Abstract
Background
In total knee arthroplasty (TKA) a flexible intramedullary rod can be used to account for sagittal bowing of the distal femur. Although patients report better post-operative functional outcome when the flexible rod was used, it is unknown how the use of the flexible rod affects the placement of the femoral TKA component, and how this relates to activities of daily living. It is expected that the use of the flexible rod will result in a more flexed femoral component, a larger patellar tendon moment arm, and consequently in better functional outcome. The goal of this study is to compare the flexible rod to the standard intramedullary rod in primary TKA in terms of fit of the TKA, functional outcome, and sizing of the femoral component.
Methods
A single-blind randomized controlled trial with two groups (flexible vs standard rod), with patients blinded for group allocation, and 2 years post-operative follow-up. The fit of the TKA is quantified by two parameters: (1) the flexion angle of the TKA in the sagittal plane and (2) the sagittal profile of the distal femur compared between the pre-operative bone and the TKA. Both parameters are calculated in 3D volume images obtained using fluoroscopy. Functional outcome will be measured using (1) the timed Get-up and Go test (2), the stair climbing test (3), knee power output, and (4) patient and clinician reported outcomes. Different parameters will be measured during the TKA procedure to account for the invisibility of cartilage in the 3D volumes and to study if the amount of bone removed during the procedure is affected by group allocation.
Discussion
The sagittal fit of TKA is not a standardized outcome measure. We discuss our choice of parameters to define the sagittal fit (i.e., flexion angle and sagittal profile), our choice for the parameters we measure during the TKA procedure to account for the lack of cartilage thickness in fluoroscopy, and our choice for the parameters to study if the amount of bone removed during the procedure is affected by group allocation. Lastly, we discuss the merits of this planned trial.
Trial registration
Netherlands Trial Register, 4888, registered 30 March 2015. https://www.trialregister.nl/trial/4888
Similar content being viewed by others
Background
Total knee arthroplasty (TKA) is a cost-effective surgical procedure for degenerative knee osteoarthritis. Although TKA has good long-term results, there are still some patients with poor long-term results [13]. Also, the good results are not always related to patient satisfaction and functional outcome [3, 7]. This may be caused by the fact that the kinematics of the TKA do not properly mimic the kinematics of the natural knee. Manufacturers of TKAs have addressed this by integrating aspects of natural knee kinematics into their TKA designs and instruments [9].
One aspect influencing natural knee kinematics is sagittal placement of the femoral component [18]. The instruments used for alignment of the femoral component are classically used intramedullary, and implanting the femoral component can be performed either according to the longitudinal axis of the femur in the sagittal plane or according to the anatomy of the distal femur, taking into account sagittal bowing of the femur. The amount of sagittal bowing can differ between individuals [30], and therefore to account for sagittal bowing during placement of the femoral component, a flexible intramedullary rod (flex rod) can be used instead of the standard intramedullary rod [17].
Up till now, the literature has not been conclusive which of the rods should be used, and acknowledges the dilemma of implanting the femoral component either according to the anatomy of the distal femur, or according to the longitudinal axis of the femur in the sagittal plane, thus ignoring sagittal bowing. The standard rod is considered a safe option. However, it is hypothesized that with the flex rod, less bone needs to be removed because the femoral component better fits to the natural anatomy of the distal femur in the sagittal plane. This would lead to more natural femorotibial kinematics through increasing the patellar tendon moment arm, and thus the possibility for larger knee power output without influencing muscle strength. This effect has been found in a modeling study [24] and could potentially lead to better functional outcome in vivo. Indeed, patient-reported functional outcome is higher up till 2 years in patients operated using the flex rod when compared the standard rod [17]. It is expected that due to less oversizing in the TKA procedure using the flex rod activities of daily living (ADL) such as stairclimbing, and rising from a chair can be performed more optimally [6]. Marra et al. have also shown in their modeling study that less oversizing resulted in better patellofemoral kinematics during rising from a chair [24]. It is also expected that with the use of the flex rod, the proper fit of the TKA will result in an overall smaller size chosen by the surgeon, as has been found in a retrospectively analyzed cohort and a modeling study [8, 17].
Methods/design
Aim
The primary aim of this study is to compare the flex rod with the standard rod in terms of post-operative sagittal fit of the distal femur. In addition, sizing and post-operative functional outcome will be evaluated.
Design and setting
A single-blind randomized controlled trial with two groups (flex rod vs standard rod), with the participants blinded for group allocation. Participants will be measured on five occasions in our outpatient clinic (pre-operative, per-operative/post-operative, 3 months, 1 year, and 2 years post-operative). The study ends after 2- year follow-up. The trial has been set up according to CONSORT guidelines [29], and ethical approval for this study was obtained from Medisch-ethische toetsingscommissie Slotervaartziekenhuis en Reade (approval ID P1453).
Participants
Subjects will be selected from the waiting list for patients scheduled for elective TKA in our hospital. Inclusion and exclusion criteria are shown in Table 1. Eligible subjects will be contacted, after their surgeon’s consent, by phone by a research nurse explaining the study. When they are interested in participating, the patient information sheet will be send. Eligible patients are contacted again after a minimum of 1 week for verbal consent. Written informed consent is obtained at the first study contact in our outpatient clinic.
Description of materials
All participants will receive the posterior stabilized model Triathlon (Stryker, Mahwah, NJ, USA). This model is used to standardize the posterior placement of the femoral component (in contrast to a cruciate-retaining model), which is expected to yield differences in the anterior part of the distal femur and therefore differences in patellofemoral function between the groups.
TKA procedure
The surgical technique is a standardized bone-referenced technique. A tibial bone cut is made based on the standard extramedullary jig from the system aiming at resection of 9 mm bone from the unworn lateral side, or 3 mm below a worn medial compartment. The amount of valgus of the distal cut femur jig is set on the femur angle measured on hip-knee-ankle radiographs of the individual patient. A special custom-made aiming device with an offset of 5 mm (for the flex rod) and 10 mm (for the standard rod) referencing from the intercondylar notch is used to standardize the entry points of the two rods (Fig. 1). Sizing of the femur component is done with posterior referencing. Standard 3° of femoral external rotation is used in these series. Necessary soft tissue releases are done for gap balancing. A standard patella resurfacing is done, and the total implant is placed cemented.
Investigational product
For this study, the difference between the two groups is the use of the flex rod or the standard rod during TKA. The rods are used in TKA to position the femoral component in the correct sagittal orientation (Fig. 2). The difference between the two rods is that the flex rod allows for bowing in the sagittal plane, thus following the natural anatomy of the femur. There are no extra risks involved with use of the flex rod. Both rods are part of the operating instruments used for placement of the TKA (Stryker, Mahwah, NJ, USA).
Study parameters
The fit of the femoral component is quantified in two parameters: the flexion angle of the TKA in the sagittal plane (Fig. 3), and the sagittal profile at three slices: medial, lateral, and at level of the trochlea (Fig. 4). Both parameters will be measured in a 3D volume image of the knee, reconstructed from images obtained with fluoroscopic assessments using the MultiDiagnost Eleva (MDE, Philips, Amsterdam, The Netherlands). All reconstructions and measurements are performed with custom-written software using Matlab (The MathWorks, Natick, MA, USA), and results are exported to a data sheet for analysis.
Primary study parameter
The flexion angle of the TKA in the sagittal plane is measured to determine difference in femoral component placement between the groups. This angle is commonly defined as the angle ‘γ’ between the bone stub axis (BSA) of the femur [25] and the neutral line of the femoral component of the prosthesis [12] (Fig. 3). Due to artifacts in the fluoroscopic assessment caused by the metal in the prosthesis, the neutral line is unidentifiable. However, the most anterior plane on the inside of the femoral component of the prosthesis, the frontal flange, can be detected in the fluoroscopic assessment (Fig. 3). Therefore, the angle ‘θ’ between the frontal flange and the BSA is measured. The angle between the frontal flange and the neutral line is 97°; thus, the formula γ = θ − 7° can be used for future comparison between gamma and theta measurements. To obtain the angle ‘θ’, the BSA is aligned with the vertical y-axis by coronal and sagittal rotating the 3D volume image. Furthermore, the transversal plane is rotated towards an anatomic orientation to correct any femur rotation deviations from acquisition. The latter is necessary to be able to compare different cases. Once the orientation of the femur is standardized within the 3D volume image, the frontal flange can be manually annotated and compared with the y-axis, the BSA. For annotation, the sagittal slice with the most prominent frontal flange is chosen and to minimize measurement error from manual annotation an average over 5 iterations is used. We hypothesize that the flexion angle will be larger in the flex rod group (i.e., better fit to the distal anatomy of the femur, Fig. 2).
Secondary study parameters
The sagittal profile is calculated as the length of lines connecting the intersection of the posterior condylar offset of the distal femur in the sagittal plane (PCOi), and the outer border of either the native femur or femoral component (Fig 4a). The PCOi must be at the exact same location before and after TKA to allow comparison. For this reason, the 3D volume images from before and after TKA are aligned using an automatic intensity-based images registration algorithm in Matlab, based on the shaft of the femur. For constant measurements, the exact same slice has to be selected in every subject. To achieve these constant measurements we pre-align a random 3D volume image with a prosthesis as calibration set. The 3D volume images after TKA are aligned with this calibration set using the same intensity-based image registration algorithm. Subsequently, the pre-operative images (without TKA) are aligned using the same transformation variables.
Due to artifacts in the 3D volume images, the sagittal profile of the trochlea is unidentifiable and sagittal profiles of the condyles are affected. For the best representation of the sagittal profiles, the original designs of the manufacturer are converted into a 3D volume image and added with the pre-aligned calibration set. To avoid image registration errors a pre-aligned calibration set is created for each prosthesis size.
The sagittal profiles of the native femur and the TKA will be calculated at 3 slices: (Fig. 4b) (1) the slice through the most distal point on the medial condyle, oriented at 17.1% of the prosthesis width from the medial side; (2) the slice through the trochlea at the most cranial point of the distal femur, oriented at exactly the middle of the prosthesis; and (3) the slice through the most distal point on the lateral condyle, oriented at 17.1% of the prosthesis width from the lateral side.
One goal of the TKA procedure is restoration of the joint line. This would mean that the amount of implant material at the medial and lateral condyle in full extension of the knee is ± 1 mm from the native situation (DIST, Fig. 4b). A second goal is restoration of the flexion radius. This is operationalized as the amount of implant material at straight posterior at 90° of knee flexion (POST, Fig. 4b) which should be the same as in the native situation (± 1 mm). For mid flexion stability (at 30° of knee flexion), it is also important that the amount of implant material equals the native situation (FLEX30, Fig. 4b). Patellofemoral function can be determined at the trochlear slice. At PCO straight anterior (ANT, Fig. 4b), the amount of implant should equal the amount of native bone (± 1 mm). When the knee flexes, the patella rests on the medial and lateral condyles instead of in the trochlea. So also on the medial and lateral slices at 30° from straight anterior, the length will be calculated (ANT30, Fig 4b). The lengths of the sagittal profile lines are automatically calculated based on pixel density as representative for bone density. In case the segmentation of the femur is incorrect due to artifacts (caused by other bone density than the algorithm expects due to osteoporosis, sclerosis, lower radiation dose, etc.), the lengths are manually corrected. For all five lines, the length in the native situation is subtracted from the length in the component, resulting in a positive sign when the implant is bigger than the native femur (i.e., “overstuffing”), and a negative sign when the implant is smaller than the native femur (i.e., “understuffing”). The margins are defined as ± 1 mm (range of 2 mm). We hypothesize that the differences of the anterior and posterior sagittal contours on the three slices between the pre-operative situation and the post-operative image will be smaller in the flex rod group (i.e., better fit). In the images of the native femur, articular cartilage cannot be seen [20]. Therefore, an average cartilage thickness of 2 mm will be added to the lengths in the native femur [20].
Functional outcome
Functional outcome will be measured using (1) the timed Get-up and Go test, to measure functional mobility [27]; (2) the stair climbing test to measure ability to ascend and descend a flight of stairs of 14 steps [28]; (3) the Leg Extensor Power Rig (Queens Medical Centre, Nottingham, UK) to measure knee power output [4, 15]. The rig consists of a seat and footplate connected via a lever and chain to a flywheel. Application of force accelerates the flywheel from rest and output is recorded as both maximal wattage (W) generated and as relative power-bodyweight ratio (%) of a single leg extension. The power output is recorded five times per leg and the highest output of the leg in study will be used for analysis and (4) Oxford Knee Score [10], KOOS-PS [26], KUJALA patella score [22], and Knee Society Score [19] to measure patient and clinician reported outcomes. Table 2 shows the schedule for the functional outcome measurements. The hypothesis is that patients in the flex rod group will have better results on all functional outcome scores.
Other study parameters
Baseline information
The following demographic and disease-related data will be collected pre-operatively: age, gender, height, weight, strength (Medical Research Council scale), physical activities (UCLA activity score [2]), primary diagnosis, side, alignment of involved knee, and previous surgery of the affected knee.
Operative information
See Table 2 for an overview of the parameters recorded during surgery.
Radiological information
The following radiological parameters will be measured (see Table 3 for an overview of the radiology schedule): (1) posterior condylar offset (PCO) measured on regular true lateral X-ray [5], (2) patellar tendon angle (PTA) measured on weight bearing with 45° flexion lateral X-ray; (3) lateral patellar tilt [23], and lateral patellar displacement of patellofemoral joint measured on supine sky-line patella X-ray [14], (4) Slope of the tibial component with respect to the tibia measured on the post-operative lateral X-ray.
All patients will receive post-operative standard care according to the usual practice of our hospital. Post-operative blood loss and/or complications will be documented on the case report form. Post-operative data through time of hospital discharge will be reported on the discharge form.
Statistical analysis
Sagittal flexion angle of the TKA will be reported as mean and standard deviation (or median and interquartile range (IQR)). An independent two-sample t test (or non-parametric equivalent) will be used to analyze the effect of group (flex rod versus standard rod). All secondary outcome parameters will be reported as mean and SD (or median and IQR). Repeated mixed model analysis techniques will be used to analyze the effects of group, time, and their interaction. All other study parameters will be reported as mean and standard deviation (or median and IQR). Independent t tests (or non-parametric equivalent) and Chi square (in case of frequencies) will be used to analyze the effect of group. All statistical analyses will be performed using Stata/IC for Windows, Release 13 (StataCorp LLC, College Station, TX, USA).
The sample size calculation is based on the primary study parameter: flexion angle of the prosthesis in the sagittal plane. Our hypothesis is that this angle at 3 months will be smaller in the standard rod group compared to the flex rod group. First, the flexion angle in posterior stabilized TKA has been reported to be about 6° [1]. Second, the mean intra-observer difference between measurements for this angle is 2.3° with an estimated standard deviation of 5° [16]. We assume that a clinical relevance of a mean difference of 5° between the rod groups can be measured [21], but this will be further studied in our study. Using G*Power (version 3.1.7 [11]), a sample size was calculated for an independent two-sample t test, with the variables m1 = 6°, m2 = 11°, σ1 = 5°, σ2 = 5°, α = 0.05, and power = 0.80. This calculation yielded a sample size for the total group of N = 54, with 27 for each group. Taking into account a drop out of participants during the course of the study of 10%, a sample size of N = 60 subjects is needed, with 30 subjects in each group.
Discussion
This randomized controlled trial has been set up to assess how the use of a flex rod affects the fit of the TKA and if this results in better functional outcome. As the sagittal fit of TKA is not a standardized outcome measure, we chose to define the fit with two outcome measures: the first being the flexion angle of the femoral component. It is expected that the flexion angle will be larger in the flex rod group as the femoral component is expected to have a better fit to the distal anatomy of the femur. The second outcome measure is the contour of the distal femur, quantified in the anterior sagittal profile and the posterior sagittal profile at various positions. It is expected that for the flex rod group, due to smaller sizing of the femoral component, the distal anatomy of the femur with TKA will be closer to the pre-operative distal anatomy. We expect that this will be most apparent in the anterior profile where the effect of overstuffing after use of the standard rod is expected to be largest [17]. Also, for this reason, patients in this trial will receive a posterior stabilized TKA, standardizing the posterior placement of the femoral component (in contrast to a cruciate-retaining TKA), and with differences in the anterior part of the distal femur as the only degree of freedom.
Custom-made software is used for calculating the parameters of sagittal fit of the TKA. Due to the unavailability of existing software for measuring the specific parameters for the specific model of TKA, this software was produced in-house. The advantages of this software is that it is tailored made for the study-specific parameters. Disadvantages of this software are that currently we have no data on the reliability and validity of the measurements and that by using imaging of the knee, it is unable to account for articular cartilage thickness [20]. During the TKA procedure, the cartilage of the different cuts will be scored. Using this information, the measurements of the sagittal profile can be corrected for cartilage thickness of 0, 1, and 2 mm, with the values based on the literature [20]. Other parameters that are measured during the TKA procedure are the sagittal thickness (in millimeters) of the cuts, and the area of the cuts. It is expected that due to a better fit of the femoral component to the distal anatomy in the flexible IM rod group, less bone has to be cut, resulting in lower thickness and smaller area of these cuts.
We expect that the use of the flex rod will result in an overall better functional outcome. This is supported by Hitt et al. [17], who studied patient and clinician reported outcome measures in a RCT. The modeling study by Marra et al. [24] showed that such effects may be explained by the increase in the patellar tendon moment arm. Our aim is to study the effects of a flex rod in a prospective RCT combining imaging, surgery details, functional outcome, and patient-reported outcomes, thus providing a data set on which our hypotheses can be thoroughly tested.
Availability of data and materials
Not applicable.
Abbreviations
- ADL:
-
Activities of daily living
- BSA:
-
Bone stub axis
- IQR:
-
Interquartile range
- PCO:
-
Posterior condylar offset
- PCOi:
-
Intersection of the posterior condylar offset of the distal femur in the sagittal plane
- PTA:
-
Patellar tendon angle
- RCT:
-
Randomized controlled trial
- TKA:
-
Total knee arthroplasty
References
Almeida F, Martin JM, Silvestre A, Gomar F. Posterior-stabilized total in knee replacement in rheumatoid arthritis. J Orthop. 2008;5(3):e3.
Amstutz H, Thomas B, Jinnah R, Kim W, Grogan T, Yale C. Treatment of primary osteoarthritis of the hip: a comparison of total joint and surface replacement arthroplasty. J Bone Jt Surg Am. 1984;66(2):228–41.
Anderson J, Wixson R, Tsai D, Stolberg S, Chang R. Functional outcome and patient satisfaction in total knee patients over the age of 75. J Arthroplast. 1996;11(7):831–40.
Bassey EJ, Short AH. A new method for measuring power output in a single leg extension: feasibility, reliability and validity. Eur J Appl Physiol Occup Physiol. 1990;60(5):385–90.
Bellemans J, Banks S, Victor J, Vandenneucker H, Moemans A. Fluoroscopic analysis of the kinematics of deep flexion in total knee arthroplasty. Influence of posterior condylar offset. J Bone Jt Surg Br. 2002;84(1):50–3.
Bonnin MP, Schmidt A, Basiglini L, Bossard N, Dantony E. Mediolateral oversizing influences pain, function, and flexion after TKA. Knee Surg Sports Traumatol Arthrosc. 2013;21(10):2314–24.
Bourne RB, Chesworth BM, Davis AM, Mahomed NN, Charron KDJ. Patient satisfaction after total knee arthroplasty: who is satisfied and who is not? Clin Orthop Relat Res [Internet]. 2010 [cited 2015 Oct 7];468(1):57–63. Available from: http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=2795819&tool=pmcentrez&rendertype=abstract.
Chen S, Zeng Y, Yan M, Yue B, Zhang J. Morphological evaluation of the sagittal plane femoral load-bearing surface in computer-simulated virtual total knee arthroplasty implantation at different flexion angles. Knee Surgery Sport Traumatol Arthrosc. 2016;25(9):2880–6.
Cushner FD, Scuderi GR. A modern approach to TKA design. Tech Orthop. 2018;33(1).
Dawson J, Fitzpatrick R, Murray D, Carr A. Questionnaire on the perceptions of patients about total knee replacement. J Bone Jt Surg Br. 1998;80(1):63–9.
Erdfelder E, FAul F, Buchner A, Lang AG. Statistical power analyses using G*power 3.1: tests for correlation and regression analyses. Behav Res Methods. 2009;41(4):1149–60.
Ewald FC. The knee society total knee arthroplasty roentgenographic evaluation and scoring system. Clin Orthop. 1989;248:9–12.
Gill G, Joshi A, Mills D. Total condylar knee arthroplasty: 16- to 21-year results. Clin Orthop Relat Res. 1999;367:210–5.
Grelsamer R, Bazos A, Proctor C. Radiographic analysis of patellar tilt. J Bone Jt Surg Br. 1993;75(5):822–4.
Hamilton DF, Gaston P. Simpson a HRW. Is patient reporting of physical function accurate following total knee replacement? J Bone Joint Surg Br. 2012;94(11):1506–10.
Hirschmann MT, Konala P, Amsler F, Iranpour F, Friederich NF, Cobb JP. The position and orientation of total knee replacement components: a comparison of conventional radiographs, transverse 2D-CT slices and 3D-CT reconstruction. J Bone Joint Surg Br. 2011;93(5):629–33.
Hitt KD, Pierce TP, Jauregui JJ, Cherian JJ, Elmallah RDK, Logan S, et al. Use of a flexible intramedullary rod and its influence on patient satisfaction and femoral size in total knee arthroplasty. J Long-Term Eff Med Implants. 2015;25(3):201–8.
Hood B, Blum L, Holcombe S, Wang S, Urquhart A, Goulet J, et al. Variation in optimal sagittal alignment of the femoral component in total knee arthroplasty. Orthopedics. 2016;40(2):102–6.
Insall JN, Dorr LD, Scott RD, Scott WN. Rationale of the knee society clinical rating system. Clin Orthop. 1989;248:13–4.
Johal P, Hassaballa MA, Eldridge JD, Porteous AJ. The posterior condylar offset ratio. Knee. 2012;19(6):843–5.
Kaya Bicer E, Servien E, Lustig S, Demey G, Ait Si Selmi T, Neyret P. Sagittal flexion angle of the femoral component in unicompartmental knee arthroplasty: is it same for both medial and lateral UKAs? Knee surgery. Sport Traumatol Arthrosc. 2010;18(7):928–33.
Kujala U, Jaakkola L, Koskinen S, Taimela S, Hurme M, Nelimarkka O. Scoring of patellofemoral disorders. Arthroscopy. 1993;9(2):159–63.
Laurin C, Dussault R, Levesque H. The tangential x-ray investigation of the patellofemoral joint: x-ray technique, diagnostic criteria and their interpretation. Clin Orthop Relat Res. 1979;144:16–26.
Marra MA, Strzelczak M, Heesterbeek PJC, Van De Groes SAW, Janssen D. Flexing and downsizing the femoral component is not detrimental to patellofemoral biomechanics in posterior-referencing cruciate- retaining total knee arthroplasty. Knee Surgery Sport Traumatol Arthrosc [Internet]. 2018;26(11):3377–3385. Available from: http://dx.doi.org/https://doi.org/10.1007/s00167-018-4900-z.
Nuño N, Ahmed AM. Sagittal profile of the femoral condyles and its application to femorotibial contact analysis. J Biomech Eng. 2001;123(1):18–26.
Perruccio AV, Stefan Lohmander L, Canizares M, Tennant A, Hawker GA, Conaghan PG, et al. The development of a short measure of physical function for knee OA KOOS-physical function Shortform (KOOS-PS) - an OARSI/OMERACT initiative. Osteoarthr Cartil. 2008;16(5):542–50.
Podsiadlo D, Richardson S. The timed “up & go”: a test of basic functional mobility for frail elderly persons. J Am Geriatr Soc. 1991;39(2):142–8.
Rejeski W, Ettinger W Jr, Schumaker S, James P, Burns R, Elam J. Assessing performance-related disability in patients with knee osteoarthritis. Osteoarthr Cartil. 1995;3(3):157–67.
Schulz K, Altman D, Moher D. CONSORT 2010 statement: updated guidelines for reporting parallel group randomised trials. BMJ. 2010;340:c332.
Yehyawi TM, Callaghan JJ, Pedersen DR, O’Rourke MR, Liu SS. Variances in sagittal femoral shaft bowing in patients undergoing TKA. Clin Orthop Relat Res. 2007;464:99–104.
Acknowledgements
We would like to thank everyone involved in data collection (research nurse, surgeons, and operating room personnel).
Funding
The study is an investigator-initiated study and the hospital received funding from Stryker (Mahwah, NJ, USA) for carrying out all the study procedures.
Author information
Authors and Affiliations
Contributions
All authors made significant contributions to the conception of this study protocol. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
The consent of the patients will be obtained in written form. A detailed description of ethical approval and consent procedures is provided in the study protocol.
Consent for publication
Not applicable.
Competing interests
The authors declare that they have no competing interests.
Additional information
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
About this article
Cite this article
Bénard, M.R., van Doremalen, R.F.M., Wymenga, A.B. et al. Flexible versus standard intramedullary rod in posterior stabilized primary total knee arthroplasty: protocol for a randomized controlled trial. J Orthop Surg Res 15, 472 (2020). https://doi.org/10.1186/s13018-020-01989-9
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s13018-020-01989-9