3D kinematics of mobile-bearing total knee arthroplasty using X-ray fluoroscopy

  • Original Article
  • Published:
International Journal of Computer Assisted Radiology and Surgery Aims and scope Submit manuscript

Abstract

Purpose

  Total knee arthroplasty (TKA) 3D kinematic analysis requires 2D/3D image registration of X-ray fluoroscopic images and a computer-aided design (CAD) model of the knee implant. However, these techniques cannot provide information on the radiolucent polyethylene insert, since the insert silhouette does not appear clearly in X-ray images. Therefore, it is difficult to obtain the 3D kinematics of the polyethylene insert, particularly the mobile-bearing insert. A technique for 3D kinematic analysis of a mobile-bearing insert used in TKA was developed using X-ray fluoroscopy. The method was tested and a clinical application was evaluated.

Methods

Tantalum beads and a CAD model of the mobile-bearing TKA insert are used for 3D pose estimation of the mobile-bearing insert used in TKA using X-ray fluoroscopy. The insert model was created using four identical tantalum beads precisely located at known positions in a polyethylene insert using a specially designed insertion device. Finally, the 3D pose of the insert model was estimated using a feature-based 2D/3D registration technique, using the silhouette of beads in fluoroscopic images and the corresponding CAD insert model. In vitro testing for the repeatability of the positioning of the tantalum beads and computer simulations for 3D pose estimation of the mobile-bearing insert were performed.

Experimental results

The pose estimation accuracy achieved was sufficient for analyzing mobile-bearing TKA kinematics (RMS error: within 1.0 mm and 1.0\(^{\circ }\), except for medial–lateral translation). In a clinical application, nine patients with mobile-bearing TKA were investigated and analyzed with respect to a deep knee bending motion.

Conclusions

A 3D kinematic analysis technique was developed that enables accurate quantitative evaluation of mobile-bearing TKA kinematics. This method may be useful for improving implant design and optimizing TKA surgical techniques.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7
Fig. 8
Fig. 9

Similar content being viewed by others

References

  1. Banks SA, Hodge WA (1996) Accurate measurement of three-dimensional knee replacement kinematics using single-plane fluoroscopy. IEEE Trans Biomed Eng 43:638–649

  2. Zuffi S, Leardini A, Catani F, Fantozzi S, Cappello A (1999) A model-based method for the reconstruction of total knee replacement kinematics. IEEE Trans Med Imaging 18:981–991

    Article  CAS  PubMed  Google Scholar 

  3. Mahfouz MR, Hoff WA, Komistek RD, Dennis DA (2003) A robust method for registration of three-dimensional knee implant models to two-dimensional fluoroscopy images. IEEE Trans Med Imaging 22:1561–1574

    Article  PubMed  Google Scholar 

  4. Yamazaki T, Watanabe T, Nakajima Y, Sugamoto K, Tomita T, Yoshikawa H, Tamura S (2004) Improvement of depth position in 2-D/3-D registration of knee implants using single-plane fluoroscopy. IEEE Trans Med Imaging 23:602–612

    Article  PubMed  Google Scholar 

  5. Kobashi S, Tomosada T, Shibanuma N, Yamaguchi M, Muratsu H, Kondo K, Yoshiya S, Hata Y, Kurosaka M (2005) Fuzzy image matching for pose recognition of occluded knee implants using fluoroscopy images. J Adv Comput Intell Intell Inf 9:181–195

    Google Scholar 

  6. Bingham J, Li G (2006) An optimized image matching method for determining in-vivo TKA kinematics with a dual–orthogonal fluoroscopic imaging system. J Biomech Eng 128:588–595

    Article  PubMed  Google Scholar 

  7. Hirokawa S, Abrar Hossain M (2008) A 3D kinematic estimation of knee prosthesis using X-ray projection images: clinical assessment of the improved algorithm for fluoroscopy images. Med Biol Eng Comput 46:1253–1262

    Article  PubMed  Google Scholar 

  8. Markelj P, Tomaževič D, Likar B, Pernuš F (2012) A review of 3D/2D registration methods for image-guided interventions. Med Image Anal 16:642–661

    Article  CAS  PubMed  Google Scholar 

  9. Fantozzi S, Leardini A, Banks SA, Marcacci M, Giannini S, Cantani F (2004) Dynamic in-vivo tibio-femoral and bearing motions in mobile bearing knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 12:144–151

    Article  CAS  PubMed  Google Scholar 

  10. Komistek RD, Dennis DA, Mahfouz MR, Walker S, Outten J (2004) In vivo polyethylene bearing mobility is maintained in posterior stabilized total knee arthroplasty. Clin Orthop Rel Res 428:207–213

    Article  Google Scholar 

  11. Chouteau J, Lerat J-L, Testa R, Moyen B, Fessy M-H, Banks SA (2009) Mobile-bearing insert translational and rotational kinematics in a PCL-retaining total knee arthroplasty. Orthop Traumatol Surg Res 95:254–259

    Article  CAS  PubMed  Google Scholar 

  12. Wolterbeek N, Garling EH, Mertens B, Valstar ER, Nelissen RGHH (2009) Mobile bearing knee kinematics change over time. A fluoroscopic study in rheumatoid arthritis patients. Clin Biomech 24:441–445

    Article  CAS  Google Scholar 

  13. Haneishi H, Yagihashi Y, Miyake Y (1995) A new method for distortion correction of electronic endoscope images. IEEE Trans Med Imaging 14:548–555

    Article  CAS  PubMed  Google Scholar 

  14. Weng J, Cohen P, Herniou M (1992) Camera calibration with distortion models and accuracy evaluation. IEEE Trans Pattern Anal Machine Intell 14:965–980

    Article  Google Scholar 

  15. Canny J (1986) A computational approach to edge detection. IEEE Trans Pattern Anal Mach Intell 8:679–698

    Article  CAS  PubMed  Google Scholar 

  16. Lavallee S, Szeliski R (1995) Recovering the position and orientation of free-form objects from image contours using 3D distance maps. IEEE Trans Pattern Anal Mach Intell 17:378–390

    Article  Google Scholar 

  17. Luenberger DG (1984) Linear and nonlinear programming, Facsimile edition. Addison-Wesley, Massachusetts

    Google Scholar 

  18. Insall JN, Hood RW, Flawn LB, Sullivan DJ (1983) The total condylar knee prosthesis in gonarthrosis. A five to nine-year follow-up of the first one hundred consecutive replacements. J Bone Joint Surg Am 65:619–628

    CAS  PubMed  Google Scholar 

  19. Grood ES, Suntay WJ (1983) A joint coordinate system for the clinical description of three-dimensional motions: application to the knee. J Biomech Eng 105:136–144

    Article  CAS  PubMed  Google Scholar 

  20. Dennis DA, Mahfouz MR, Komistek RD, Hoff W (2005) In vivo determination of normal and anterior cruciate ligament–deficient knee kinematics. J Biomech 38:241–253

    Article  PubMed  Google Scholar 

Download references

Acknowledgments

A part of this study was supported by the Global COE Program “in silico medicine” at Osaka University, and by Grants-in-Aid for Hybrid Organ Project and Scientific Research for Young Scientists (B) (23700206) from The Ministry of Education, Culture, Sports, Science and Technology, Japan.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Takaharu Yamazaki.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Yamazaki, T., Futai, K., Tomita, T. et al. 3D kinematics of mobile-bearing total knee arthroplasty using X-ray fluoroscopy. Int J CARS 10, 487–495 (2015). https://doi.org/10.1007/s11548-014-1093-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11548-014-1093-x

Keywords

Navigation