Abstract
The paper describes technological advances in high-field (7 T) transceiver-phased arrays developed for magnetic resonance imaging of the human brain. The first part of this work describes an 8-element inductively decoupled split elliptical transceiver-phased array with selectable geometry, which provides an easy and efficient way of compensating for changes in mutual inductive coupling associated with difference in loading due to variability in head shape and size. The second part of the work describes a double-row 16-element (2 × 8) transceiver array to extend the homogeneous transmit B 1 profile in the longitudinal direction. Multiplexing eight transmit channels between the two rows of the array provides homogeneous excitation over the entire volume. The final section describes design and construction of a double-tuned 31P/1H 16-element (8 at each frequency) array. The array improves transmission efficiency and B 1 homogeneity at 1H frequency in comparison with 31P/1H quadrature transverse electromagnetic volume coil. For 31P studies, the array also improves transmission efficiency (38%), signal-to-noise ratio (SNR) for central brain locations (20%) and provides substantially greater SNR (up to 400%) for peripheral locations.
Similar content being viewed by others
References
J.T. Vaughan, M. Garwood, C.M. Collins, W. Liu, L. DelaBarre, G. Adriany, G.P. Andersen, H. Merkle, R. Goebel, M.B. Smith, K. Ugurbil, Magn. Reson. Med. 46, 24 (2001)
P.-F. Van de Moortele, C. Akgun, G. Adriany, S. Moeller, J. Ritter, C.M. Collins, M.B. Smith, J.T. Vaughan, K. Ugurbil, Magn. Reson. Med. 54, 1503 (2005)
C.M. Collins, W. Liu, W. Schreiber, Q.Y. Yang, M.B. Smith, J. Magn. Reson. Imag. 21, 192 (2005)
N.I. Avdievich, H.P. Hetherington, A.M. Kuznetsov, J.W. Pan, J. Magn. Reson. Imag. 29, 461 (2009)
A. Nabetani, G. McKinnon, T. Nakada, in Proceedings of the 14th annual meeting of ISMRM, Seattle, 2006, p. 2608
G. Adriany, P.-F. Van de Moortele, F. Wiesinger, S. Moeller, J.P. Strupp, P. Andersen, C. Snyder, X. Zhang, W. Chen, K.P. Pruessmann, P. Boesiger, J.T. Vaughan, K. Ugurbil, Magn. Reson. Med. 53, 434 (2005)
R.G. Pinkerton, E.A. Barberi, R.S. Menon, Magn. Reson. Med. 54, 499 (2005)
G. Adriany, P.F. Van de Moortele, J. Ritter, S. Moeller, E.J. Auerbach, C. Akgun, C.J. Snyder, T. Vaughan, K. Ugurbil, Magn. Reson. Med. 59, 590 (2008)
N.I. Avdievich, J.W. Pan, J.M. Baehring, D.D. Spencer, H.P. Hetherington, Magn. Reson. Med. 62, 17 (2009)
W. Mao, M.B. Smith, C.M. Collins, Magn. Reson. Med. 56, 4–918 (2006)
T.S. Ibrahim, L. Tang, J. Magn. Reson. Imag. 25, 6–1235 (2007)
J. Tropp, T. Schirmer, J. Magn. Reson. 151, 146 (2001)
N.I. Avdievich, J.W. Pan, A.M. Kuznetsov, H.P. Hetherington, in Proceedings of the 17th annual meeting of ISMRM, Honolulu, 2009, p. 3002
H.P. Hetherington, N.I. Avdievich, J.W. Pan, in Proceedings of the 18 th annual meeting of ISMRM, Stockholm, 2010, p. 3822
N.I. Avdievich, J.W. Pan, H.P. Hetherington, in Proceedings of the 19th annual meeting of ISMRM, Montreal, 2011, p. 328
N.I. Avdievich, J.W. Pan, H.P. Hetherington, in Proceedings of the 18th annual meeting of ISMRM, Stockholm, 2010, p. 1501
N.I. Avdievich, H.P. Hetherington, J. Magn. Reson. 186, 341 (2007)
Q.X. Yang, J. Wang, C.M. Collins, M.B. Smith, X. Zhang, K. Ugurbil, W. Chen, Magn. Reson. Med. 52, 1016 (2004)
J.T. Vaughan, H.P. Hetherington, J.O. Otu, J.W. Pan, G.M. Pohost, Magn. Reson. Med. 32, 206 (1994)
M.D. Harpen, Magn. Reson. Med. 29, 5–713 (1993)
M. Sucher, J. Fox, Handbook of microwave measurements (Interscience Publishers, Polytechnic Press of the Polytechnic Institute of Brooklyn/New York, 1963)
J.P. Strupp, E.J. Auerbach, A. Gozubuyuk, G. Adriany, K. Ugurbil, P.-F. Van De Moortele, in Proceedings of the 16th annual meeting of ISMRM, Toronto, 2008, p. 1135
J.W. Pan, D.B. Twieg, H.P. Hetherington, Magn. Reson. Med. 40, 363 (1998)
H.P. Hetherington, N.I. Avdievich, J.W. Pan, in Proceedings of the 19th annual meeting of ISMRM, Montreal, 2011, p. 163
H.P. Hetherington, N.I. Avdievich, A.M. Kuznetsov, J.W. Pan, Magn. Reson. Med. 63, 9 (2010)
A.R. Rath, J. Magn. Reson. 86, 488 (1990)
J. Murphy-Boesch, R. Srinivasan, L. Carvajal, T.R. Brown, J. Magn. Reson. B 103, 103 (1994)
G.X. Shen, J.F. Wu, F.E. Boada, K.R. Thulborn, Magn. Reson. Med. 41, 268 (1999)
M.D. Schnall, V.H. Subramanian, J.S. Leigh, B. Chance, J. Magn. Reson. 65, 122 (1985)
J.R. Fitzsimmons, B.L. Beck, H.R. Brooker, Magn. Reson. Med. 30, 107 (1993)
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Avdievich, N.I. Transceiver-Phased Arrays for Human Brain Studies at 7 T. Appl Magn Reson 41, 483–506 (2011). https://doi.org/10.1007/s00723-011-0280-y
Received:
Revised:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00723-011-0280-y