Programmable Shunt Assistant Tested in Cambridge Shunt Evaluation Laboratory

  • Conference paper
  • First Online:

Part of the book series: Acta Neurochirurgica Supplementum ((NEUROCHIRURGICA,volume 113))

Abstract

Objective: The programmable Shunt Assistant (capped!) (ProSA) shunt system has recently been introduced into clinical use. The system can be in vivo adjusted magnetically, and this adjustability is supposed to affect CSF drainage only in the vertical body position.

Materials and methods: We tested a combination of ProSA with fixed differential pressure valve designed to drain cerebrospinal fluid (CSF) from the brain ventricles into the peritoneal cavity.

Results: The ProSA showed good mechanical durability and stability of hydrodynamic performance over a 60-day period. The flow-pressure performance curves and operating pressures were stable, fell within the limits specified by the manufacturer, and changed according to the programmed performance levels. The ProSA system has higher than usual hydrodynamic resistance (around 8.8 mmHg/(mL/min)) in the vertical position. Operating pressure in the vertical position reacted repeatedly to changes of settings within the limits 0–40 cmH2O and was reduced gradually when the axis of the valve declined from vertical to horizontal. External programming proved to be reliable. Strong magnetic fields (3-T MRI) were not able to change the programming of ProSA.

Conclusion: ProSA works in the horizontal position as differential fixed pressure low hydrodynamic resistance and in the vertical position as an adjustable, normal hydrodynamic resistance valve. It is able to compensate for posture-related overdrainage

This is a preview of subscription content, log in via an institution.

Buying options

Chapter
USD   29.95
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
eBook
USD   189.00
Price excludes VAT (USA)
  • Available as EPUB and PDF
  • Read on any device
  • Instant download
  • Own it forever
Softcover Book
USD   159.99
Price excludes VAT (USA)
  • Compact, lightweight edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info
Hardcover Book
USD   219.99
Price excludes VAT (USA)
  • Durable hardcover edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info

Tax calculation will be finalised at checkout

Purchases are for personal use only

Learn about institutional subscriptions

References

  1. Aschoff A, Kremer P, Benesch C, Fruh K, Klank A (1995) Kunze S Overdrainage and shunt technology. A critical comparison of programmable, hydrostatic and variable-resistance valves and flow-reducing devices. Childs Nerv Syst 11(4):193–202

    Article  PubMed  CAS  Google Scholar 

  2. Czosnyka Z, Czosnyka M, Richards HK, Pickard JD (2002) Laboratory testing of hydrocephalus shunts – conclusion of the U.K. Shunt evaluation programme. Acta Neurochir (Wien) 144(6):525–38

    Article  CAS  Google Scholar 

  3. Momjian S, Owler BK, Czosnyka Z, Czosnyka M, Pena A, Pickard JD (2004) Pattern of white matter regional cerebral blood flow and autoregulation in normal pressure hydrocephalus. Brain 127(Pt 5):965–72

    Article  PubMed  Google Scholar 

  4. Petrella G, Czosnyka M, Smielewski P, Allin D, Guazzo EP, Pickard JD, Czosnyka ZH (2009) In vivo assessment of hydrocephalus shunt. Acta Neurol Scand 120(5):317–23

    Article  PubMed  CAS  Google Scholar 

Download references

Conflicts of interest statement

This evaluation study was commissioned by B. Brown Ltd UK, in agreement with the University of Cambridge, which partly contributes to Z.C.’s salary. The study was conducted independently, and the manufacturer’s comments were provided unabridged.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Marek Czosnyka .

Editor information

Editors and Affiliations

Rights and permissions

Reprints and permissions

Copyright information

© 2012 Springer-Verlag/Wien

About this paper

Cite this paper

Czosnyka, M., Czosnyka, Z., Pickard, J.D. (2012). Programmable Shunt Assistant Tested in Cambridge Shunt Evaluation Laboratory. In: Aygok, G., Rekate, H. (eds) Hydrocephalus. Acta Neurochirurgica Supplementum, vol 113. Springer, Vienna. https://doi.org/10.1007/978-3-7091-0923-6_15

Download citation

  • DOI: https://doi.org/10.1007/978-3-7091-0923-6_15

  • Published:

  • Publisher Name: Springer, Vienna

  • Print ISBN: 978-3-7091-0922-9

  • Online ISBN: 978-3-7091-0923-6

  • eBook Packages: MedicineMedicine (R0)

Publish with us

Policies and ethics