Introduction

Fourier transform ion cyclotron resonance mass spectrometry (FT-ICR MS) provides the highest resolving power and mass accuracy for molecular structural analysis [16]. Top-down proteomics and petroleomics, because of their extreme requirements for analytical performance, are the primary application areas of FT-ICR MS nowadays [3, 7]. Over the past decade, other application areas of interest were primarily addressed using Orbitrap FTMS and high-resolution time-of-flight (TOF) MS [4, 811]. Nevertheless, the extreme analytical challenges in molecular structural analysis require further increase of FTMS performance. Owing to its fundamental power, further development of FT-ICR MS is feasible and may allow bridging the gap between the analytical requirements and MS capabilities. For that, a corresponding increase in high-performance analysis throughput is required. For example, even at the current data acquisition rate, the exceptional FT-ICR MS capabilities trigger development of novel approaches of molecular structural analysis [12]. Among those, the uses of isotopic fine structure information in qualitative bottom-up proteomics and mass defect information in quantitative bottom-up proteomics appear particularly attractive [1315]. Increased, for example, by a factor of 2–10, data acquisition rate should allow interrogation of more peptides at the required structural level and provide improved qualitative and quantitative protein analyses. Development of novel concepts for ion conditioning and manipulation in ICR cells based on improved understanding of ion motion fundamentals has thus become a target of recent innovations in the field [6, 16].

Since the inception of FT-ICR MS by Marshall and Comisarow, the design of ICR cells has been primarily based on the use of wide aperture detection electrodes and sinusoidal ion signals (that is, generation of transient signals whose spectral energy is concentrated primarily at the fundamental ion frequency) [12, 17]. Since the last decade or so, the most common ICR cell configuration has been a cylindrical cell with a pair of 90°-wide excitation electrodes and a pair of similar 90°-wide detection electrodes [4, 18]. The detected image charge ion signal is averaged during ion motion along the detection electrode surface, requiring low frequency (as determined from Nyquist criteria) sampling. The drawbacks of this configuration include a certain limitation on the acquisition rate, substantial perturbation of excitation radio frequency RF field by detection electrodes, and a low degree of design flexibility.

Marshall and co-workers have recently demonstrated by simulations that 120°-wide excitation electrodes are optimal for coherent ion dipolar excitation and elimination of the third harmonic [19]. However, implementation of such electrodes for ion excitation would reduce the efficiency of ion detection using the currently employed ICR cell design concept. A feasible, although technically challenging, solution to this problem is the use of external switching electronic circuitry allowing for ion excitation and detection using the same pair of 120°-wide electrodes.

Quadrupolar modes of ion excitation, including the quadrature excitation, are other attractive approaches for improving the coherence of excited ion motion [2023]. ICR cell operation in these modes also suffers from the same low flexibility of the current ICR cell design, when wide aperture detection electrodes compete with the excitation electrodes for space. Improved coherence of ion motion in ICR cells has been efficiently addressed via harmonization of the trapping electric fields by a number of approaches, including those of Tolmachev et al. [24, 25], Gross et al. [26], Bruce et al. [27, 28], and Nikolaev and co-workers [29, 30]. The obtained improvements in FT-ICR MS performance are quite dramatic. Specifically, the transient lifetime has been substantially, about 10- to 100-fold, increased. The dynamically harmonized ICR cell, following the original idea of Boldin and Nikolaev [29], is particularly impressive, with demonstrated resolution of up to 40 million on a peptide (single peak) and a routinely achieved isotopic fine structure resolution on peptide ions even in moderate, 7 T, magnetic fields [31]. The main drawback of the harmonized cells is that they do not improve the throughput (or the acquisition rate) of FT-ICR MS.

Signal detection at multiples of ion (reduced) cyclotron frequency, which is a feasible approach for increasing the throughput [3236], is also limited by the design of those cells: they also comprise relatively wide aperture excitation and detection electrodes, resulting in a restricted factor of the generated multiples (e.g., ~22° electrodes and the fourth multiple for a 16-electrode ICR cell). A number of attempts have been made to reduce the aperture of detection electrodes. For example, Nikolaev and co-workers suggested an ICR cell with four detection electrodes made of conducting wires [32]. However, the experimental implementation of the suggested ICR cell was deferred. One of the possible reasons for the failure was the positioning of these electrodes directly below the excitation electrodes in the areas with strong excite RF field, leading to excitation field distortion and pre-amplifier saturation prior to ion detection.

Here, we describe a concept, design, and implementation of an ICR cell with narrow aperture (flat) detection electrodes (NADEL) addressing the limitations of FT-ICR MS design and capabilities specified above. The analytical characteristics of thus introduced NADEL ICR cells are presented and their particular advantages are outlined.

Experimental

Sample Preparation

Substance P, MRFA, Ultramark, and ubiquitin were purchased from Sigma-Aldrich (Buchs, Switzerland). ESI calibration mixture pre-formulated at low concentration was purchased from Agilent (Basel, Switzerland). Serum albumin (bovine) was purchased from Fisher Scientific (Schwerte, Germany). Polyphosphoric acid was obtained from Fluka (Buchs, Switzerland). C-terminally amidated peptides P1, P2, and P3: P1 – EESTR (monoisotopic mass 619.29255 Da), P2 – PMMSR (619.29342 Da), and P3 – HDGHR (619.29389 Da) were synthesized in-house with solid state peptide synthesis. All peptide and protein samples were used without further purification. LC-MS grade acetonitrile and water were obtained from Fluka (Buchs, Switzerland). Formic acid was obtained from Merck (Zug, Switzerland). Peptide and protein solutions were prepared in 1:1 (v/v) water/acetonitrile solvent mixtures containing 0.1% (v/v) of formic acid. The final sample concentration for all samples was ~1 μM, except ~10 μM for albumin. The polyphosphoric calibration mixture was prepared by adding 1% (v/v) of polyphosphoric acid to 1 mL of water. Maltene fraction of Venezuelan crude oil was dissolved in dichloromethane, followed by 1:23 dilution in acetonitrile with the addition of 2% of formic acid (by volume), as described previously [9].

NADEL ICR Cell Design and Mass Spectrometry

The design of the NADEL ICR cell employed here was developed based on the open-ended cylindrical ICR cell (Ultra Cell; Thermo Scientific, Bremen, Germany). The commercially available Ultra cell contains two pairs, inner and outer, of segmented trapping ring electrodes and excitation grids, placed over the entire length of the cell, in addition to the conventional 90° electrodes (a pair of which are used for ion detection and another pair is grounded) [4]. Ultra cell’s inner diameter is of ~56 mm and length of middle section (detection) electrodes is of ~70 mm. In NADEL ICR cell, the standard 90° detection electrodes were substituted with narrow aperture (flat) detection electrodes with a thickness of 1 mm. Importantly, the new flat detection electrodes were inserted radially closer to the center of the cell, Figure 1 and Figure S1 (Supporting Information). In the NADEL ICR cell evaluated here, the detection electrodes were positioned at 17 mm radially away from the cell center.

Figure 1
figure 1

Schematic representation of the narrow aperture detection electrodes (NADEL) ICR cell: (top panel) a 3D view and (bottom panel) its unrolled surface. The NADEL cell contains two pairs, outer and inner, of trapping ring electrodes, excitation grids, conventional 90° excitation, and narrow aperture (flat) detection electrodes with a thickness of 2 mm and radius of curvature R = 125 mm

The NADEL as well as a standard Ultra (90° detection electrodes) ICR cells were mounted, one at a time, onto a hybrid linear ion trap Fourier transform ion cyclotron resonance mass spectrometer (LTQ FT-ICR MS; Thermo Scientific, Bremen, Germany) equipped with a 10 T actively shielded superconducting magnet (Oxford Nanoscience, Oxon, UK), described elsewhere. Instrument control was performed by the standard electronics and software (Xcalibur, Thermo Scientific). The ions were formed with robotic chip-based nano-electrospray ionization source (Triversa Nanomate; Advion Biosciences, Ithaca, NY, USA) and transported to the linear ion trap (LTQ) through an inlet set of multipole ion guides. A predetermined number of charges to be transferred into the ICR cell using another set of multipole ion guides were controlled by the automatic gain control (AGC) function of the LTQ. For petroleomics measurements the AGC function was disabled and accumulation time in the LTQ was set to 50 ms, while the potentials of the inlet transfer system were further optimized manually. For petroleomics analysis, these experimental parameters provided the best correlation of ion distribution envelope (the total width and position of the maximum of ion distribution) between ion detection in LTQ and in the ICR cell. Vacuum conditions in FT-ICR MS with NADEL or Ultra ICR cells were identical (2E-10 Torr).

Ion trapping in both ICR cells was achieved with two pairs of trapping ring electrodes. All sections of outer trapping rings were grounded throughout the complete experimental sequence. A potential of 3 V was applied to the grid-free sections of inner trapping rings to confine transferred ions inside the ICR cell during ion trapping and relaxation events. The same level of trapping potential was kept during ion excitation event, whereas it was user-defined in the range of 0–10 V during the ion detection event. Approximately 4.6-fold higher potential was applied to the sections of inner rings covered with grid. The shape of the detection electrodes in NADEL ICR cell was designed to form trapping potential distribution closer to quadratic. For the NADEL ICR cell reported here the distance r 0 from the detector electrode edge to the ICR cell’s axis z = 0 was selected to be 17 mm, which corresponds to 60% radius of the original Ultra ICR cell. The shape of the detection electrodes was calculated using the following formula:

$$ r(z)={r}_0+R\left(1-\sqrt{1-\frac{z^2}{R^2}}\right) $$

where z is an axial coordinate and R is the electrode’s radius of curvature. For the NADEL ICR cell reported here, with 70 mm long detection electrodes, we selected a compromise value of R, 125 mm, as it provides a reasonable ratio of radial and axial trapping electric fields along the z-axis (as stated above, trapping potential distribution should be close to quadratic). SIMION calculations confirmed that the selected shape of the detection electrodes provides improved trapping field compared to rectangular narrow aperture detection electrodes, Figures S1 and S2 (Supporting Information). The parameters r 0 and R can be optimized both numerically (e.g., using SIMION, and experimentally in further studies.

Ion excitation with NADEL ICR cell was technically performed in the same way as with Ultra ICR cell [4]. Coherent motion of ions was excited by standard dipolar frequency-sweep excitation in the range of 72–2150 kHz applied during 10 ms to the excitation grids. The peak-to-peak amplitude of the excitation RF voltage was varied from 0 to 100 V. The particular advantage of the NADEL ICR cell compared with other ICR cell designs, including the Ultra cell, is in the improved excitation field distribution, Figure S3, top left (Supporting Information). Due to the positioning of the detection electrodes in the central plane, symmetrically between the top and bottom excitation electrodes, narrow aperture detection electrodes disturb the RF excite field significantly less than the standard 90° detection electrodes, as employed in Ultra cell. Therefore, ion excitation in the NADEL cell is more efficient than for a regular ICR cell, especially when exciting to a larger radius. Overall, the maximum excitation radius is limited by the location of the detection electrodes. However, to the best of our knowledge, ion excitation to large, >0.7 radius, orbits is not employed in modern FT-ICR MS [22, 24, 37]. Moreover, according to the estimation by the authors, most of the ICR cells function with ion excitation to 0.3–0.4 cell radius because of the increasing of the inhomogeneity of magnetic field at higher radii, although the exact data could not be found in publications. Therefore, the detection electrodes were placed at the corresponding positions, allowing ion excitation to 0.6 cell radius, and thus not reducing the effective ion volume. Note, a finite initial magnetron radius would correspondingly reduce the maximum achievable excitation radius. Figure S3, bottom (Supporting Information) demonstrates the difference in the ion detection conditions between NADEL and Ultra cells. Compared with the standard 90° detection electrodes, the narrow aperture detection electrodes increase the non-linear character of ion detection as a function of a post-excitation radius. Importantly, for NADEL cell sensitivity level, detection electrode capacitance of the installed NADEL cell (measured together with the connecting wires and feedthrough) was reduced four-fold compared with the Ultra ICR cell, as expected.

Ion detection was performed using the commercial ion detection configuration of the Ultra ICR cell. Briefly, standard dipolar differential detection was employed to acquire time domain signals (transients). The transients of variable length in the range of 96 ms–25 s were recorded in MIDAS format at 1–5 MHz sampling frequency (a regime with sinusoidal signals) using the advanced software interface of the built-in data acquisition system (Thermo Scientific). A given number (e.g., 10–100) of transients were averaged to obtain the final transients, which were Hann-apodized and zero-filled once before fast Fourier transformation (FFT) to yield Fourier spectra, which were further transformed to mass spectra using standard data analysis software (Xcalibur; Thermo Scientific) and the pyFTMS framework developed in-house.

Finally, the offset DC potentials in the range from –100 to +100 mV were applied to the NADEL and Ultra ICR cell detection and excitation electrodes independently relative to each other to vary the position of the ion cloud prior to ion excitation and during ion detection events. The influence of offset potentials on ion motion was monitored with diverse quality attributes of mass spectra (e.g., peak shape and magnetron sidebands [38]). Overall, it was possible to optimize the offset potentials to effectively decrease the magnetron sideband intensities below 1% relative to the corresponding reduced cyclotron frequency peak for each experiment at the optimal post-excitation ion cyclotron radius.

Results and Discussion

NADEL ICR Cell Initial Performance Tests: MS and MS/MS Data Acquisition

A typical broadband mass spectrum obtained with NADEL FT-ICR MS demonstrates efficient simultaneous confinement and detection of ions during 3 s in a mass range of m/z 300–3300, Figure 2. Similarly, Figure S4 (Supporting Information) shows a broadband mass spectrum of a calibration mixture, revealing simultaneous detection of ions spanning a broad range of m/z 150–1900 with acquisition period of 0.384 s. Direct comparison of NADEL FT-ICR mass spectra with those obtained with a linear ion trap (LTQ) on the same instrument demonstrates that for all constituent signals the relative ion abundances are similar (data not shown). The decrease of ion abundance was observed only for m/z less than 300, attributable to expected m/z-dependent time-of-flight discrimination of ions during their transfer between LTQ and ICR cell [39]. The insets in Figure 2 and Figure S4 (Supporting Information) demonstrate the typical peak shape for monoisotopic peaks of selected compounds.

Figure 2
figure 2

Broadband mass spectrum of a polyphosphoric acid sample, acquired with 10 T FT-ICR MS equipped with NADEL ICR cell. The mass spectrum was obtained from a sum of 30 transients recorded in the mass range of 300–3300 m/z with the acquisition period 3.072 s and number of charges 1E6 (AGC)

In-cell tandem mass spectrometry (MS/MS) experiments with NADEL ICR cell demonstrate efficiency comparable to Ultra cell. Typical electron capture dissociation (ECD) [40], infrared multiphoton dissociation (IRMPD) [41], and electron induced dissociation (EID) [42, 43] FT-ICR mass spectra are shown in Figure S5 (Supporting Information). MS/MS broadband mass spectra contain analyte peaks with a magnitude spreading over 3–4 decimal orders. Particularly, the spectral dynamic range (ion abundance ratio of the highest to the lowest peaks within a single mass spectrum) [44] of 4.4 orders was achieved for EID MS/MS measurements. Thus, NADEL ICR cell provides dipolar excitation and detection of ions, either injected from an external ion source or produced inside of the cell, in the wide mass range without the discrimination of different m/z ions and at a dynamic range comparable to the state-of-the-art FT-ICR MS performance. Note, the employed ion excitation for this mode of operation (generation of sinusoidal ion signals) aims to excite ions to moderate radii, when ions do not yet approach the detection electrodes. The resulting transient components for these regimes are primarily sinusoidal, with a very low (<5%) high-order harmonic content.

NADEL ICR Cell Performance: Sensitivity

To estimate fundamental characteristics of the NADEL ICR cell, in particular the value of magnetron frequency, ω , and optimal post-excitation ion radius (ion radius at optimized excitation parameters), the reduced cyclotron frequency ω + and the signal-to-noise ratio (SNR) of peaks corresponding to ω + frequency, its harmonics and interharmonics ( + , nω + ± - , where n, k = 1, 2, 3) were measured as functions of excitation energy for isolated singly protonated ions of MRFA (m/z 524.3) at different trapping potentials applied during ion detection event, Figure 3. As expected, the SNR of the peak corresponding to the reduced cyclotron frequency ω + linearly increases with excitation amplitude (cyclotron radius). However, a significant decrease of the SNR was observed for excitation amplitudes higher than approximately 40 V. On the other hand, the magnitude ratio of the harmonic 3ω + to the ω + peak increased with the excitation amplitude, Figure 3 top left, meaning that the post-excitation cyclotron radius continues to increase with the corresponding growth of the high-order harmonics. The SNR decline for the fundamental frequency, ω + , is presumably due to the reduced harmonicity of the trapping electric potential and homogeneity of the excitation electric field for high excitation amplitudes, which cause phase decoherence of ion clouds, leading to a decrease of ion signal amplitude. Additionally, even harmonics of ω + were observed in the mass spectra. The presence of even harmonics testifies for a non-zero magnetron radius and an initial radial shift of an ion cloud from the axis of the ICR cell. Notably, even harmonics’ magnitudes were reduced to a level less than 2% relative to the ω + peak via fine tuning of the offset DC potentials applied to the excitation and detection electrodes independently for each trapping potential.

Figure 3
figure 3

Characterization of the NADEL ICR cell in 10 T FT-ICR MS. (Top left): relative intensity of the harmonic 3ω + (w.r.t. the peak ω + ) and signal-to-noise ratio (SNR), as absolute intensity w.r.t. the mean of baseline noise, of the peak ω + , as a function of the excitation amplitude. (Top right) dependencies of the SNR of the peak ω + on the defined number of charges (logarithmic scales) were obtained under identical instrumental parameters for both the standard (open circles) and NADEL (solid circles) ICR cells. (Bottom left) dependencies of the frequency (triangles) and the SNR value (circles) of the peak ω + on the user-defined value of number of charges (AGC) in the wide range 2E1–2E6. (Bottom right) dependence of the frequency of the peak ω + on the excitation amplitude for different trapping potentials. All experimental data were acquired for isolated singly protonated MRFA peptide (m/z 524.3) with the acquisition period of 3.072 s and 10 ms of excitation duration

The sensitivity levels of the NADEL ICR cell and the standard Ultra cell with 90° detection electrodes were compared for an isolated singly protonated MRFA peptide (m/z 524.3) under identical optimal instrumental parameters for both cells. As expected, a signal-to-noise ratio (SNR) of peak corresponding to the reduced cyclotron frequency ω + linearly increased with the number of charges (ion abundance) for both NADEL and Ultra ICR cells, Figure 3 top right. The SNR values reached a plateau at around 1E5–2E5 number of charges in the isolation mode for both cells as well. This is likely due to the limitations of accumulation in LTQ and ion transport from LTQ to ICR cell for ions with the same nominal mass-to-charge ratio. Importantly, Figure 3 top right shows the increase of the peak SNR almost two times for the NADEL cell compared with the standard Ultra cell. Therefore, the experimental results demonstrate higher signal voltage developed across the detection circuit of the NADEL ICR cell compared with the corresponding signal in the standard 90° electrode Ultra ICR cell, despite the significantly reduced surface area of detection electrodes of the NADEL cell. This result can be explained by the substantially lower capacitance of the flat detect electrodes and shorter distance between the orbit of excited ions and detection electrodes. It is thus not surprising that only a few charges, approximately 10 ions (AGC = 30) per cell load, of triply protonated substance P (m/z 449.9) ions were detected with acquisition period of 384 ms and five transients averaging, Figure S6 (Supporting Information). This result correlates well with the known sensitivity performance level of 50 charges of the state-of-the-art FT-ICR MS with standard ICR cells.

NADEL ICR Cell Performance: Mass Accuracy

For low trapping potentials (<1 V) , the reduced cyclotron frequency ω + peak remained within the frequency range of 1 ppm over the entire range of excitation amplitudes where ion signal can be detected, except the region of low excitation amplitudes, Figure 3, bottom right. However, frequency reduced with an increase in the trapping potential for high excitation amplitudes due to reduced harmonicity of the trapping electric potential. Thus, the electric field distribution of the NADEL ICR cell is uniform in the wide range of excitation amplitudes (post-excitation radii) at low trapping potentials. Note, the DC offset potentials were adjusted for each trapping potential independently when constructing these dependencies. Furthermore, at low excitation amplitudes (small post-excitation radii) for all trapping potentials a space charge of significant ion abundance (AGC = 5E4) is the reason for the observed frequency shift, as reported earlier [45]. A scaling factor of the magnetron frequency was experimentally determined to be 2.7 Hz/V, Figure 3, bottom right. Experimental results allowed us to define optimal trapping potentials 0.8–1.0 V and excitation amplitude of 30–40 V with a duration of 10 ms. Additionally, the stability of the reduced cyclotron frequency measured for isolated singly protonated ions of MRFA 524.3 m/z with trapping potential of 1.0 V and excitation amplitude of 30 V indicates the absence of space charge effect in the wide range of number of charges, AGC = 1E2–1E5, for the NADEL ICR cell, Figure 3, bottom left.

Mass calibration of the FT-ICR mass spectrometer equipped with the NADEL ICR cell was performed via the standard two parametric calibration function using nine monoisotopic ions of ESI-L low concentration tuning mixture (Agilent). The mass calibration function was determined using two AGC values (5E5 and 2E6) in the mass range of m/z 300–1400 (four calibrants) with acquisition period of 0.568 s and trapping potentials of 1 and 2 V, as well as in the range of m/z 300–3000 (nine calibrants) with acquisition period of 1.336 s and trapping potential of 1 V resulting in six calibration parameter sets, Table 1.

Table 1 RMS Mass Accuracy of Agilent ESI-L Low Concentration Tuning Mixture (Internal Calibration) and Polyphosphoric Acid (External Calibration, Based on the ESI-L Mixture) Obtained with Two Values of a Number of Charges (AGC Equal to 5E5 or 2E6)

Mass calibration was evaluated by considering the mass measurement accuracies achieved in the same mass ranges for analyte peaks of polyphosphoric acid Hn+2PnO3n+1, n = 3–30, (monoisotopic peaks of protonated and ammonium adducts, as well as the 18O isotopologue of the protonated species), Table 1. Only analyte peaks with SNR higher than 3 were taken into account. The mass error did not exceed 750 ppb level with the root-mean-square (RMS) value of 330 ppb in the mass range 300–1400 m/z for both AGC settings and both trapping potentials. These results are comparable to those obtained using conventional ICR cells with external mass calibration [24, 46, 47]. Mass error increased for wider mass range, especially for higher AGC value; the corresponding dependencies are given in Figure S7 (Supporting information).

To rationalize these results, the effect of space charge on mass accuracy was investigated using polyphosphoric acid with internal calibration. Specifically, mass calibration was performed for three mass ranges and varied AGC values (3E5–6E6) with acquisition period of 1336 ms, Table 2. Figure S7 (Supporting Information) demonstrates the SNR values of analyte peaks linearly increasing with an increase of the AGC value in the current range. The RMS mass accuracy of 60 ppb was achieved for eight monoisotopic peaks of polyphosphoric acid in the range of 300–1000 m/z and AGC = 7E5. Furthermore, the RMS mass error remained within 200 ppb for the number of charges up to AGC = 2E6 for low mass calibration range up to 1000 m/z, Table 2 and Figure S7 (Supporting Information). These results are comparable to those obtained with conventional and harmonized ICR cells for internal calibration [24, 46, 48]. The mass error became more pronounced for higher AGC values, 3E6, and more as can be expected due to an increased space-charge field [45, 49], especially for the highest m/z peaks due to higher analyte density per nominal mass in that m/z region, Figure 2 and Figure S7 (Supporting Information).

Table 2 RMS Mass Accuracy of Polyphosphoric Acid (Internal Calibration) as a Function of Number of Charges

Thus, mass accuracy performance with NADEL ICR cell is comparable to those obtained with conventional and harmonized ICR cells. This conclusion also follows from the analysis of a crude oil fraction using NADEL ICR cell, vide infra. Application of absorption mode FT signal processing allows improving RMS mass accuracy values two times, as expected [5052]. However, the observed increase in mass inaccuracy when a very wide m/z range is analyzed, Figure S7 (Supporting Information) and Table 1, needs further investigation. Particularly, influence of the modified trapping field should be considered. Indeed, Figure S1 (Supporting Information) shows that the detection electrode configuration described here introduces an azimuthal component of a trapping electric field, previously avoided in FT-ICR MS cell designs. On the other hand, it is known that the offset DC potentials typically employed to reduce harmonic content in FT-ICR mass spectra and improve peak shapes may introduce a similar azimuthal component to the trapping field distribution. Therefore, the exact character, destructive or constructive, of the modified trapping field is to be evaluated. Finally, further optimization of narrow aperture detection electrodes may reduce the introduced modulation of the trapping electric field.

NADEL ICR Cell Performance: Resolving Power

The resolving power performance of the NADEL ICR cell was evaluated in the analysis of peptide and protein samples with maximum detection period of 24.576 s for the employed hardware configuration. Acquisition of longer transients was not possible because of the technical limitations imposed by the on-board LTQ FT computer memory, whereas the lifetime of excited and coherent ion motion apparently surpasses this temporal limitation; for example see Figure S8 (Supporting Information). Notably, the achieved ion detection period of ~25 s is the record for LTQ FTMS instruments reports.

Figure 4 shows a mass spectrum of three isobaric singly protonated peptides P1, P2, and P3 (nominal m/z of monoisotopic peaks is 620) with mass differences of 0.9 and 0.5 mDa between the pairs of peptides. Notably, mass of 0.5 mDa is close to the mass of an electron. Expanded segments of the mass spectrum show baseline-resolved monoisotopic and 13C peaks for the three peptides of interest with the resolving power reaching 3′300′000 for a single scan. To avoid the peak coalescence effect, a relatively low number of ions were employed (AGC setting of 3E4). The particular importance of the result reported in Figure 4 is in the long time of ion cloud coherence after excitation, which allows recording of 25 s and (potentially) longer transients. The reported separation of this peptide triplet can be compared with the notable example of a peptide doublet separation (mass difference close to the mass of an electron) by Marshall and co-workers on a 9.4 T FT-ICR MS [53].

Figure 4
figure 4

Isobaric peptide mixture measurements with NADEL cell in 10 T FT-ICR MS. The resolving power 3′300′000 was achieved for a single scan with acquisition period of 24.576 s and number of charges (AGC) of 30′000 for isolated ions of singly protonated peptides P1, P2, and P3

Another notable example of resolving power performance provided by NADEL cell for peptide and protein analysis is the acquisition of the isotopic fine structure distribution of triply protonated substance P (449.9 m/z) and major isotopologues of 11+ charge state of bovine ubiquitin (779.0 m/z), Figure 5. The mass spectrum of substance P was obtained for 10 summed transient signals; the summed signal is shown in Figure S8 (Supporting Information). Mass spectrum of bovine ubiquitin was obtained from 20 summed single transients. Isotopic fine structure components of substance P and bovine ubiquitin are resolved with resolution of 4′400′000 and 2′900′000, respectively.

Figure 5
figure 5

Peptide (substance P, top panel) and protein (ubiquitin, bottom panel) isotopic fine structure measurements with NADEL cell in 10 T FT-ICR MS. Mass spectrum of isolated triply protonated substance P 449.9 m/z was obtained from a sum of 10 transients of 24.576 s each. The corresponding transient is shown in Figure S8 (Supporting Information). Mass spectrum of bovine ubiquitin was obtained from sum of 20 transients of 24.576 s acquired for isolated 11+ ions

To evaluate NADEL ICR cell performance for heavy protein analysis, protein mass measurements were performed for a bovine serum albumin (BSA) with direct infusion ESI. Broadband mass spectrum of BSA obtained by averaging of 50 single scans of 1.536 s shows the charge state distribution in the range from 35+ up to 72+, Figure S9 (Supporting Information). Furthermore, the resolving power of 320′000 was achieved for an isolated BSA48+ charge state with a single scan and acquisition period of 6.144 s, Figure S10 (Supporting Information).

Finally, a petroleomic sample was chosen to evaluate the NADEL ICR cell performance for complex mixture analysis. A positive ESI broadband mass spectrum of a fraction of South American crude oil was acquired with an acquisition period of 3.072 s, Figure 6. The insets demonstrate mass scale expansions at 451 m/z nominal mass, revealing typical representative spectral compositions for a petroleum sample with resolving power of 450'000. Mass assignments are given in Table 3. Additionally, mass accuracy values for the 2715 monoisotopic peaks identified in the broadband mass spectrum of a crude oil fraction in the mass range of 300–700 m/z were measured as a function of m/z and SNR by using standard two-parametric equation and internal calibration based on 29 monoisotopic peaks of a prominent homologous series, Figure S11 (Supporting information). Overall, 5142 peaks were detected with the intensity higher than 2% relative to the highest magnitude peak in mass spectrum. RMS mass accuracy and mean values were 236 ppb and 0.7 ppb, respectively. As expected, mass error increases with decreasing SNR and increasing m/z value.

Figure 6
figure 6

ESI broadband mass spectrum of a crude oil fraction obtained with NADEL ICR cell in 10 T FT-ICR MS from the averaging of 600 transients of 3.072 s each in the positive-ion mode. Insets show expanded regions around (top) 451 m/z and (bottom) 445–453 m/z for the displayed broadband mass spectrum. Only peaks above a certain threshold (dotted red line) were considered. Annotated peaks are listed in Table 3

Table 3 Peak Assignments and Mass Errors in the Segments Around 451 m/z of the Positive ESI Broadband Mass Spectrum of a Crude Oil fraction. Corresponding Inset is Shown in Figure 6

To conclude, the combined benefits of trapping and excitation field configuration in the NADEL cell provide comparable performance to standard ICR cells in routine analysis in regard to resolving power performance. The obtained characteristics approach those of the most sophisticated ICR cells with complex approaches to electric field harmonization, either statically or dynamically, for the investigated 10 T magnetic field [24, 29, 46]. The described NADEL ICR cell has been employed to perform routine mass analysis in our laboratory and demonstrated an ability to achieve the required analytical objectives of molecular structural analysis in a number of applications, for example in the analysis of supramolecular complexes [54].

NADEL ICR Cell Concept Advantages

The particular benefits of the NADEL ICR cell concept for current and envisioned FT-ICR MS configurations include: (1) enhanced dipolar ion excitation conditions because of improved configuration of excite electric field. The latter is achieved by reduced excitation field disturbance by radially inserted detection electrodes compared with the standard 90° degrees wide detection electrodes, Figure S3 (Supporting Information). Further improvement of excite electric field distribution can be presumably obtained by utilization of wide aperture excitation electrodes with an optimized angle, which can be now varied up to 180°. Interestingly, flat (wide aperture) excitation electrodes (capacitor configuration) may be considered to provide an optimum excitation field configuration; (2) simultaneous detection of a number of transients or detection at frequency multiples when several NADEL pairs are employed, for instance leading to an optimized design of the recently implemented quadruple frequency multiple detection ICR cell [36]; (3) enabling implementation of efficient quadrature ion excitation and quadrupolar ion detection schemes when two pairs of 90° wide excite electrodes and two NADEL pairs are employed without the need for a high-frequency switch between excitation and detection modes. Quadrupolar ion detection should further benefit high mass accuracy and high dynamic range demanding applications by providing measurements of trapping potential-insensitive unperturbed ion cyclotron frequency [23, 55]; (4) creation of special ion motion conditions in ICR cell even with dipolar ion excitation leading to ion detection at unperturbed cyclotron frequency [56]; (5) facilitating and improving the efficiency of fluorescence-based ion spectroscopy due to substantially increased optical access to and from the cell [5759]; and, last but not least, (6) enabling high throughput high-resolution mass spectrometry by accelerating ions to high post-excitation radii when, significantly, an order of magnitude higher resolving power can be obtained in the same ion detection period. The latter regime requires the development of the matching signal processing methods, capable of efficient analysis of periodic non-sinusoidal transients with high order harmonics [60].

Conclusions

We designed and implemented an ICR cell with a pair of narrow aperture detection electrodes (NADEL) for improving the analytical capabilities and increasing the mass analyzer design flexibility of FT-ICR MS. The obtained results demonstrate that the narrow aperture (flat) detection electrodes provide comparable or superior performance relative to the standard (commercial) ICR cells in modern FT-ICR MS. As proven here, the ability to replace the currently employed wide aperture (azimuthally large) detection electrodes in all ICR cell designs with the narrow aperture ones opens new avenues for ICR cell development and FT-ICR MS capabilities. A number of advantages and envisioned benefits validate the importance of the described ICR cell development. Rationalizing the underlying ion physics phenomena leading to these capabilities (e.g., unperturbed ion cyclotron frequency detection with dipolar ion detection) requires in-depth theoretical analysis of the described NADEL ICR cell and its further modifications.