Abstract
Objective: The programmable Shunt Assistant (capped!) (ProSA) shunt system has recently been introduced into clinical use. The system can be in vivo adjusted magnetically, and this adjustability is supposed to affect CSF drainage only in the vertical body position.
Materials and methods: We tested a combination of ProSA with fixed differential pressure valve designed to drain cerebrospinal fluid (CSF) from the brain ventricles into the peritoneal cavity.
Results: The ProSA showed good mechanical durability and stability of hydrodynamic performance over a 60-day period. The flow-pressure performance curves and operating pressures were stable, fell within the limits specified by the manufacturer, and changed according to the programmed performance levels. The ProSA system has higher than usual hydrodynamic resistance (around 8.8 mmHg/(mL/min)) in the vertical position. Operating pressure in the vertical position reacted repeatedly to changes of settings within the limits 0–40 cmH2O and was reduced gradually when the axis of the valve declined from vertical to horizontal. External programming proved to be reliable. Strong magnetic fields (3-T MRI) were not able to change the programming of ProSA.
Conclusion: ProSA works in the horizontal position as differential fixed pressure low hydrodynamic resistance and in the vertical position as an adjustable, normal hydrodynamic resistance valve. It is able to compensate for posture-related overdrainage
This is a preview of subscription content, log in via an institution.
Buying options
Tax calculation will be finalised at checkout
Purchases are for personal use only
Learn about institutional subscriptionsReferences
Aschoff A, Kremer P, Benesch C, Fruh K, Klank A (1995) Kunze S Overdrainage and shunt technology. A critical comparison of programmable, hydrostatic and variable-resistance valves and flow-reducing devices. Childs Nerv Syst 11(4):193–202
Czosnyka Z, Czosnyka M, Richards HK, Pickard JD (2002) Laboratory testing of hydrocephalus shunts – conclusion of the U.K. Shunt evaluation programme. Acta Neurochir (Wien) 144(6):525–38
Momjian S, Owler BK, Czosnyka Z, Czosnyka M, Pena A, Pickard JD (2004) Pattern of white matter regional cerebral blood flow and autoregulation in normal pressure hydrocephalus. Brain 127(Pt 5):965–72
Petrella G, Czosnyka M, Smielewski P, Allin D, Guazzo EP, Pickard JD, Czosnyka ZH (2009) In vivo assessment of hydrocephalus shunt. Acta Neurol Scand 120(5):317–23
Conflicts of interest statement
This evaluation study was commissioned by B. Brown Ltd UK, in agreement with the University of Cambridge, which partly contributes to Z.C.’s salary. The study was conducted independently, and the manufacturer’s comments were provided unabridged.
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2012 Springer-Verlag/Wien
About this paper
Cite this paper
Czosnyka, M., Czosnyka, Z., Pickard, J.D. (2012). Programmable Shunt Assistant Tested in Cambridge Shunt Evaluation Laboratory. In: Aygok, G., Rekate, H. (eds) Hydrocephalus. Acta Neurochirurgica Supplementum, vol 113. Springer, Vienna. https://doi.org/10.1007/978-3-7091-0923-6_15
Download citation
DOI: https://doi.org/10.1007/978-3-7091-0923-6_15
Published:
Publisher Name: Springer, Vienna
Print ISBN: 978-3-7091-0922-9
Online ISBN: 978-3-7091-0923-6
eBook Packages: MedicineMedicine (R0)