PubMed  Google Scholar 

  27. Gruen TA, Poggie RA, Lewallen DG, Hanssen AD, Lewis RJ, O’Keefe TJ, et al. Radiographic evaluation of a monoblock acetabular component: a multicenter study with 2- to 5-year results. J Arthroplasty. 2005;20:369–78.

    Article  PubMed  Google Scholar 

  28. DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976;121:20–32.

    PubMed  Google Scholar 

  29. Kurtz SM, Gawel HA, Patel JD. History and systematic review of wear and osteolysis outcomes for first-generation highly crosslinked polyethylene. Clin Orthop Relat Res. 2011;469(8):2262–77.

    Article  PubMed  Google Scholar 

  30. Komarasamy B, Vadivelu R, Bruce A, Kershaw C, Davison J. Clinical and radiological outcome following total hip arthroplasty with an uncemented trabecular metal monoblock acetabular cup. Acta Orthop Belg. 2006;72:320–5.

    PubMed  Google Scholar 

  31. Xenakis TA, Macheras GA, Stafilas KS, Kostakos AT, Bargiotas K, Malizos KN. Multicentre use of a porous tantalum monoblock acetabular component. Int Orthop. 2009;33(4):911–6.

    Article  PubMed  Google Scholar 

Download references

Conflict of interest

Y. N. received lecture fees from Zimmer. All other authors have declared no conflicts of interest.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Yasuharu Nakashima.

About this article

Cite this article

Nakashima, Y., Mashima, N., Imai, H. et al. Clinical and radiographic evaluation of total hip arthroplasties using porous tantalum modular acetabular components: 5-year follow-up of clinical trial. Mod Rheumatol 23, 112–118 (2013). https://doi.org/10.1007/s10165-012-0618-9

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10165-012-0618-9

Keywords

Navigation