Introduction

Fixed-bearing (FB) total knee arthroplasty is a successful operation with well-documented excellent long-term results [1, 2]. However, because of changing demographics in patients who require total knee arthroplasty, i.e., shifting to a younger population with higher functional demands, newer designs have been developed to achieve greater survivorship and clinical outcomes.

Mobile bearings (MBs) were designed to reduce the peak loading stress and backside wear observed as a cause of aseptic loosening in FB designs [3]. To achieve this, they have a more conforming superior articular surface which, in theory, reduces the contact stresses [1, 48]. The introduction of a second bearing interface results in a decoupling of the complex multidirectional motions which occur in FB designs producing unidirectional motion at the two bearing interfaces of the MB implant which, in theory, should reduce polyethylene wear. There have been concerns raised, however, about the risk of MB dislocation and some reports of early backside wear in some clinical studies [1, 7].

In addition, MB designs have the potential to correct any rotational malalignment of the femoral and tibial components by allowing the patellar tendon to self-align throughout a range of motion, enhancing both patellofemoral and tibiofemoral mechanics [9]. Little attention has been given to the potential effects that this decoupling may have on the patellofemoral joint portion of the articulation. In theory, the self-alignment seen with MB designs should reduce the rate of lateral release of the patella, which is usually performed to optimise patellofemoral mechanics.This study reports on the lateral release rates for P.F.C.® Sigma MB posterior-stabilised TKR compared with its FB equivalent.

Materials and methods

Three hundred and fifty-two patients were randomised to receive a PFC Sigma© total knee replacement (TKR) with either FB or MB implants. The randomisation occurred at the pre-operative assessment stage with the inclusion of patients who had a pre-operative diagnosis of osteoarthritis. Patients who had undergone previous knee surgery, inflammatory arthropathy or had a significant co-morbidities were excluded from the trial.

The study was granted full ethical approval from the Multisite Research Ethics Committee and the Local Research Ethics Committee. Informed consent was obtained from each patient following a full explanation and provision of all necessary patient information.

A single knee design was used in this study (PFC Sigma® Posterior Stabilised; DePuy Inc., Warsaw, IN, USA) with all components being cemented using Palacos®cement. The femoral component was constant for all patients with the tibial component being randomised into two main groups (MB vs FB) using a third party computerised randomisation process.

Each patient was randomised into receiving either FB or MB prosthesis, and sub-randomisation was performed to determine whether the patella would be resurfaced or not. The need for lateral release was determined at the time of surgery using a ‘no thumb technique’ and releasing the tourniquet if required. Lateral release was performed where tilting or subluxation of the patella occurred as the knee was taken through a range of motion, before retinacular closure. There was no difference between lateral release rates between surgeons (Fig. 1).

Fig. 1
figure 1

CONSORT diagram

The surgical details of the 352 patients recruited into the trial were used. The two groups were matched for age, sex and body mass index (Table 1). Statistical analysis of the data was performed by an independent statistician. For normally distributed data, the two-sample t-test was used. Where the data had unequal variance or was not normally distributed, the Wilcoxon rank sum test was used.

Table 1 Cohort demographics

Results

The lateral release rate was the same for both the FB and MB designs with 17 patients in each group requiring lateral release (10 %) (p = 0.9) (Table 2).

Table 2 Lateral release rates of fixed versus mobile bearings

There was a statistically significant difference, however, in the lateral release rates between the patients who had their patella resurfaced and those who did not (6 vs 14 %) (p = 0.0179) (Table 3).

Table 3 Lateral release rates of patella resurfacing versus retention

Closer analysis of the data including sub-randomisation of the bearing type, revealed an insignificant difference in the lateral release rates between those who had patellar resurfacing and those who did not in the FB group (8 vs 11 %) (p = 0.4). However, there was a significantly lower rate of lateral release in the patients in the MB group who had patellar resurfacing compared to those who did not (3 vs 16 %) (p = 0.009) (Table 4).

Table 4 Lateral release rates of fixed versus mobile versus patella

Discussion

Lateral release has been performed with FB TKA to optimise patellar tracking [10]; however, it is not without complications by jeopardising soft tissue and wound healing [10, 11]. Lateral release has also been proposed as a cause of avascular necrosis of the patella by interrupting the blood supply [11, 12]. Scuderi et al. demonstrated a higher incidence of vascular compromise to the patella when lateral release was performed [12]. If MB reduces the lateral release rate it may therefore reduce the rate of these complications.

In MB total knee arthroplasty there is potential for self-alignment of the bearing with the femoral component [13]. In an FB design that is inserted with internal rotation of the tibial component, the tibial tubercle becomes lateralized; however, with an MB design the self-alignment potentially permits correction in this circumstance [9, 13]. Rees et al. [14] provided evidence in support of this theory with in vivo fluoroscopic studies. Sawaguchi et al. [13] demonstrated in an intra-operative kinematic study that there was significantly improved patellar tracking with decreased patellofemoral contact stresses. Despite this theoretical advantage, there is no evidence as yet to demonstrate better clinical outcomes [9].

Design improvements of the femoral component of the PFC Sigma® system created a more anatomic trochlear groove that has favourably enhanced patella tracking [15]. In this study, Ballantyne et al. demonstrated a lateral release rate of 15.1 % for the newer FB PFC Sigma® design [15] compared to the older press-fit condylar prosthesis in prospective groups of patients. The addition of a rotating platform tibial component had no impact on the lateral release rate in our study; however, there was a statistically significant positive advantage for patellar resurfacing. This suggests that it is patellofemoral congruency rather than patellofemoral alignment that determines the need for lateral release in TKR.

The data also show the positive effects of patellar resurfacing and MB TKR which together gave the lowest lateral release rates of all groups. Perhaps the benefits offered by the rotating platform design which allows self-alignment of the patella are only realised once the patellofemoral geometry has been optimised. We believe that patellar resurfacing may therefore reduce the need for lateral release in MB knees and should be considered when tracking is suboptimal at the time of assessment with trial components in situ.