Abstract

The present study forms the basis in the development of a new acetabular revision system capable of providing improved primary fixation as well as postoperative electro-stimulation of the adjacent bone stock. The development is aimed at the treatment of patient individual situations achieving enhanced fixation of the acetabular cup by additional polyaxial and angular stable fixation pegs as well as a modular adaptable lateral flap.

Electrostimulation of the bone was realized by the senior author in form of the BISS system indicated for fracture healing and treatment of avascular necrosis of the femoral head. In the present research work a new concept was developed to introduce electrostimulation of the bone in the reconstruction of defects in case of acetabular cup revision. Thereby, the secondary coils, which are the source for generation of the electric field in the surrounding bone stock, are placed inside the fixation pegs of the revision cup. The external oscillating magnetic field needed for energy transfer into the human body is generated by a primary coil placed around the patient’s hip.

We conducted preclinical examinations of the acetabular revision system with regard to the safety of the angular stable locking mechanism of the fixation pegs. In addition, the concept of integrating electro-stimulation into the fixation pegs is demonstrated. In conclusion, a promising concept to satisfy the needs for sufficient primary implant stability and secondary periprosthetic bone regeneration is presented.

This is a preview of subscription content, log in via an institution to check access.

Access this chapter

Chapter
USD 29.95
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
eBook
USD 429.00
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
Softcover Book
USD 549.99
Price excludes VAT (USA)
  • Compact, lightweight edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info

Tax calculation will be finalised at checkout

Purchases are for personal use only

Institutional subscriptions

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. Malchau H, Herberts P, Eisler T et al. (2002) The swedish total hip replacement register. J Bone Joint Surg Am 84-ASuppl 2:2–20

    Google Scholar 

  2. Kurtz S, Mowat F, Ong K et al. (2005) Prevalence of primary and revision total hip and knee arthroplasty in the United States from 1990 through 2002. J Bone Joint Surg Am 87:1487–1497

    Article  Google Scholar 

  3. D’Antonio J A, Capello W N, Borden L S et al. (1989) Classification and management of acetabular abnormalities in total hip arthroplasty. Clin Orthop Relat Res 420:96–100

    Google Scholar 

  4. Kluess D, Martin H, Mittelmeier W et al. (2007) Finite-Element-analysis into a newly developed implant system for acetabular cup revision. 53rd Annual Meeting of the Orthopaedic Research Society, San Diego

    Google Scholar 

  5. Krüger T, Hein W, Plitz W et al. (2000) Einfluss der magnetisch induzierten Elektrostimulation (MIES) auf die Kallusfestigkeit nach Distraktion der Schaftstibia. Osteologie 9:157–164

    Google Scholar 

  6. Mittelmeier W, Lehner S, Kraus W et al. (2004) BISS: concept and biomechanical investigations of a new screw system for electromagnetically induced internal osteostimulation. Arch Orthop Trauma Surg 124:86–91

    Article  Google Scholar 

  7. Aaron RK, Ciombor DM, Simon BJ (2004) Treatment of nonunions with electric and electromagnetic fields. Clin Orthop Relat Res 419:21–29

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Daniel Kluess .

Editor information

Editors and Affiliations

Rights and permissions

Reprints and permissions

Copyright information

© 2009 Springer-Verlag Berlin Heidelberg

About this paper

Cite this paper

Kluess, D., Souffrant, R., Bader, R., van Rienen, U., Ewald, H., Mittelmeier, W. (2009). A New Concept of an Electrostimulative Acetabular Revision System with Patient Individual Additional Fixation. In: Vander Sloten, J., Verdonck, P., Nyssen, M., Haueisen, J. (eds) 4th European Conference of the International Federation for Medical and Biological Engineering. IFMBE Proceedings, vol 22. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-540-89208-3_440

Download citation

  • DOI: https://doi.org/10.1007/978-3-540-89208-3_440

  • Publisher Name: Springer, Berlin, Heidelberg

  • Print ISBN: 978-3-540-89207-6

  • Online ISBN: 978-3-540-89208-3

  • eBook Packages: EngineeringEngineering (R0)

Publish with us

Policies and ethics