Abstract
This chapter illustrates a practical radiation protection problem experienced at an installation of the second PET/CT equipment at a nuclear medicine department. It had to be installed in an existing building and where the choice of localities had to be done for reasons other than radiation protection. The process and the methods for calculations and measurements are described and references made to up-to-date reports describing shielding calculation methodologies for PET and X-ray facilities.
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
References
NCRP (2004) Structural shielding design for medical X-ray imaging facilities. NRCP report no. 147. http://www.ncrponline.org/
Madsen MT, Anderson JA, Halama JR, Kleck J, Simpkin DJ, Votaw JR, Wendt RE III, Williams LE, Yester M (2006) AAPM task group 108: PET and PET/CT shielding requirements. Med Phys 33(1):4–15