Reliability of the volatile agent consumption display in the Draeger Primus™ anesthesia machine

  • Original Research
  • Published:
Journal of Clinical Monitoring and Computing Aims and scope Submit manuscript

Abstract

Knowledge of the consumed amount of volatile anesthetic (VA) expressed in liquid agent is necessary to enable agent sparing dosing measures and for billing purposes. The widespread Draeger Primus™ anesthesia machine displays in its logbook the amount of consumed VA at the end of each anesthesia, but the reliability of this parameter is yet unknown. The objective was to evaluate the precision and reliability of the inbuilt VA consumption display in Draeger Primus™ anesthesia machines as compared with the gold standard of weighing the vaporizer before and after anesthesia. In this prospective laboratory investigation we compared the VA consumption displayed by the Draeger Primus™ anesthesia machine with measured vaporizer weight differences before and after 10 sevoflurane and 10 desflurane anesthesias. We assessed the average difference and spread of values between the predicted (displayed) and measured (control) values for VA consumption. The displayed sevoflurane consumption overestimated the measured values by 4.3 ± 5.4 ml (7.6 %). The displayed desflurane consumption underestimated the measured values by −3.5 ± 6.3 ml (6.2 %). Nine from 10 sevoflurane pairs of values and all desflurane pairs of values were within ±1.96 SD. The displayed VA consumption calculations for sevoflurane and desflurane in the Draeger Primus™ are sufficiently reliable to estimate the pharmacoeconomic impact of VA delivery during inhalational anesthesia.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2

Similar content being viewed by others

References

  1. Rapeport D, Harper S, Flisberg P. Pharmacoeconomic viewpoints on volatile anaesthesia—“much ado about nothing?”. Anaesth Intensive Care. 2005;33:144–5.

    CAS  PubMed  Google Scholar 

  2. Weinberg L, Story D, Nam J, McNicol L. Pharmacoeconomics of volatile inhalational anaesthetic agents: an 11-year retrospective analysis. Anaesth Intensive Care. 2010;38:849–54.

    CAS  PubMed  Google Scholar 

  3. Hinz J, Rieske N, Schwien B, et al. Cost analysis of two anaesthetic machines: “Primus(R)” and “Zeus(R)”. BMC Res Notes. 2012;5:3.

    Article  PubMed  PubMed Central  Google Scholar 

  4. Honemann C, Hagemann O, Doll D. Inhalational anaesthesia with low fresh gas flow. Indian J Anaesth. 2013;57:345–50.

    Article  PubMed  PubMed Central  Google Scholar 

  5. Coetzee JF, Stewart LJ. Fresh gas flow is not the only determinant of volatile agent consumption: a multi-centre study of low-flow anaesthesia. Br J Anaesth. 2001;88:46–55.

    Article  Google Scholar 

  6. Cotter SM, Petros AJ, Dore CJ, Barber ND, White DC. Low-flow anaesthesia. Practice, cost implications and acceptability. Anaesthesia. 1991;46:1009–12.

    Article  CAS  PubMed  Google Scholar 

  7. Eger EI 2nd. Characteristics of anesthetic agents used for induction and maintenance of general anesthesia. Am J Health Syst Pharm. 2004;61(Suppl 4):S3–10.

    CAS  PubMed  Google Scholar 

  8. Loscar M, Conzen P. Volatile anesthetics. Anaesthesist. 2004;53:183–98.

    Article  CAS  PubMed  Google Scholar 

  9. Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res. 1999;8:135–60.

    Article  CAS  PubMed  Google Scholar 

  10. Glen J, Marshall S. Gas leak related to Draeger Primus anaesthetic machine. Anaesthesia. 2010;65:750.

    Article  CAS  PubMed  Google Scholar 

  11. Lockwood GG, White DC. Measuring the costs of inhaled anaesthetics. Br J Anaesth. 2001;87:559–63.

    Article  CAS  PubMed  Google Scholar 

Download references

Conflict of interest

The authors declare that there is no financial interest involved.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Peter Biro.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Biro, P., Kneschke, O. & Theusinger, O.M. Reliability of the volatile agent consumption display in the Draeger Primus™ anesthesia machine. J Clin Monit Comput 29, 601–604 (2015). https://doi.org/10.1007/s10877-014-9639-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10877-014-9639-6

Keywords

Navigation