Severe Metal-induced Osteolysis Many Years After Unipolar Hip Endoprosthesis

  • Case Report
  • Published:
Clinical Orthopaedics and Related Research®

Abstract

Background

Modularity of the femoral head-neck junction provides increased intraoperative flexibility to the surgeon. Complications of this modularity include damage to the trunnion, with subsequent bone and/or soft tissue loss from adverse reactions to metal debris.

Case Description

We describe two cases of severe metal-induced osteolysis and soft tissue damage requiring revision 10 and 13 years following implantation of a unipolar endoprosthesis. Damage to the trunnion resulted in severe acetabular and trochanteric osteolysis and soft tissue loss requiring complex revision surgery.

Literature Review

Several reports have shown the trunnion, the head-neck interface, and the neck-stem couple as the causes of this early failure secondary to metal ion release from mechanical fretting corrosion or from crevice corrosion at these modular interfaces. These reports have been in association with a total hip prosthesis rather than a unipolar endoprosthesis. Revision of a unipolar endoprosthesis is most commonly attributable to stem loosening or acetabular erosion from the large femoral head articulating on the host acetabular cartilage and not owing to failure of the trunnion.

Purposes and Clinical Relevance

Trunnion damage resulting in a severe reaction to metal debris with acetabular osteolysis, erosion of the greater trochanter, and loss of the abductor mechanism can occur years after implantation of a cementless unipolar endoprosthesis. This raises questions regarding long-term safety of the modular interface of a contemporary cementless stem and a large-diameter unipolar head. We recommend long-term followup of patients with a unipolar endoprosthesis as early recognition and treatment are required to avoid a potentially complex revision.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1A–E
Fig. 2A–C

References

  1. Andersson G, Nielsen JM. Results after arthroplasty of the hip with Moore’s prosthesis. Acta Orthop Scand. 1972;43:397–410.

    Article  PubMed  CAS  Google Scholar 

  2. Atwood SA, Patten EW, Bozic KJ, Pruitt LA, Ries MD. Corrosion-induced fracture of a double-modular hip prosthesis: a case report. J Bone Joint Surg Am. 2010;92:1522–1525.

    Article  PubMed  Google Scholar 

  3. Bolland BJ, Culliford DJ, Langton DJ, Millington JP, Arden NK, Latham JM. High failure rates with a large-diameter hybrid metal-on-metal total hip replacement: clinical, radiological and retrieval analysis. J Bone Joint Surg Br. 2011;93:608–615.

    PubMed  CAS  Google Scholar 

  4. Botti TP, Gent J, Martell JM, Manning DW. Trunion fracture of a fully porous-coated femoral stem: case report. J Arthroplasty. 2005;20:943–945.

    Article  PubMed  Google Scholar 

  5. Brown SA, Flemming CA, Kawalec JS, Placko HE, Vassaux C, Merritt K, Payer JH, Kraay MJ. Fretting corrosion accelerates crevice corrosion of modular hip tapers. J Appl Biomater. 1995;6:19–26.

    Article  PubMed  CAS  Google Scholar 

  6. Collier JP, Surprenant VA, Jensen RE, Mayor MB. Corrosion at the interface of cobalt-alloy heads on titanium-alloy stems. Clin Orthop Relat Res. 1991;271:305–312.

    PubMed  Google Scholar 

  7. Collier JP, Surprenant VA, Jensen RE, Mayor MB, Surprenant HP. Corrosion between the components of modular femoral hip prostheses. J Bone Joint Surg Br. 1992;74:511–517.

    PubMed  CAS  Google Scholar 

  8. Cook SD, Barrack RL, Clemow AJ. Corrosion and wear at the modular interface of uncemented femoral stems. J Bone Joint Surg Br. 1994;76:68–72.

    PubMed  CAS  Google Scholar 

  9. Cooper HJ, Della Valle CJ, Berger RA, Tetreault M, Paprosky WG, Sporer SM, Jacobs JJ. Corrosion at the head-neck taper as a cause for adverse local tissue reactions after total hip arthroplasty. J Bone Joint Surg Am. 2012;94:1655–1661.

    Article  PubMed  Google Scholar 

  10. Cossey AJ, Goodwin MI. Failure of Austin Moore hemiarthroplasty: total hip replacement as a treatment strategy. Injury. 2002;33:19–21.

    Article  PubMed  CAS  Google Scholar 

  11. Garbuz DS, Tanzer M, Greidanus NV, Masri BA, Duncan CP. The John Charnley Award: Metal-on-metal hip resurfacing versus large-diameter head metal-on-metal total hip arthroplasty: a randomized clinical trail. Clin Orthop Relat Res. 2010;468:318–325.

    Article  PubMed  Google Scholar 

  12. Gilbert JL, Buckley CA, Jacobs JJ. In vivo corrosion of modular hip prosthesis components in mixed and similar metal combinations: the effect of crevice, stress, motion, and alloy coupling. J Biomed Mater Res. 1993;27:1533–1544.

    Article  PubMed  CAS  Google Scholar 

  13. Gilbert JL, Buckley CA, Jacobs JJ, Bertin KC, Zernich MR. Intergranular corrosion-fatigue failure of cobalt-alloy femoral stems: a failure analysis of two impants. J Bone Joint Surg Am. 1994;76:110–115.

    PubMed  CAS  Google Scholar 

  14. Gilbert JL, Mehta M, Pinder B. Fretting crevice corrosion of stainless steel stem-CoCr femoral head connections: comparisons of materials, initial moistrure, and offset length. J Biomed Mater Res B Appl Biomater. 2009;88:162–173.

    PubMed  Google Scholar 

  15. Goldberg JR, Gilbert JL. In vitro corrosion testing of modular hip tapers. J Biomed Mater Res B Appl Biomater. 2003;64:78–93.

    Article  PubMed  Google Scholar 

  16. Goldberg JR, Gilbert JL, Jacobs JJ, Bauer TW, Paprosky W, Leurgans S. A multicenter retrieval study of the taper interfaces of modular hip prostheses. Clin Orthop Relat Res. 2002;401:149–161.

    Article  PubMed  Google Scholar 

  17. Kaltsas DS, Klugman DJ. Acetabular erosion: a comparison between the Austin Moore and Monk hard top prostheses. Injury. 1986;17:230–236.

    Article  PubMed  CAS  Google Scholar 

  18. Kop AM, Swarts E. Corrosion of a hip stem with a modular neck taper junction: a retrieval study of 16 cases. J Arthroplasty. 2009;24:1019–1023.

    Article  PubMed  Google Scholar 

  19. Lam LO, Stoffel K, Kop A, Swarts E. Catastrophic failure of 4 cobalt-alloy Omnifit hip arthroplasty femoral components. Acta Orthop. 2008;79:18–21.

    Article  PubMed  Google Scholar 

  20. Langton DJ, Jameson SS, Joyce TJ, Gandhi JN, Sidaginamale R, Mereddy P, Lord J, Nargol AV. Accelerating failure rate of the ASR total hip replacement. J Bone Joint Surg Br. 2011;93:1011–1016.

    PubMed  CAS  Google Scholar 

  21. Langton DJ, Jameson SS, Joyce TJ, Hallab NJ, Natu S, Nargol AV. Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: a consequence of excess wear. J Bone Joint Surg Br. 2010;92:38–46.

    PubMed  CAS  Google Scholar 

  22. Lewis G. Geometric element analysis of fretting in a model of a modular femoral component of a hip implant. Biomed Mater Eng. 2004;14:43–51.

    PubMed  Google Scholar 

  23. Lieberman JR, Rimnac CM, Garvin KL, Klein RW, Salvati EA. An analysis of the head-neck taper interface in retrieved hip prostheses. Clin Orthop Relat Res. 1994;300:162–167.

    PubMed  Google Scholar 

  24. Phillips TW. Thompson hemiarthroplasty and acetabular erosion. J Bone Joint Surg Am. 1989;71:913–917.

    PubMed  CAS  Google Scholar 

  25. Shareef N, Levine D. Effect of manufacturing tolerances on the micromotion at the Morse taper interface in modular hip implants using the finite element technique. Biomaterials. 1996;17:623–630.

    Article  PubMed  CAS  Google Scholar 

  26. Soreide O, Lillestol J, Aiho A, Hvidsten K. Acetabular protrusion following endoprostetic hip surgery: a multifactorial study. Acta Orthop Scand. 1980;51:943–948.

    Article  PubMed  CAS  Google Scholar 

  27. Warwick D, Hubble M, Sarris I, Strange J. Revision of failed hemiarthroplasty for fractures at the hip. Int Orthop. 1998;22:165–168.

    Article  PubMed  CAS  Google Scholar 

  28. Wright G, Sporer S, Urban R, Jacobs J. Fracture of a modular femoral neck after total hip arthroplasty: a case report. J Bone Joint Surg Am. 2010;92:1518–1521.

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Michael Tanzer MD.

Additional information

Each author certifies that he or she, or a member of his or her immediate family, has no funding or commercial associations (eg, consultancies, stock ownership, equity interest, patent/licensing arrangements, etc) that might pose a conflict of interest in connection with the submitted article.

All ICMJE Conflict of Interest Forms for authors and Clinical Orthopaedics and Related Research editors and board members are on file with the publication and can be viewed on request.

Each author certifies that his or her institution approved the reporting of this case report, that all investigations were conducted in conformity with ethical principles of research, and that informed consent for participation in the study was obtained.

About this article

Cite this article

Mann, M.A., Tanzer, D. & Tanzer, M. Severe Metal-induced Osteolysis Many Years After Unipolar Hip Endoprosthesis. Clin Orthop Relat Res 471, 2078–2082 (2013). https://doi.org/10.1007/s11999-012-2780-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11999-012-2780-y

Keywords

Navigation