Abstract
Introduction
The aim of this investigation was to evaluate the introduction of the VA-LCP anterior clavicle plate in the treatment of clavicle fractures.
Material and methods
From March 2011 to March 2013, 42 clavicle fractures were treated; 40 were middle-third and 2 lateral-third, and 13/42 (31 %) patients were treated due to painful nonunion. Patient age ranged from 16 to 81 years.
Results
Complications were screw placement through the AC-joint, one superficial wound infection and one neuropraxia of the nervus radialis with dropping hand. We had some difficulties prebending both lateral to low and lateral to high but without clinical consequences. In all cases, the fracture healed with full functionality. After 1 year, 4 patients underwent a removal of the hardware.
Conclusion
The VA-LCP anterior plate showed good reliability and sufficient stability with both middle-third, lateral and nonunion fractures of the clavicle.
Similar content being viewed by others
References
McKee RC, Whelan DB, Schemitsch EH, McKee MD (2012) Operative versus nonoperative care of displaced midshaft clavicular fractures: a meta-analysis of randomized clinical trials. J Bone Joint Surg Am 94:675–684
Hill JM, McGuire MH, Crosby LA (1997) Closed treatment of displaced middle-third fractures of the clavicle gives poor results. J Bone Joint Surg Br 79:537–539
Matis N (1999) Effects of clavicle shortening after clavicle fracture. Unfallchirurg 275:314–315
Houwert RM, Wijdicks FJG, Bisschop CS et al (2012) Plate fixation versus intramedullary fixation for displaced midshaft clavicle fractures: a systematic review. Int Orthop 36:579–585
Wijdicks FJG, Houwert RM, Dijkgraaf MGW, et al. A randomized controlled trial comparing plate fixation versus intramedullary fixation for dislocated midshaft clavicle fractures: short term results. [thesis]
Canadian Orthopaedic Trauma Society (2007) Nonoperative treatment compared with plate fixation of displaced midshaft clavicular fractures. A multicenter, randomized clinical trial. J Bone Joint Surg Am 89:1–10
Taylor PR, Day RE, Nicholls RL et al (2011) The comminuted midshaft clavicle fracture: a biomechanical evaluation of plating methods. Clin Biomech (Bristol, Avon) 26:491–496
Sinha A, Edwin J, Sreeharsha B et al (2011) A radiological study to define safe zones for drilling during plating of clavicle fractures. J Bone Joint Surg Br 93:1247–1252
Formaini N, Taylor BC, Backes J, Bramwell TJ (2013) Superior versus anteroinferior plating of clavicle fractures. Orthopedics 36(7):898–904
Houwert RM, Wijdicks FJ, Steins Bisschop C et al (2012) Plate fixation versus intramedullary fixation for displaced mid-shaft clavicle fractures: a systematic review. Int Orthop 36:579–585
Lenza M, Belloti JC, Gomes Dos Santos JB, et al (2009) Surgical interventions for treating acute fractures or non-union of the middle third of the clavicle. Cochrane Database Syst Rev (4):CD007428
Virtanen KJ, Remes V, Pajarinen J et al (2012) Sling compared with plate osteosynthesis for treatment of displaced midshaft clavicular fractures. J Bone Joint Surg Am 94:1546–1553
Virtanen KJ, Malmivaara AO, Remes VM, Paavola MP (2012) Operative and nonoperative treatment of clavicle fractures in adults. Acta Orthop 83:65–73
Conflict of interest
None.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
van Olden, G.D.J. VA-LCP anterior clavicle plate: The anatomically precontoured fixation system with angular stability for clavicle shaft. Musculoskelet Surg 98, 217–223 (2014). https://doi.org/10.1007/s12306-013-0302-z
Received:
Accepted:
Published:
Issue Date: