Skip to main content

Part of the book series: Clinical Hypertension and Vascular Diseases ((CHVD))

Abstract

With the increasing marketing of automated and semiautomated devices for the measurement of blood pressure, there is a need for potential purchasers to be able to satisfy themselves that such devices have been evaluated according to agreed-upon criteria (1). With this need in mind, the Association for the Advancement of Medical Instrumentation (AAMI) published a standard for electronic or aneroid sphygmomanometers in 1987 (2), which included a protocol for the evaluation of the accuracy of devices; this was followed in 1990 by the protocol of the British Hypertension Society (BHS) (3). Both protocols were revised in 1993 (4,5). These protocols, which differed in detail, had a common objective, namely the standardization of validation procedures to establish minimum standards of accuracy and performance and to facilitate comparison of one device with another (6).

This is a preview of subscription content, log in via an institution to check access.

Access this chapter

Chapter
USD 29.95
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
eBook
USD 149.00
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever

Tax calculation will be finalised at checkout

Purchases are for personal use only

Institutional subscriptions

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. O’Brien E, Waeber B, Parati G, Staessen G, Myers MG, on behalf of the European Society of Hypertension Working Group on Blood Pressure Monitoring. Blood pressure measuring devices: validated instruments. BMJ 2001;322:531–536.

    Article  PubMed  CAS  Google Scholar 

  2. Association for the Advancement of Medical Instrumentation. The National Standard of Electronic or Automated Sphygmomanometers. Arlington, VA, AAMI, 1987.

    Google Scholar 

  3. O’Brien E, Petrie J, Littler W, et al. The British Hypertension Society Protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems. J Hypertens 1990;8:607–619.

    Article  PubMed  CAS  Google Scholar 

  4. O’Brien E, Petrie J, Littler WA, et al. The British Hypertension Society Protocol for the evaluation of blood pressure measuring devices. J Hypertens 1993;11(suppl 2):S43–S63.

    Google Scholar 

  5. American National Standard. Electronic or automated sphygmomanometers. Association for the Advancement of Medical Instrumentation, Arlington, VA, 1993.

    Google Scholar 

  6. O’Brien E, Atkins N. A comparison of the BHS and AAMI protocols for validating blood pressure measuring devices: can the two be reconciled? J Hypertension 1994;12:1089–1094.

    CAS  Google Scholar 

  7. O’Brien E, Pickering T, Asmar R, et al., with the statistical assistance of Atkins N and Gerin W on behalf of the Working Group on Blood Pressure Monitoring of the European Society of Hypertension. International protocol for validation of blood pressure measuring devices in adults. Blood Press Monit 2002;7:3–17.

    Article  PubMed  Google Scholar 

  8. O’Brien E, Mee F, Atkins N, O’Malley K. Inaccuracy of the Hawksley randomzero sphygmomanometer. Lancet 1990;336:1465–1468.

    Article  PubMed  CAS  Google Scholar 

  9. O’Brien E, Mee F, Atkins N, O’Malley K. Inaccuracy of seven popular sphygmomanometers for home-measurement of blood pressure. J Hypertens 1990;8:621–634.

    Article  PubMed  CAS  Google Scholar 

  10. O’Brien E, More O’Ferrall J, et al. An evaluation of the Accutracker II non-invasive ambulatory blood pressure recorder according to the AAMI standard. J Amb Monit 1991;4:27–33.

    Google Scholar 

  11. O’Brien E, Atkins N, Mee F, O’Malley K. Evaluation of the SpaceLabs 90202 according to the AAMI standard and BHS criteria. J Hum Hypertens 1991;5:223–226.

    PubMed  CAS  Google Scholar 

  12. O’Brien E, Mee F, Atkins N, O’Malley K. Accuracy of the Del Mar Avionics Pressurometer IV determined by the British Hypertension Society Protocol. J Hypertens 1991;9(Suppl 5):S1–S7.

    Google Scholar 

  13. O’Brien E, Mee F, Atkins N, O’Malley K. Accuracy of the Novacor DIASYS 200 determined by the British Hypertension Society Protocol. J Hypertens 1991;9(Suppl 5):S9–S15.

    Google Scholar 

  14. O’Brien E, Mee F, Atkins N, O’Malley K. Accuracy of the Takeda TM-2420/TM-2020 determined by the British Hypertension Society Protocol. J Hypertens 1991;9(Suppl 5):S17–S23.

    Google Scholar 

  15. O’Brien E, Mee F, Atkins N, O’Malley K. Accuracy of the SpaceLabs 90207 determined by to the British Hypertension Society Protocol. J Hypertens 1991;9(Suppl 5):S25–S31.

    Google Scholar 

  16. O’Brien E, Mee F, Atkins N, O’Malley K. Accuracy of the SpaceLabs 90207, Novacor DIASYS 200, Del Mar Avionics Presurometer IV and Takeda TM-2420 ambulatory systems according to the AAMI and BHS criteria. J Hypertens 1991;9(Suppl 6):S332–S333.

    CAS  Google Scholar 

  17. O’Brien E, Mee F, Atkins N, Halligan A, O’Malley K. Accuracy of the SpaceLabs 90207 ambulatory blood pressure measuring system in normotensive pregnant women determined by the British Hypertension Society Protocol. J Hypertens 1993;11(Suppl 5):869–873.

    PubMed  Google Scholar 

  18. O’Brien E, Mee F, Atkins N, O’Malley K. Short Report: accuracy of the Dinamap portable monitor, Model 8100, determined by the British Hypertension Society protocol. J Hypertens 1993;11:761–763.

    Article  PubMed  CAS  Google Scholar 

  19. O’Brien E, Mee F, Atkins N, O’Malley K. Accuracy of the CH-Druck/Pressure ERKA ambulatory blood pressure measuring system determined by the British Hypertension Society protocol. J Hypertens 1993;11(Suppl 2):S1–S7.

    Google Scholar 

  20. O’Brien E, Mee F, Atkins N, O’Malley K. Accuracy of the Profilomat ambulatory blood pressure measuring system determined by the British Hypertension Society protocol. J Hypertens 1993;11(Suppl 2):S9–S15.

    Google Scholar 

  21. Mee F, O’Brien E, Atkins N, O’Malley K. Comparative accuracy of the CHDruck, Profilomat, SpaceLabs 90207, DIASYS 200, Pressurometer IV and Takeda TM-2420 ambulatory blood pressure measuring (ABPM) devices determined by the British hypertension Society (BHS) protocol. J Human Hypertens 1993;7:98.

    Google Scholar 

  22. O’Brien E, Mee F, Atkins N, Thomas M. Evaluation of three devices for self-measurement of blood pressure: according to the revised British Hypertension Society Protocol: the Omron HEM-705CP, Phillips HP5332, and Nissei DS-175. Blood Press Monit 1996;1:55–61.

    PubMed  Google Scholar 

  23. Mee F, Atkins N, O’Brien E. Evaluation of the Profilomat II® ambulatory blood pressure system according to the protocols of the British Hypertension Society and the Association for the Advancement of Medical Instrumentation. Blood Press Monit 1999;3:353–361.

    Google Scholar 

  24. O’Brien E, Mee F, Atkins N. Evaluation of the Schiller BR-102 ambulatory blood pressure system according to the protocols of the British Hypertension Society and the Association for the Advancement of Medical Instrumentation. Blood Press Monit 1999;4:35–43.

    Article  PubMed  CAS  Google Scholar 

  25. Conroy R, O’Brien E, O’Malley K, Atkins N. Measurement error in the Hawksley random zero sphygmomanometer: What damage has been done and what can we learn? BMJ 1993;306:1319–1322.

    PubMed  CAS  Google Scholar 

  26. O’Brien E, Petrie J, Littler WA, et al. Blood Pressure Measurement: Recommendations of the British Hypertension Society, 3rd ed. BMJ Publishing Group, 1997.

    Google Scholar 

  27. O’Brien E, Mee F, Atkins N, O’Malley K, Tan S. Training and assessment of observers for blood pressure measurement in hypertension research. J Human Hypertens 1991;5:7–10.

    CAS  Google Scholar 

  28. Mee F, Atkins N, O’Brien E. Problems associated with observer training in blood pressure measurement. J Human Hypertens 1994;8:293.

    Google Scholar 

  29. O’Brien E, Atkins N, Mee F, Coyle D, Syed S. A new audio-visual technique for recording blood pressure in research: the Sphygmocorder. J Hypertens 1995;13:1734–1737.

    PubMed  CAS  Google Scholar 

  30. Atkins N, O’Brien E, Wesseling KH, Guelen I. Increasing observer objectivity with audio-visual technology: the sphygmocorder. Blood Press Monit 1997;2:269–272.

    PubMed  Google Scholar 

  31. Forsberg SA, DE Guzman M, Berlind S. Validity of blood pressure measurement with cuff in the arm and forearm. Acta Med Scand 1979;188:389–396.

    Article  Google Scholar 

  32. Blackburn H, Kihlberg J, Brozek J. Arm versus forearm blood pressure in obesity. Am Heart J 1965;69:423.

    Article  PubMed  CAS  Google Scholar 

  33. Rytand DA, Boyer SH. Auscultatory determination of arterial pressure at wrist and ankle. Am J Med 1954;17:112.

    Article  Google Scholar 

  34. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;i:307–310.

    Google Scholar 

  35. Longo D, Bertolo O, Toffanin G, Frezza P, Palatini P. Validation of the A&D UA-631 (UA-779 Life Source) device for self-measurement of blood pressure and relationship between its performance and large artery compliance. Blood Press Monit 2002;7:243–248.

    Article  PubMed  Google Scholar 

  36. Longo D, Toffanin G, Garbelotto R, Zaetta V, Businaro L, Palatini P. Performance of the UA-787 oscillometric blood pressure monitor according to the European Society of Hypertension Protocol. Blood Press Monit 2003;8:91–95.

    Article  PubMed  Google Scholar 

  37. El Assad M, Topouchian J, Asmar R. Evaluation of two devices for self-measurement of blood pressure according to the international protocol: the Omron M5-I and the Omron 705IT. Blood Press Monit 2003;8(3):127–133.

    Google Scholar 

  38. O’Brien E, Atkins N, Murphy A, Lyons S. Validation of the Rossmax blood pressure measuring monitor according to the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults. Blood Press Monit 2003;8:261–265.

    Article  PubMed  Google Scholar 

  39. O’Brien E, Atkins N, Murphy A, Lyons S. Validation of the Tonoport blood pressure measuring monitor according to the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults. Blood Press Monit 2003;8:261–265.

    Article  PubMed  Google Scholar 

  40. Graves JW, Tibor M, Murtagh B, Klein L, Sheps SG. The Accoson Greenlight 300—the first non-automated mercury-free blood pressure measurement device to pass the International Protocol for blood pressure measuring devices in adults. Blood Press Monit 2004;9:13–17.

    Article  PubMed  Google Scholar 

  41. Nolly H, Romero M, Nolly A, Osso P, Reinoso O, Nolly M. Home blood pressure measurement: validation of the Braun BP 2550 (UG) monitor according to the ESH International Protocol. Blood Press Monit 2004;9:53–58.

    Article  PubMed  Google Scholar 

  42. Jones SC, Bilous M, Winship S, Finn P, Goodwin J. Validation of the OSCAR 2 oscillometric 24-hour ambulatory blood pressure monitor according to the International Protocol for the validation of blood pressure measuring devices. Blood Press Monit 2004;9:219–223.

    Article  PubMed  Google Scholar 

  43. El Assaad M, Topouchian J, Orobinskaia L, Asmar R. Validation de l’appareil de mesure automatique de la pression artérielle au niveau du poignet “OMRON RX-3” selon le protocole international P94 at XXIVes journées de l’HyperTension Artérielle Paris, Palais des Congrès, December 16–17, 2004.

    Google Scholar 

  44. Haensel A, Utech K, Langewitz W. Validation of the TONOPORT V blood pressure measuring monitor in adults. J Human Hypertens 2005;19:745–749.

    Article  CAS  Google Scholar 

  45. Mallion JJM, Pierre H, Neuder Y, Ormezzano O, Baguet JP. Validation of the AGILIS ambulatory blood pressure monitor according to the European Society of Hypertension International Protocol for Validation of Blood Pressure Measuring Devices in Adults. Blood Press Monit 2005;10:97–101.

    Article  Google Scholar 

  46. Topouchian JA, El Assad MA, Orobinskaia LV, El Feghali RN, Asmar RG. Validation of two devices for self-measurement of brachial blood pressure according to the International Protocol of the European Society of Hypertension: the SEINEX SE-9400 and the Microlife BP 3AC1-1. Blood Press Monit 2005;10:325–331.

    Article  PubMed  Google Scholar 

  47. Pereira T, Maldonado J. Validação de aparelho de medição automática da pressão arterial COLSON MAM BP3AA1-2 segundo o Protocolo da Sociedade Europeia de Hipertensão Rev Port Cardiol 2005;24(11):1341–1351.

    PubMed  Google Scholar 

  48. Altunkan Ş, Özta ŞK, Altunkan E. Validation of the Omron 637IT wrist blood pressure measuring device with a position sensor according to the International Protocol in adults and obese adults. Blood Press Monit 2006;11:79–85.

    Article  PubMed  Google Scholar 

  49. Altunkan Ş, Altunkan E. Validation of the Omron 637IT wrist blood pressure measuring device with a position sensor according to the International Protocol in the elderly Blood Press Monit 2006;11:97–102.

    Article  PubMed  Google Scholar 

  50. Rogoza AN. Accuracy of blood pressure estimation by automated monitors. Functional Diagnostics 2003;1:56–64. (In Russian).

    Google Scholar 

  51. Coleman A, Freeman P, Steel S, Shennan A. Validation of the Omron MX3 Plus oscillometric blood pressure monitoring device according to the European Society of Hypertension international protocol. Blood Press Monit 2005;10:165–168.

    Article  PubMed  Google Scholar 

  52. Stergiou GS, Giovas PP, Neofytou MS, Adamopoulos DN. Validation of the Microlife BP A 100 Plus device for self-home blood pressure measurement according to the International Protocol. Blood Press Monit 2006;11:157–160.

    Article  PubMed  Google Scholar 

  53. Wilton A, Shabeeh H, Cuckson C, Shennan AH. Validation of a non-mercury digital auscultatory device with manual pressure registration (PMS Mandaus). Blood Press Monit 2006;11:161–164.

    Article  PubMed  Google Scholar 

  54. Topouchian JA, El Assaad MA, Orobinskaia LV, El Feghali RN, Asmar RG. Validation of two automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension: the Omron M6 (HEM-7001-E) and the Omron R7 (HEM 637-IT). Blood Press Monit 2006;11:165–171.

    Article  PubMed  Google Scholar 

  55. Reinders A, Reggiori F, Shennan AH. Validation of the DINAMAP ProCare blood pressure device according to the international protocol in an adult population. Blood Press Monit 2006;11:293–296.

    Article  PubMed  Google Scholar 

  56. Li L, Wang Hui W, Jin Lan W. Validation of the Oregon Scientific BPW810 blood pressure monitor according to the European Society of Hypertension Protocol. Blood Press Monit 2006;11:343–347.

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Editor information

Editors and Affiliations

Rights and permissions

Reprints and permissions

Copyright information

© 2007 Humana Press Inc.

About this chapter

Cite this chapter

O’Brien, E., Atkins, N. (2007). Validation and Reliability of Blood Pressure Monitors. In: White, W.B. (eds) Blood Pressure Monitoring in Cardiovascular Medicine and Therapeutics. Clinical Hypertension and Vascular Diseases. Humana Press. https://doi.org/10.1007/978-1-59259-978-3_5

Download citation

  • DOI: https://doi.org/10.1007/978-1-59259-978-3_5

  • Publisher Name: Humana Press

  • Print ISBN: 978-1-58829-512-5

  • Online ISBN: 978-1-59259-978-3

  • eBook Packages: MedicineMedicine (R0)

Publish with us

Policies and ethics