Abstract
This ablation rate study has been carried out with a fibre-deliverable, 308 nm, XeCl laser, producing long pulses of 200 ns as opposed to the usual 10–20 ns. This in-depth study aimed to evaluate the ablation rates for this longer pulse length. The effects on the ablation depth of dentine, enamel, soft tissue and bone were investigated ex vivo. Radiant exposure, number of pulses, pulse repetition rate and spot size were independently varied. For all tissues, ablation depth per pulse was found to increase, initially, linearly with radiant exposure. For both dentine and soft tissue a saturation radiant exposure was determined; thereafter the ablation rate decreased. The depth per pulse increased linearly with repetition rate but decreased logarithmically with both number of pulses and spot size. The ablation depth due to a 200 ns pulse is comparable to that caused by a 10–20 ns pulse but has the advantage of fibre delivery.
Similar content being viewed by others
References
Keller U, Hibst R (1993) Lasers in dentistry clinical application today and tomorrow. In: SPIE Proceedings, dental applications lasers 2080:2–7
White JM, Gekelman D, Shin KB, Park JS, Swenson TO, Rouse BP, Tran KT, Bullard SL, Scott-Beckles LB, Oto MG, Buhler JS, Yamamoto A (2002) Laser interaction with dental soft tissues: what do we know from our years of applied scientific research? In: SPIE Proceedings, lasers in dentistry VIII 4610:39–48
Frentzen M, Braun A, Koort HJ (2002) Lasers in endodontics: an overview. In: SPIE Proceedings, lasers in dentistry VIII 4610:1–8
Wigdor HA, Walsh JT, Featherstone JDB, Visuri SR, Fried D, Waldvogel JL (1995) Lasers in dentistry. Lasers Surg Med 16:103–133
Wigdor H (1997) Patient’s perception of lasers in dentistry. Lasers Surg Med 20:47–50
Keller U, Hibst R (1997) Effects of Er:YAG laser in caries treatment: a clinical pilot study. Lasers Surg Med 20:32–38
Harris M, Goodis HA, White J, Arcoria C, Simon J, Burkart J, Yessik MJ, Myers TD (2000) Pulsed Nd:Yag laser selective ablation of surface enamel caries II: histology and clinical trials. In: SPIE Proceedings, lasers in dentistry VI 3910:184–192
Li ZZ, Code JE, Van De Merwe WP (1992) Er:YAG laser ablation of enamel and dentin of human teeth: determination of ablation rates at various fluences and pulse repetition rates. Lasers Surg Med 12:625–630
Fried D, Seka W, Glena RE, Featherstone JDB (1996) Thermal response of hard dental tissues to 9- through 11-μm CO2-laser irradiation. Optical Eng 35:1976–1984
Boulnois JL (1986) Photophysical processes in recent medical laser developments: a review. Lasers Med Sci 1:47–66
Sciberras HC, Boyd N, Dair GT (2002) Morphological study of dentin irradiated by solid-state ultraviolet lasers: a comparison of wavelength and repetition rate. J Clin Laser Med Surg 20:263–268
Sciberras HC, Dair GT, Vansaarloos PP, Boyd N (2001) Combined UV and IR laser ablation of dentine. J Clin Lasers Surg Med 19:21–27
Frenten M, Koort HJ, Thiensiri I (1992) Excimer lasers in dentistry: future possibilities with advanced technology. Quintessence Int 23:117–131
Neev J, Liaw LHL, Stabholtz A, Torabinejad M, Fujishige JT, Berns MW (1992) Tissue alteration and thermal characteristics of excimer laser interaction with dentin. In: SPIE Proceedings, laser surgery: advanced characterization, therapeutics, and systems III 1643:386–397
Pearson GJ, McDonald AV (1994) Invited review: use of infra-red and ultra-violet lasers in the removal of dental hard tissue. Lasers Med Sci 9:227–237
Tasev E, Delacretaz G, Woste L (1990) Drilling in human enamel and dentin with lasers: a comparative study. In: SPIE Proceedings, laser surgery: advanced characterization, therapeutics, and systems II 1200:437–445
Liesenhoff T (1992) 308 nm Excimer laser in endontics. In: SPIE Proceedings, laser surgery: advanced characterization, therapeutics, and systems III 1643:380–385
Pini R, Salimbeni R, Vannini M, Barone R, Clauser C (1989) Laser dentistry: a new application of excimer in root canal therapy. Lasers Surg Med 9:352–357
Lewandrowski KU, Lorente C, Schomacker KT, Flotte TJ, Wilkes JW, Deutsch TF (1996) Use of Er:YAG laser for improved plating in maxillofacial surgery: comparison of bone healing in laser and drill osteotomies. Lasers Surg Med 19:40–45
Neev J, Liaw LHL, Berns MW (1995) Surface morphology and ablation characteristics of two excimer lasers: a comparison of pulse duration effect. In: SPIE Proceedings, lasers in dentistry 2394:180–187
Farrar SR, Attrill DC, Dickinson MR, Kng TA, Blinkthorn AS (1997) Etch rate and spectroscopic ablation studies of Er:YAG laser-irradiated dentine. Appl Optics 36:5641–5646
Willms L, Herschel A, Niemz MH, Pioch T (1996) Preparation of dental hard tissue with picosecond laser pulses. Lasers Med Sci 11:45–51
Gijsbers GHM, van den Broecke DG, Sprangers RLH, van Gemert MJC (1992) Effect of force on ablation depth for a XeCl excimer laser beam delivered by an optical fiber in contact with arterial tissue under saline. Lasers Surg Med 12:576–584
Pearson GJ (1994) The application of excimer laser radiation to dental hard tissue. Nanobiology 3:223–232
Jacques SL (1993) Role of tissues optics and pulse duration on tissue effects during high-power laser irradiation. Appl Optics 32:2447–2454
Neev J, Raney DV, Whalen WE, Fujishige JT, Ho PD, McGrann JV, Berns MW (1992) Dentin ablation with two excimer lasers: a comparative study of physical characteristics. Lasers Life Sci 5:129–153
Hame H, Voss R, Papioannou T, Grundfest W, Johnson R (1990) The effect of the 308 nm excimer laser on tooth dentin. In: SPIE Proceedings, laser surgery: advanced characterisation, therapeutics, and systems II 1200:452–458
Izatt JA, Albagli D, Britton M, Jubas JM, Itzkan I, Feld MS (1991) Wavelength dependence of pulsed laser ablation of calcified tissue. Lasers Surg Med 11:238–249
Venugopalan V, Nishioka NS, Mikic BB (1991) The effect of CO2 laser pulse repetition rate on tissue ablation and thermal damage. IEEE Trans Biomed Eng 38:1049–1052
Dressel M, Jahn R, Neu W, Jungbluth KH (1991) Studies in fiber guided excimer laser surgery for cutting and drilling bone and meniscus. Lasers Surg Med 11:569–579
Eyett M, Bauerle D (1987) Influence of the beam spot size on ablation rates in pulsed-laser processing. Appl Phys Lett 51:2054–2055
Lee C, Ragadio J, Fried D (2000) Influence of wavelength and pulse duration on peripheral thermal and mechanical damage to dentin and alveolar bone during IR laser ablation. In: SPIE Proceedings, lasers in dentistry VI 3910:193–203
Acknowledgements
This work was carried out under an Engineering and Physical Science Research Council grant.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Murray, A.K., Dickinson, M.R. Tissue ablation-rate measurements with a long-pulsed, fibre-deliverable 308 nm excimer laser. Lasers Med Sci 19, 127–138 (2004). https://doi.org/10.1007/s10103-004-0315-9
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10103-004-0315-9