Medium-term result of Elite Plus hip arthroplasty: the second modular evolution of the original Charnley low-friction arthroplasty

  • Original Article
  • Published:
Journal of Orthopaedic Science

Abstract

Background

The Elite Plus total hip arthroplasty (THA) system was introduced as the second modular evolution of the original Charnley low-friction arthroplasty. However, the results of the Elite Plus THA are currently considered to be inconsistent. The aim of this study was to clarify the medium-term results and the factors affecting the results of Elite Plus THA.

Methods

The results of 97 Elite Plus THAs in 87 patients at 5 years or more postoperatively were reviewed. The patients’ mean age was 65 years, and follow-up averaged 8.0 years. The Elite Plus stem was used in all hips, and three designs of socket, Hylamer Ogee in 40 hips, Wroblewski Offset Bore in 38, and Charnley Ogee in 19, were implanted. The survival rates with loosening and revision as endpoints were analyzed. To clarify the risk factors affecting the survival rates, Charnley’s functional classification, Bombelli’s classification, Crowe’s classification, structural bone grafting of the acetabulum, cup design, stem position, and grading of the cement mantle were evaluated.

Results

The stem survival rate with aseptic loosening as the endpoint was 96 %, and the rate with revision was 100 %. The cup survival rate with aseptic loosening as the endpoint was 80 %, and the rate with revision was 93 %. Using a Cox proportional hazards model, cup design was a potential confounding factor for acetabular loosening (p = 0.001). The loosening rates were 31 % for Hylamer Ogee sockets at 9.4 years, 8 % for Wroblewski Offset Bore sockets at 7.4 years, and 0 % for Charnley Ogee sockets at 5.6 years. No factors were significantly related to acetabular revision and femoral loosening.

Conclusions

The results of Elite Plus stem THA were excellent in the medium term. The cup design was a potential confounding factor for acetabular loosening, and the Hylamer Ogee socket was associated with a poor result.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2

Similar content being viewed by others

References

  1. Wroblewski BM, Siney PD. Charnley low-friction arthroplasty of the hip. Long-term results. Clin Orthop Relat Res. 1993;292:191–201.

    Google Scholar 

  2. Walton NP, Darrah C, Shepstone L, Donell ST, Phillips H. The Elite Plus total hip arthroplasty: the need for radiological surveillance. J Bone Joint Surg Br. 2005;87(4):458–62.

    Article  PubMed  CAS  Google Scholar 

  3. Norton MR, Yarlagadda R, Anderson GH. Catastrophic failure of the Elite Plus total hip replacement, with a Hylamer acetabulum and Zirconia ceramic femoral head. J Bone Joint Surg Br. 2002;84(5):631–5.

    Article  PubMed  CAS  Google Scholar 

  4. Rowsell M, Der Tavitian J, Birtwistle S, Power R. Survivorship of the Charnley Elite Plus cemented femoral stem. Int Orthop. 2005;29(4):214–8.

    Article  PubMed  CAS  Google Scholar 

  5. Hauptfleisch J, Glyn-Jones S, Beard DJ, Gill HS, Murray DW. The premature failure of the Charnley Elite-Plus stem: a confirmation of RSA predictions. J Bone Joint Surg Br. 2006;88(2):179–83.

    Article  PubMed  CAS  Google Scholar 

  6. Kalairajah Y, Azurza K, Molloy S, Hulme C, Cronin M, Drabu KJ. Is the Charnley evolution working? A five-year outcome study. Acta Orthop Belg. 2004;70(4):315–21.

    PubMed  Google Scholar 

  7. Kanno D, Haruhuji T, Sasaki S, Kojima H, Sigematsu S, Kaneko T, Masuda T, Abe S. X-ray assessment of Charnley Elite Plus THA (over five year follow-up). Proceedings of the Japanese Research Association for Replacement Arthroplasty 2006;38:34-5 (in Japanese).

  8. Kim YH, Kim JS, Yoon SH. Long-term survivorship of the Charnley Elite Plus femoral component in young patients. J Bone Joint Surg Br. 2007;89(4):449–54.

    Article  PubMed  Google Scholar 

  9. Chmell MJ, Poss R, Thomas WH, Sledge CB. Early failure of Hylamer acetabular inserts due to eccentric wear. J Arthroplasty. 1996;11(3):351–3.

    Article  PubMed  CAS  Google Scholar 

  10. Livingston BJ, Chmell MJ, Spector M, Poss R. Complications of total hip arthroplasty associated with the use of an acetabular component with a Hylamer liner. J Bone Joint Surg Am. 1997;79(10):1529–38.

    PubMed  CAS  Google Scholar 

  11. Graeter JH, Nevins R. Early osteolysis with Hylamer acetabular liners. J Arthroplasty. 1998;13(4):464–6.

    Article  PubMed  CAS  Google Scholar 

  12. Vaughn BK, Dameron TB Jr. Bauer TW, Mochida Y, Akisue T, Eberle RW. Early osteolysis following total hip arthroplasty with use of a Hylamer liner in combination with a modular ceramic femoral head. A case report. J Bone Joint Surg Am. 1999;81(10):1446–9.

    PubMed  CAS  Google Scholar 

  13. Charnley J. The long-term results of low-friction arthroplasty of the hip performed as a primary intervention. J Bone Joint Surg Br. 1972;54(1):61–76.

    PubMed  CAS  Google Scholar 

  14. Merle D’Aubigné R, Postel M. Functional results of hip arthroplasty with acrylic prostesis. J Bone Joint Surg Am. 1954;36:451–75.

    Google Scholar 

  15. DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop. 1976;121:20–32.

    PubMed  Google Scholar 

  16. Gruen TA, McNeice GM, Amstutz HC. Modes of failure of cemented stem-type femoral components; a radiographic analysis of loosening. Clin Orthop. 1979;141:17–27.

    PubMed  Google Scholar 

  17. Hodgkinson JP, Shelley P, Wroblewski BM. The correlation between the roentgenographic appearance and operative findings at the bone-cement junction of the socket in Charnley low friction arthroplasties. Clin Orthop Relat Res. 1988;228:105–9.

    PubMed  Google Scholar 

  18. Harris WH, McCarthy JC Jr, O’Neill DA. Femoral component loosening using contemporary techniques of femoral cement fixation. J Bone Joint Surg Am. 1982;64(7):1063–7.

    PubMed  CAS  Google Scholar 

  19. Bombelli R. Classification of osteoarthritis of the hip. In: Osteoarthritis of the hip: classification and pathogenesis: the role of osteotomy as a consequent therapy. 2nd ed. Berlin, etc: Springer, 1983:89- 99.

  20. Crowe JF, Mani VJ, Ranawat CS. Total hip replacement in congenital dislocation and dysplasia of the hip. J Bone Joint Surg Am. 1979;61(1):15–23.

    PubMed  CAS  Google Scholar 

  21. Mulroy WF, Estok DM, Harris WH. Total hip arthroplasty with use of so-called second-generation cementing techniques. A fifteen-year-average follow-up study. J Bone Joint Surg Am. 1995;77(12):1845–52.

    PubMed  CAS  Google Scholar 

  22. Espehaug B, Furnes O, Engesaeter LB, Havelin LI. 18 years of results with cemented primary hip prostheses in the Norwegian Arthroplasty Register: concerns about some newer implants. Acta Orthop. 2009;80(4):402–12.

    Article  PubMed  Google Scholar 

  23. Wroblewski BM, Fleming PA, Siney PD. Charnley low-frictional torque arthroplasty of the hip. 20-to-30 year results. J Bone Joint Surg Br. 1999;81(3):427–30.

    Article  PubMed  CAS  Google Scholar 

  24. Kobayashi S, Takaoka K, Saito N, Hisa K. Factors affecting aseptic failure of fixation after primary Charnley total hip arthroplasty. Multivariate survival analysis. J Bone Joint Surg Am. 1997;79(11):1618–27.

    PubMed  CAS  Google Scholar 

  25. Berry DJ, Harmsen WS, Cabanela ME, Morrey BF. Twenty-five-year survivorship of two thousand consecutive primary Charnley total hip replacements: factors affecting survivorship of acetabular and femoral components. J Bone Joint Surg Am 2002; 84-A (2):171-7.

  26. Garcia-Cimbrelo E, Cruz-Pardos A, Cordero J, Sanchez-Sotelo J. Low-friction arthroplasty in patients younger than 40 years old: 20- to 25-year results. J Arthroplasty. 2000;15(7):825–32.

    Article  PubMed  CAS  Google Scholar 

  27. Chougle A, Hemmady MV, Hodgkinson JP. Long-term survival of the acetabular component after total hip arthroplasty with cement in patients with developmental dysplasia of the hip. J Bone Joint Surg Am. 2006;88(1):71–9.

    Article  PubMed  Google Scholar 

  28. Callaghan JJ, Forest EE, Olejniczak JP, Goetz DD, Johnston RC. Charnley total hip arthroplasty in patients less than fifty years old. A twenty to twenty-five-year follow-up note. J Bone Joint Surg Am. 1998;80(5):704–14.

    PubMed  CAS  Google Scholar 

  29. Wroblewski BM, Siney PD, Fleming PA. Wear of enhanced ultra-high molecular-weight polyethylene (Hylamer) in combination with a 22.225 mm diameter zirconia femoral head. J Bone Joint Surg Br. 2003;85(3):376–9.

    Article  PubMed  CAS  Google Scholar 

  30. Barrack RL, Mulroy RD Jr. Harris WH. Improved cementing techniques and femoral component loosening in young patients with hip arthroplasty. A 12-year radiographic review. J Bone Joint Surg Br. 1992;74(3):385–9.

    PubMed  CAS  Google Scholar 

Download references

Conflict of interest

The authors did not receive and will not receive any benefits or funding from any commercial party related directly or indirectly to the subject of this article.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Taichi Irie.

About this article

Cite this article

Irie, T., Oyama, M., Kita, A. et al. Medium-term result of Elite Plus hip arthroplasty: the second modular evolution of the original Charnley low-friction arthroplasty. J Orthop Sci 17, 699–704 (2012). https://doi.org/10.1007/s00776-012-0296-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00776-012-0296-7

Keywords

Navigation