Abstract
Purpose
To evaluate the safety and efficacy of three-dimensional (3D) navigation-guided polymethylmethacrylate (PMMA)-augmented thoracic and lumbar pedicle screw fixation for the osteoporotic patients.
Methods
27 consecutive osteoporosis patients with a variety of spinal disorders who underwent 3D navigation-guided PMMA-augmented pedicle screw fixation were evaluated clinically and radiologically in the perioperative and 1-year follow-up period. The improvement of Japanese Orthopaedic Association (JOA) scores was analyzed. PMMA leakage and other complications were inspected intraoperatively and postoperatively. Screw loosening and bone fusion were evaluated radiographically during follow-up.
Results
8 patients had thoracic and lumbar fractures; 18 patients had degenerative spinal disorders; one patient had revision surgery. One patient died of postoperative pneumonia. Each of the other 26 patients was followed up regularly at 3, 6, 12 and 18 months postoperatively. The mean therapeutic improvement rate is 39.6 % evaluated by JOA scores. 2 patients (7.4 %) had leakage into the spinal canal in front of the posterior longitudinal ligament and two patients (7.4 %) had leakage into the prevertebral soft tissue inspected by the postoperative CT scans. No pedicle cortex breach and cement leakage surrounding pedicle cortex were observed. None of patients complained of dyspnoea and showed evidence of pulmonary embolism. Bone fusions were found in 20 patients (bony fusion rate 76.9 %) at the 12-month follow-up and no screw loosening occurs.
Conclusion
The results show favorable outcome using 3D navigation-guided PMMA-augmented thoracic and lumbar pedicle screw fixation for the osteoporosis patients both clinically and radiologically.
Similar content being viewed by others
References
Cummings SR, Melton LJ (2002) Epidemiology and outcomes of osteoporotic fractures. Lancet 359(9319):1761–1767
Paxinos O, Tsitsopoulos PP, Zindrick MR, Voronov LI et al (2010) Evaluation of pullout strength and failure mechanism of posterior instrumentation in normal and osteopenic thoracic vertebrae. J Neurosurg Spine 13(4):469–476
Halvorson TL, Kelley LA, Thomas KA, Whitecloud TS 3rd, Cook SD (1994) Effects of bone mineral density on pedicle screw fixation. Spine (Phila Pa 1976) 19(21):2415–2420
Garfin SR, Yuan HA, Reiley MA (2001) New technologies in spine: kyphoplasty and vertebroplasty for the treatment of painful osteoporotic compression fractures. Spine (Phila Pa 1976) 26(14):1511–1515
Frankel BM, Jones T, Wang C (2007) Segmental polymethylmethacrylate-augmented pedicle screw fixation in patients with bone softening caused by osteoporosis and metastatic tumor involvement: a clinical evaluation. Neurosurgery 61(3):531–537 (discussion 537–538)
Fuentes S, Blondel B, Metellus P, Gaudart J, Adetchessi T, Dufour H (2010) Percutaneous kyphoplasty and pedicle screw fixation for the management of thoraco-lumbar burst fractures. Eur Spine J 19(8):1281–1287
Becker S, Chavanne A, Spitaler R, Kropik K et al (2008) Assessment of different screw augmentation techniques and screw designs in osteoporotic spines. Eur Spine J 17(11):1462–1469
Gao M, Lei W, Wu Z, Liu D, Shi L (2011) Biomechanical evaluation of fixation strength of conventional and expansive pedicle screws with or without calcium based cement augmentation. Clin Biomech (Bristol, Avon) 26(3):238–244
Waits C, Burton D, McIff T (2009) Cement augmentation of pedicle screw fixation using novel cannulated cement insertion device. Spine (Phila Pa 1976) 34(14):E478–E483
Ren H, Shen Y, Zhang YZ, Ding WY et al (2010) Correlative factor analysis on the complications resulting from cement leakage after percutaneous kyphoplasty in the treatment of osteoporotic vertebral compression fracture. J Spinal Disord Tech 23(7):e9–e15
Wilkes RA, Mackinnon JG, Thomas WG (1994) Neurological deterioration after cement injection into a vertebral body. J Bone Joint Surg Br 76(1):155
Rajan VV, Kamath V, Shetty AP, Rajasekaran S (2010) Iso-C3D navigation assisted pedicle screw placement in deformities of the cervical and thoracic spine. Indian J Orthop 44 (2):163–168
Kosmopoulos V, Schizas C (2007) Pedicle screw placement accuracy: a meta-analysis. Spine (Phila Pa 1976) 32(3):E111–E120
Engelke K, Adams JE, Armbrecht G, Augat P et al (2008) Clinical use of quantitative computed tomography and peripheral quantitative computed tomography in the management of osteoporosis in adults: the 2007 ISCD official positions. J Clin Densitom 11(1):123–162
Sawakami K, Yamazaki A, Ishikawa S, Ito T, Watanabe K, Endo N (2012) Polymethylmethacrylate augmentation of pedicle screws increases the initial fixation in osteoporotic spine patients. J Spinal Disord Tech 25(2):E28–E35
Gaines RW Jr (2000) The use of pedicle-screw internal fixation for the operative treatment of spinal disorders. J Bone Joint Surg Am 82-A(10):1458–1476
Lane NE (2006) Epidemiology, etiology, and diagnosis of osteoporosis. Am J Obstet Gynecol 194(2 Suppl):S3–S11
Sarzier JS, Evans AJ, Cahill DW (2002) Increased pedicle screw pullout strength with vertebroplasty augmentation in osteoporotic spines. J Neurosurg 96(3 Suppl):309–312
Wittenberg RH, Lee KS, Shea M, White AA 3rd, Hayes WC (1993) Effect of screw diameter, insertion technique, and bone cement augmentation of pedicular screw fixation strength. Clin Orthop Relat Res 296:278–287
Frankel BM, Jones T, Wang C (2007) Segmental polymethylmethacrylate-augmented pedicle screw fixation in patients with bone softening caused by osteoporosis and metastatic tumor involvement: a clinical evaluation. Neurosurgery 61(3):531–537 (discussion 537–538)
Chang MC, Liu CL, Chen TH (2008) Polymethylmethacrylate augmentation of pedicle screw for osteoporotic spinal surgery: a novel technique. Spine (Phila Pa 1976) 33(10):E317–E324
Rajasekaran S, Vidyadhara S, Ramesh P, Shetty AP (2007) Randomized clinical study to compare the accuracy of navigated and non-navigated thoracic pedicle screws in deformity correction surgeries. Spine (Phila Pa 1976) 32(2):E56–E64
Hulme PA, Krebs J, Ferguson SJ, Berlemann U (2006) Vertebroplasty and kyphoplasty: a systematic review of 69 clinical studies. Spine (Phila Pa 1976) 31(17):1983–2001
Merloz P, Troccaz J, Vouaillat H, Vasile C et al (2007) Fluoroscopy-based navigation system in spine surgery. Proc Inst Mech Eng H 221(7):813–820
Chen LH, Tai CL, Lee DM, Lai PL et al (2011) Pullout strength of pedicle screws with cement augmentation in severe osteoporosis: a comparative study between cannulated screws with cement injection and solid screws with cement pre-filling. BMC Musculoskelet Disord 12:33
Conflict of interest
None of the authors has any potential conflict of interest.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Yuan, Q., Zhang, G., Wu, J. et al. Clinical evaluation of the polymethylmethacrylate-augmented thoracic and lumbar pedicle screw fixation guided by the three-dimensional navigation for the osteoporosis patients. Eur Spine J 24, 1043–1050 (2015). https://doi.org/10.1007/s00586-013-3131-8
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00586-013-3131-8