Combined oral appliance and positive airway pressure therapy for obstructive sleep apnea: a pilot study

  • Original Article
  • Published:
Sleep and Breathing Aims and scope Submit manuscript

Abstract

Background

The high efficacy of continuous positive airway pressure (CPAP) in treating obstructive sleep apnea (OSA) is limited by poor compliance often related to pressure intolerance. Mandibular advancement devices (MADs) are proven alternative therapy although not universally effective. A combination of nasal CPAP and MAD may provide another option for CPAP-intolerant patients with incomplete response to MAD.

Methods

Ten patients with residual apnea/hypopnea events on MAD who were intolerant to CPAP were recruited prospectively from the sleep clinic. After a washout period of 1 week off MAD, subjects were asked to use an auto-CPAP unit along with their prescribed MAD for three consecutive nights. Oxygen desaturations were obtained from overnight oximetry. Efficacy of the combination therapy was evaluated by the Epworth Sleepiness Scale and Smartcard data recordings.

Results

The combination of MAD and nasal CPAP was well tolerated by all participants. Compared to CPAP alone, the optimal CPAP pressure required to eliminate all obstructive events on the combination therapy was reduced from 9.4 ± 2.3 to 7.3 ± 1.4 cm H2O (p = 0.001). The residual apnea hypopnea index on the MAD decreased from 11.2 ± 3.9 to 3.4 ± 1.5 on the combination therapy (p < 0.001). The number of oxygen desaturations was also less with the combination therapy than with MAD (p < 0.001). Both the MAD and the combination therapy were effective in reducing daytime sleepiness from 12.7 ± 2.1 at baseline to 9.7 ± 3.1 (p = 0.04) and 7.5 ± 4.1 (p = 0.007), respectively.

Conclusions

Combination therapy of MAD and nasal CPAP is effective in normalizing respiratory disturbances of sleep apnea in selected OSA patients who are intolerant to CPAP.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

References

  1. Gay P, Weaver T, Loube D, Iber C (2006) Evaluation of positive airway pressure treatment for sleep related breathing disorders in adults. Sleep 29:381–401

    PubMed  Google Scholar 

  2. Noseda A, Jann E, Hoffman G, Linkowski P, Kerkhofs M (2000) Compliance with nasal continuous positive airway pressure assessed with a pressure monitor: pattern of use and influence of sleep habits. Respir Med 94:76–81

    Article  PubMed  CAS  Google Scholar 

  3. Rosenberg R, Doghramji P (2009) Optimal treatment of obstructive sleep apnea and excessive sleepiness. Adv Ther 26:295–312

    Article  PubMed  Google Scholar 

  4. Hudgel DW, Fung C (2000) A long-term randomized, cross-over comparison of auto-titrating and standard nasal continuous airway pressure. Sleep 23:645–648

    PubMed  CAS  Google Scholar 

  5. Ferguson KA, Ono T, Lowe AA, al-Majed S, Love LL, Fleetham JA (1997) A short-term controlled trial of an adjustable oral appliance for the treatment of mild to moderate obstructive sleep apnoea. Thorax 52:362–368

    Article  PubMed  CAS  Google Scholar 

  6. Mayer G, Meier-Ewert K (1995) Cephalometric predictors for orthopaedic mandibular advancement in obstructive sleep apnea. Eur J Orthod 17:35–43

    PubMed  CAS  Google Scholar 

  7. Eveloff SE, Rosenberg CL, Carlisle CC, Millman RP (1994) Efficacy of a Herbst mandibular advancement device in obstructive sleep apnea. Am J Respir Crit Care Med 149:905–909

    PubMed  CAS  Google Scholar 

  8. Randerath WJ, Heise M, Hinz R, Ruehle K-H (2002) An individually adjustable oral appliance vs continuous positive airway pressure in mild-to-moderate obstructive sleep apnea syndrome. Chest 122:569–575

    Article  PubMed  Google Scholar 

  9. Ferguson KA, Cartwright R, Rogers R, Schmidt-Nowara W (2006) Oral appliances for snoring and obstructive sleep apnea: a review. Sleep 29:244–262

    PubMed  Google Scholar 

  10. Denbar MA (2002) A case study involving the combination treatment of an oral appliance and auto-titrating CPAP unit. Sleep Breath 6:125–128

    Article  PubMed  Google Scholar 

  11. Kushida C, Morgenthaler T, Littner M, Alessi CA, Bailey D, Coleman J Jr, Friedman L, Hirshkowitz M, Kapen S, Kramer M, Lee-Chiong T, Owens J, Pancer JP, American Academy of Sleep (2006) Practice parameters for the treatment of snoring and obstructive sleep apnea with oral appliances: an update for 2005. Sleep 29:240–243

    PubMed  Google Scholar 

  12. Rechtschaffen A, Kales A (1968) A manual of standardized terminology, technique, and scoring system for sleep stages of human subjects. BIS/BRI, UCLA, Los Angeles

    Google Scholar 

  13. Johns MW (1991) A new method for measuring daytime sleepiness: the Epworth sleepiness scale. Sleep 14:540–545

    PubMed  CAS  Google Scholar 

  14. Zeng B, Ng AT, Liu B, Qian J, Darendeliler MA, Cistulli PA (2006) Effect of mandibular advancement splint (MAS) on upper airway anatomy. Proc Am Thorac Soc 3:A868, Abstract

    Google Scholar 

  15. Heinzer RC, Stanchina ML, Malhotra A, Fogel RB, Patel SR, Jordan AS, Schory K, White DP (2005) Lung volume and continuous positive airway pressure requirements in obstructive sleep apnea. Am J Respir Crit Care Med 172:114–117

    Article  PubMed  Google Scholar 

  16. Ayas NT, Patel SR, Malhotra A, Schulzer M, Malhotra M, Jung D, Fleetham J, White DP (2004) Auto-titrating versus standard continuous positive airway pressure for the treatment of obstructive sleep apnea: results of a meta-analysis. Sleep 27:249–253

    PubMed  Google Scholar 

  17. Senn O, Brack T, Matthews F, Russi E, Bloch K (2003) Randomized short-term trial of two auto-CPAP devices versus fixed continuous positive airway pressure for the treatment of sleep apnea. Am J Respir Crit Care Med 168:1506–1511

    Article  PubMed  Google Scholar 

  18. Desai H, Patel A, Grant BJ, Mador MJ (2009) Accuracy of autotitrating CPAP to estimate the residual apnea–hypopnea index in patients with obstructive sleep apnea on treatment with autotitrating CPAP. Sleep Breath 13:383–390

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

We would like to thank Joe Saltino, LPN for his assistance in arranging overnight oximetry to all participants and downloading the Smartcard data recordings. Dr. Morohunfolu E. Akinnusi is supported by a grant from the American Sleep Foundation.

Conflict of interest

None

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Ali A. El-Solh.

Rights and permissions

Reprints and permissions

About this article

Cite this article

El-Solh, A.A., Moitheennazima, B., Akinnusi, M.E. et al. Combined oral appliance and positive airway pressure therapy for obstructive sleep apnea: a pilot study. Sleep Breath 15, 203–208 (2011). https://doi.org/10.1007/s11325-010-0437-1

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11325-010-0437-1

Keywords

Navigation