Background

Nitrogen containing heterocyclic ring such as pyrimidine is a promising structural moiety for drug design. Pyrimidine derivatives form a component in various useful drugs and are associated with many biological and therapeutic activities. Condensed pyrimidines have been reported as antimicrobial [1]–[3], anti-inflammatory [4],[5], analgesic [6],[7], anticancer [8]–[10], anti-HIV [11], antitubercular, antimalarial, diuretic, and cardiovascular disease [12] (Scheme 1).

Scheme 1
scheme 1

Chemical structure of BT 10 M.

The present work is a synthesis, a biological evaluation and validation of novel pyrimidine derivatives. Research workers have synthesized 50 pyrimidine derivatives (T1M-T10M, T1P-T10P, BT1M-BT10M, BT1P-BT10P, CT1M-CT5M, and CT1P-CT5P). Among them, BT10M exhibited maximum antimicrobial, anti-inflammatory, and analgesic activity. Hence, a validation study was done on BT10M. BT10M is chemically [6-Bromo-3-(6-(2,6-dichlorophenyl)-2-(morpolinomethylamino) pyrimidine4-yl) -2H-chromen-2-one]. It is a yellow crystalline powder with a molecular formula of C24H19BrCl2N4O3 and a molecular weight of 562.24. It is a potent antimicrobial, analgesic and anti-inflammatory agent among all the synthesized derivatives. Hence, the aim of present investigation is to develop a simpler, rapid, and cost-effective analytical method for the determination of pyrimidine derivative (BT10M) in bulk dosage form suitable for routine quality control analysis.

Method validation is the process used to confirm that analytical procedure employed for a specific test is suitable for its intended use. It is an integral part of any good analytical practice. Methods need to be validated or revalidated [13].

Methods

Chemical and reagent

BT10M was synthesized by research workers and then validated. Methanol and acetonitrile (1:1) were used throughout spectrophotometric method development and validation.

Instrumentation

The method was performed on a double-beam ultraviolet-visible (UV-VIS) spectrophotometer (Shimadzu model 1700 (Shimadzu, Kyoto, Japan)) having two matched quartz cells with a 1-cm light path.

Determination of maximum wavelength (λmax), methodology, and sample preparation

About 50 mg of BT10M was weighed accurately and transferred into a 50-ml volumetric flask and dissolved in 25 ml of methanol and acetonitrile (1:1) and made up to the volume with the same solvent mixture to give a standard concentration of 1,000 μg/ml. Transfer 5 ml of above solution into the 50-ml volumetric flask, dilute, and made up to the volume with the same solvent mixture to get a standard concentration of 100 μg/ml. This solution was scanned against a blank over the entire UV-VIS wavelength of 200 to 400. Based on the spectrum, a λmax of 275 nm was selected for further analysis.

Results

The method was validated with respect to linearity, accuracy, precision, specificity, robustness, ruggedness, LOD, and LOQ in Table 1.

Table 1 Validation summary

Discussion

The method was validated with respect to linearity, accuracy, precision, specificity, robustness, ruggedness, LOD, and LOQ. The method was established according to the International Conference on Harmonisation of Technical Requirements for Registration of Pharmaceuticals for Human Use (ICH) guidelines. BT10M exhibited maximum absorption at 275 nm and obeyed Beer's law in the concentration range of 50 to 150 μg/ml. The proposed method for the determination of BT10M showed linear regression y = 0.005x + 0.025 with a coefficient correlation (r2) of 0.997 (Figure 1). The precision was determined by the relative standard deviation of the six-assay sample of BT10M, and the assay of each was calculated and the obtained relative standard deviation of % assay was less than 1.5%. The percentage recovery for BT10M was found in the range of 98.97% to 99.83% which indicates that the developed method was simple, rapid, and precise. LOD was found to be 145.2 and limit of quantitation to be 440.0. The proposed method will be suitable for the analysis of newly synthesized pyrimidine derivative (BT10M) in bulk dosage form.

Figure 1
figure 1

Linearity plot. The BT10M absorbance response in the concentration range of about 50 mcg/ml to 150 mcg/ml was found to be linear to the analyte concentration in the solution with a correlation coefficient (r2) of 0.997.

Experimental

Validation

The methods were validated with respect to linearity, accuracy, precision, specificity, ruggedness, robustness, limit of detection (LOD), and limit of quantitation (LOQ).

Linearity

The linearity of an analytical method is its ability to elicit test results that are directly, or by a well-defined mathematical transformation, proportional to the concentration of analyte in the samples within a given range. For assay determination, the concentration of BT10M is 100 μg/ml. So the working range of analyte was set between 50, 62.5, 75, 87.5, 100, 112.5, 125, 137.5 and 150 μg/ml to show the linearity of the curve obtained. The observations and calibration curve are shown in Table 2 and Figure 1.

Table 2 Linearity table of BT 10 M in working standard

Accuracy (by recovery test)

Accuracy of method is by shown by recovery study and spiking working standard in the placebo at levels 80%, 100%, and 120% of the working standard. Recovery study was performed by spiking in BT10M to the placebo at levels 80%, 100%, and 120% of working standard. The samples were prepared according to the assay procedure. The results are shown in Tables 3 and 4.

Table 3 Accuracy reading
Table 4 Deviation from recovery

The percentage recovery for BT10M was found in the range of 98.97% to 99.83% with an overall relative standard deviation (%RSD) of 0.4366. From the data obtained which was given in Table 3, the method was found to be accurate. Formula of standard deviation was SD = (xi − x/n − 1)1/2 if n is very large. In case of very small data, SD = (xi − x/n)1/2.

Precision

The precision of an analytical method is the degree of agreement among individual test results when the method is applied repeatedly to multiple sampling of homogenous sample. The precision of an analytical method is usually expressed as the standard deviation or relative standard of a series of measurements. Assay preparation and standard preparation were prepared as per method of analysis of six BT10M assay sample preparations as per the experimental conditions in method of analysis. Calculated percent of BT10M in each assay sample percent by spectrophotometry and the results and observation are summarized in Tables 5 and 6.

Table 5 System precision data of BT 10 M working standard solution
Table 6 Method precision data for estimation of BT 10 M

Assay

%Assay= Abs . of S m p . Abs . of standard × W t . of standard D i l . factor × D i l . factor W t . of S m p ×100

The precision will be determined by the relative standard deviation of the six-assay sample of BT10M, and the assay of each is calculated and the obtained relative standard deviation of % assay should be less than 1.5%. The %RSD shows that precision of the method was satisfactory.

Specificity

Specificity study is designed to prove that the BT10M in the solution gives maximum absorbance at wavelength 275 nm, and there is no interference from the solvent. The purpose of this study is to establish the fact that inherent chemical stability of the molecule remains intact during its existence. If any degradation product formed, it can be monitored and resolved to quantify the nature and extent of degradation. For this, the spectrum of BT10M, placebos are studied. The sample preparation is as per methodology. The spectrum of BT10M is shown in Figure 2.

Figure 2
figure 2

Spectrum of BT 10 M.

For the spectrophotometric method, no other component of formulation shows considerable absorbance at the λmax of the analyte or at the detection wavelength of the subject analyte. In this case of BT10M, the detection wavelength is 275 nm. The placebo solution under the same condition does not show any absorbance at 275 nm.

Ruggedness

The ruggedness of the analytical method is the degree of reproducibility of test results obtained by the analysis of the sample on different days, by different chemist using different instruments. In this study, two individual assay sample preparations of BT10M drug product were prepared by different chemists for analysis. Six (6) replicate observations of the same standard solution were obtained as well as six observations of different sample solution were recorded. The assay percentage of each sample was calculated in each case. The results are summarized in Tables 5, 6, 7, 8, 9.

Table 7 System precision data of BT 10 M working standard solution
Table 8 Method precision data for BT 10 M
Table 9 Mean % assay and %RSD

The ruggedness (inter-day precision) will be determined by the relative standard deviation of the results of assay of two different chemists on different days.

Robustness

The robustness of an analytical method is a measure of its capacity to remain unaffected by small but deliberate variations in method parameters and provides an indication of its reliability under normal usage. Method robustness was determined by analyzing the same sample at normal operating conditions and also by changing some operating analytical conditions. The result and observation are summarized in Table 10.

Table 10 Percentage deviation for sample under both conditions

The robustness will be determined by the relative standard deviation of the results of assay of two different conditions by a change in original parameter.

Limit of detection and limit of quantitation

Limit of detection

LOD is defined as the lowest concentration of an analyte in a sample that can be detected, not quantified.

Limit of quantitation

The lowest concentration of an analyte in a sample that can be determined with acceptable precision and accuracy under the stated operational conditions of the method.

L O D : 3.3 S D / Slope 3.3 0.200 / 0.005 S D = 0.2200 = 145.2 mg 2 . L O Q : 10 S D / Slope 10 0.2200 / 0.005 = 440.00 mg

Conclusions

All the validation parameters for all the developed methods were studied as per ICH guidelines. All the methods were found to be accurate, simple, specific, selective, precise, and reproducible. Hence, the method can be used for routine analysis of BT10M in bulk dosage form.