Skip to main content

Advertisement

Log in

Ray tracing software for intraocular lens power calculation after corneal excimer laser surgery

  • Clinical Investigation
  • Published:
Japanese Journal of Ophthalmology Aims and scope Submit manuscript

Abstract

Purpose

To evaluate the accuracy of intraocular lens (IOL) power calculations using ray tracing software in eyes after myopic laser in situ keratomileusis (LASIK).

Methods

Twenty-four eyes of 17 cataract patients who underwent phacoemulsification and IOL implantation after myopic LASIK were analyzed retrospectively. The IOL power calculation was performed using OKULIX ray tracing software. The axial length was measured using the IOLMaster and keratometry data using TMS2N. The accuracy of the IOL power calculation using OKULIX was compared with those using the Camellin–Calossi, Shammas-PL, Haigis-L formulas and the double-K SRK/T formula using 43.5 diopters (D) for the Kpre.

Results

The mean values of the arithmetic and absolute prediction errors were 0.63 ± 0.85 and 0.80 ± 0.68 D, respectively. The arithmetic prediction error by OKULIX was a significant hyperopic shift of the distribution of the postoperative refractive errors compared to the Camellin–Calossi, Shammas-PL and Haigis-L formulas (P < 0.05), and the absolute prediction error showed no significant difference with other formulas. The prediction errors using OKULIX were within ±0.5 D in 10 eyes (41.7 %) and within ±1.0 D in 18 eyes (75.0 %). The percentages of eyes within ±1.0 D using OKULIX were comparable to those obtained using the Camellin–Calossi, the Shammas-PL formulas and the double-K SRK/T formula using 43.5 D for the Kpre, and significantly (P < 0.05) higher than that obtained using the Haigis-L formula.

Conclusions

IOL power calculations using OKULIX provided predictable outcomes in eyes that had undergone a previous myopic LASIK.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2

Similar content being viewed by others

References

  1. Hoffer KJ. Intraocular lens power calculation for eyes after refractive keratotomy. J Refract Surg. 1995;11:490–3.

    PubMed  CAS  Google Scholar 

  2. Feiz V, Mannis MJ, Garcia-Ferrer F, Kandavel G, Darlington JK, Kim E, et al. Intraocular lens power calculation after laser in situ keratomileusis for myopia and hyperopia: a standardized approach. Cornea. 2001;20:792–7.

    Article  PubMed  CAS  Google Scholar 

  3. Hoffer KJ. Calculating intraocular lens power after refractive corneal surgery. Arch Ophthalmol. 2002;120:500–1.

    Article  PubMed  Google Scholar 

  4. Aramberri J. Intraocular lens power calculation after corneal refractive surgery: double-K method. J Cataract Refract Surg. 2003;29:2063–8.

    Article  PubMed  Google Scholar 

  5. Koch DD, Wang L. Calculating IOL power in eyes that have had refractive surgery. J Cataract Refract Surg. 2003;29:2039–42.

    Article  PubMed  Google Scholar 

  6. Shammas HJ, Shammas MC, Garabet A, Kim JH, Shammas A, LaBree L. Correcting the corneal power measurements for intraocular lens power calculations after myopic laser in situ keratomileusis. Am J Ophthalmol. 2003;136:426–32.

    Article  PubMed  Google Scholar 

  7. Wang L, Booth MA, Koch DD. Comparison of intraocular lens power calculation methods in eyes that have undergone LASIK. Ophthalmology. 2004;111:1825–31.

    Article  PubMed  Google Scholar 

  8. Latkany RA, Chokshi AR, Speaker MG, Abramson J, Soloway BD, Yu G. Intraocular lens calculations after refractive surgery. J Cataract Refract Surg. 2005;31:562–70.

    Article  PubMed  Google Scholar 

  9. Rosa N, Capasso L, Lanza M, Iaccarino G, Romano A. Reliability of a new correcting factor in calculating intraocular lens power after refractive corneal surgery. J Cataract Refract Surg. 2005;31:1020–4.

    Article  PubMed  Google Scholar 

  10. Camellin M, Calossi A. A new formula for intraocular lens power calculation after refractive corneal surgery. J Refract Surg. 2006;22:187–99.

    PubMed  Google Scholar 

  11. Masket S, Masket SE. Simple regression formula for intraocular lens power adjustment in eyes requiring cataract surgery after excimer laser photoablation. J Cataract Refract Surg. 2006;32:430–4.

    Article  PubMed  Google Scholar 

  12. Borasio E, Stevens J, Smith GT. Estimation of true corneal power after keratorefractive surgery in eyes requiring cataract surgery: BESSt formula. J Cataract Refract Surg. 2006;32:2004–14.

    Article  PubMed  Google Scholar 

  13. Savini G, Barboni P, Zanini M. Intraocular lens power calculation after myopic refractive surgery: theoretical comparison of different methods. Ophthalmology. 2006;113:1271–82.

    Article  PubMed  Google Scholar 

  14. Shammas HJ, Shammas MC. No-history method of intraocular lens power calculation for cataract surgery after myopic laser in situ keratomileusis. J Cataract Refract Surg. 2007;33:31–6.

    Article  PubMed  Google Scholar 

  15. Haigis W. Intraocular lens calculation after refractive surgery for myopia: Haigis-L formula. J Cataract Refract Surg. 2008;34:1658–63.

    Article  PubMed  Google Scholar 

  16. Geggel HS. Pachymetric ratio no-history method for intraocular lens power adjustment after excimer laser refractive surgery. Ophthalmology. 2009;116:1057–66.

    Article  PubMed  Google Scholar 

  17. Kim SW, Kim EK, Cho BJ, Kim SW, Song KY, Kim TI. Use of the Pentacam true net corneal power for intraocular lens calculation in eyes after refractive corneal surgery. J Refract Surg. 2009;25:285–9.

    PubMed  Google Scholar 

  18. Hoffer KJ. Intraocular lens power calculation after previous laser refractive surgery. J Cataract Refract Surg. 2009;35:759–65.

    Article  PubMed  Google Scholar 

  19. Saiki M, Negishi K, Kato N, Arai H, Toda I, Torii H, et al. A new central-peripheral corneal curvature method for intraocular lens power calculation after excimer laser refractive surgery. Acta Ophthalmol. 2013;91:e133–9.

    Article  PubMed  Google Scholar 

  20. Saiki M, Negishi K, Kato N, Ogino R, Arai H, Toda I, et al. Modified double-K method for intraocular lens power calculation after excimer laser corneal refractive surgery. J Cataract Refract Surg. 2013;39:556–62.

    Article  PubMed  Google Scholar 

  21. Preussner PR, Wahl J, Lahdo H, Dick B, Findl O. Ray tracing for intraocular lens calculation. J Cataract Refract Surg. 2002;28:1412–9.

    Article  PubMed  Google Scholar 

  22. Preussner PR, Hoffmann P, Petermeier K. Comparison between ray-tracing and IOL calculation formulae of the 3rd generation. Klin Monbl Augenheilkd. 2009;226:83–9 [in German].

    Article  PubMed  Google Scholar 

  23. Minami K, Kataoka Y, Matsunaga J, Ohtani S, Honbou M, Miyata K. Ray-tracing intraocular lens power calculation using anterior segment optical coherence tomography measurements. J Cataract Refract Surg. 2012;38:1758–63.

    Article  PubMed  Google Scholar 

  24. Rabsilber TM, Reuland AJ, Holzer MP, Auffarth GU. Intraocular lens power calculation using ray tracing following excimer laser surgery. Eye (Lond). 2007;21:697–701.

    Article  CAS  Google Scholar 

  25. Camellin M. Proposed formula for the dioptric power evaluation of the posterior corneal surface. Refract Corneal Surg. 1990;6:261–4.

    PubMed  CAS  Google Scholar 

  26. Canovas C, Abenza S, Alcon E, Villegas E, Marin J, Artal P. Effect of corneal aberrations on intraocular lens power calculations. J Cataract Refract Surg. 2012;38:1325–32.

    Article  PubMed  Google Scholar 

  27. Preussner PR, Wahl J, Weitzel D. Topography-based intraocular lens power selection. J Cataract Refract Surg. 2005;31:525–33.

    Article  PubMed  Google Scholar 

  28. Wang L, Shirayama M, Ma XJ, Kohnen T, Koch DD. Optimizing intraocular lens power calculations in eyes with axial lengths above 25.0 mm. J Cataract Refract Surg. 2011;37:2018–27.

    Article  PubMed  Google Scholar 

Download references

Conflicts of interest

M. Saiki, None; K. Negishi, None; N. Kato, None; H. Torii, None; M. Dogru, None; K. Tsubota, None.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Kazuno Negishi.

About this article

Cite this article

Saiki, M., Negishi, K., Kato, N. et al. Ray tracing software for intraocular lens power calculation after corneal excimer laser surgery. Jpn J Ophthalmol 58, 276–281 (2014). https://doi.org/10.1007/s10384-014-0304-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10384-014-0304-x

Keywords

Navigation