Abstract
Purpose
To study possible risk factors in cases in which part of the device becomes exposed outside the conjunctiva after Ahmed glaucoma valve (AGV) implantation.
Methods
Eleven eyes in which one part of the AGV was exposed outside the conjunctivae were compared with 44 random eyes that underwent the same operation by the same surgeon during the same time period but without any complications. Preoperative diagnosis, age, sex, implant location, number of previous ocular surgeries, and both associated ocular diseases and systemic diseases were taken into account in both the patient group and the control group to determine all possible causes.
Results
Age, implant location, type of glaucoma, diabetes mellitus, and hypertension had no relation to implant exposure. However, the patient group had more ocular surgeries before AGV implantation than the control group (P = 0.006).
Conclusion
We found that only the number of previous ocular surgeries was a possible risk factor for implant exposure. More effort should be made to decrease implant exposure outside the conjunctiva, and research should be performed to determine the proper treatment for implant exposure.
Similar content being viewed by others
References
Rollett M, Moreau M. Traitement dele hypopyon par le drainage capillaire de chamber anterieure. Rev Gen Ophthalmol 1906;35:481.
El-Sayyad F, el-Maghraby A, Helal M, Amayem A. The use of releasable sutures in Molteno glaucoma procedures to reduce postoperative hypotony. Ophthalmic Surg 1991;35:82–84.
Kook MS, Jeon SK. Clinical results of Ahmed glaucoma implantation in refractory glaucoma. J Korean Ophthalmol Soc 1996;37:1893–1901.
Kim JM, Lee DH. Clinical experience with the Ahmed glaucoma valve. J Korean Ophthalmol Soc 1999;40:1332–1338.
Hu CH, Lee YG, Hong YJ. Ahmed glaucoma valve implant in refractory glaucoma. 1997;38:259–266.
Molteno ACB. New implant for drainage in glaucoma. Clinical trial. Br J Ophthalmol 1969;53:606–615.
Schocket SS, Lakhanpal V, Richards RD. Anterior chamber tube shunt to an encircling band in the treatment of neovascular glaucoma. Ophthalmology 1982;89:1188–1194.
Lloyd MA, Baerveldt G, Heuer DK, et al. Initial clinical experience with the Baerveldt implant in complicated glaucoma. Ophthalmology 1994;101:640–650.
Ahn BH, Kim CS, Kim YB. Use of e-PTFE membrane for glaucoma drainage surgery. J Korean Ophthalmol Soc 1990;31:603–614.
Coleman AL, Hill R, Wilson MR, et al. Initial clinical experience with the Ahmed glaucoma valve implant. Am J Ophthalmol 1995;120:23–31.
Krupin T, Podos SM, Becker B. Valve implants in filtering surgery. Am J Ophthalmol 1976;81:232–235.
Joseph NH, Sherwood MB, Trantas G, et al. A one-piece drainage system for glaucoma surgery. Trans Ophthalmol Soc UK 1986;105:657–664.
White TC. Clinical results of glaucoma surgery using the White glaucoma pump shunt. Ann Ophthalmol 1992;24:365–373.
Prata JA, Mermoud A, LaBree L, Minckler DS. In vitro and in vivo flow characteristics of glaucoma drainage implants. Ophthalmology 1995;102:894–904.
Ayyala RS, Zurakowski D, Smith JA, et al. A clinical study of the Ahmed glaucoma valve implant in advanced glaucoma. Ophthalmology 1998;105:1968–1976.
Heuer DK, Lloyd MA, Abrams DA, et al. Which is better? One or two? A randomized clinical trial of single-plate versus double-plate Molteno implantation for glaucomas in aphakia and pseudophakia. Ophthalmology 1992;99:1512–1519.
Heuer DK, Budenz D, Coleman A. Aqueous shunt tube erosion. J Glaucoma 2001;10:493–496.
Huang MC, Netland PA, Coleman AL, et al. Intermediate-term clinical experience with the Ahmed glaucoma valve implant. Am J Ophthalmol 1999;127:27–33.
Coleman AL, Mondino BJ, Wilson MR, et al. Clinical experience with the Ahmed glaucoma valve implant in eyes with prior or concurrent penetrating keratoplasties. Am J Ophthalmol 1997;123:54–61.
Heuer DK, Budenz D, Coleman A. Aqueous shunt tube erosion. J Glaucoma 2001;10:493–496.
Takahashi A, Okisaka S, Akeo K, Taketani P. Aging changes in conjunctiva visualized by fluorescein angiography and histopathology. Nippon Ganka Gakkai Zasshi 1996;100:296–301.
Ainsworth G, Rotchford A, Dua HS, King AJ. A novel use of amniotic membrane in the management of tube exposure following glaucoma tube shunt surgery. Br J Ophthalmol 2006;90:417–419.
Siegner SW, Netland PA, Urban RC Jr, et al. Clinical experience with the Baerveldt glaucoma drainage implant. Ophthalmology 1995;102:1298–1307.
Rai P, Lauande-Pimentel R, Barton K. Amniotic membrane as an adjunct to donor sclera in the repair of exposed glaucoma drainage devices. Am J Ophthalmol 2005;140:1148–1152.
Chen H, Zhang SX, Liu L, et al. Intermediate-term and long-term clinical evaluation of the Ahmed glaucoma valve implantation. Zhonghua Yan Ke Za Zhi 2005;41:796–802.
Author information
Authors and Affiliations
Corresponding author
About this article
Cite this article
Byun, Y.S., Lee, N.Y. & Park, C.K. Risk factors of implant exposure outside the conjunctiva after Ahmed glaucoma valve implantation. Jpn J Ophthalmol 53, 114–119 (2009). https://doi.org/10.1007/s10384-008-0630-y
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10384-008-0630-y