Intra-session repeatability of iridocorneal angle measurements provided by a Scheimpflug photography-based system in healthy eyes

  • Miscellaneous
  • Published:
Graefe's Archive for Clinical and Experimental Ophthalmology Aims and scope Submit manuscript

Abstract

Purpose

The purpose of this study was to evaluate intra-session repeatability of measurements of the iridocorneal angle at different meridians in the nasal and temporal areas in healthy eyes using the Sirius Scheimpflug photography-based system in glaucoma analysis mode.

Methods

A total of 43 eyes of 43 patients ranging in age from 36 to 79 years were enrolled in the study. All eyes received a comprehensive ophthalmologic examination including a complete anterior segment analysis with the Costruzione Strumenti Oftalmici [CSO] Sirius system. Three consecutive measurements of nasal and temporal angles at 0°, ±10°, ±20°, and ±30° meridians were obtained in order to assess the intra-session repeatability of iridocorneal angle measurements provided by the device using the glaucoma analysis mode. Within-subject standard deviation (Sw), coefficient of variation (CV), and intraclass correlation coefficient (ICC) values were calculated.

Results

The mean Sw was 1.07 ± 1.09°, 1.22 ± 1.53°, 0.66 ± 0.51°, 0.86 ± 0.57°, 0.68 ± 0.65°, 0.84 ± 0.68°, and 0.91 ± 0.70° at the temporal 30°, 20°, 10°, 0°, −10°, −20°, and −30° positions, respectively. Mean Sw was 3.13 ± 3.15°, 3.43 ± 3.63°, 2.75 ± 2.29°, 2.19 ± 1.55°, 1.90 ± 1.49°, 2.14 ± 1.74°, and 2.24 ± 2.06° at the temporal −30°, −20°, −10°, 0°, 10°, 20°, and 30° positions, respectively. Mean CV ranged from 1.36 ± 1.05 % (nasal 0° position) to 10.92 ± 13.95 % (nasal −20° position). ICC values ranged from 0.778 to 0.972.

Conclusions

The glaucoma analysis mode of the Sirius system provides consistent measurements of the iridocorneal angle at different meridians in healthy eyes, with slightly less consistency for nasal measurements. It may be considered a clinically useful non-invasive technique for the detection of potentially occludable angles.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1

Similar content being viewed by others

References

  1. Doganay S, Bozgul Firat P, Emre S, Yologlu S (2010) Evaluation of anterior segment parameter changes using the Pentacam after uneventful phacoemulsification. Acta Ophthalmol 88:601–606

    Article  PubMed  Google Scholar 

  2. Xu L, Cao WF, Wang YX, Chen CX, Jonas JB (2008) Anterior chamber depth and chamber angle and their associations with ocular and general parameters: the Beijing Eye Study. Am J Ophthalmol 145:929–936

    Article  PubMed  Google Scholar 

  3. Congdon NG, Spaeth GL, Augsburger J, Klancnik J Jr, Patel K, Hunter DG (1999) A proposed simple method for measurement in the anterior chamber angle: biometric gonioscopy. Ophthalmology 106:2161–2167

    Article  PubMed  CAS  Google Scholar 

  4. Fremont AM, Lee PP, Mangione CM, Kapur K, Adams JL, Wickstrom SL, Escarce JJ (2003) Patterns of care for open-angle glaucoma in managed care. Arch Ophthalmol 121:777–783

    Article  PubMed  Google Scholar 

  5. Friedman DS, He M (2008) Anterior chamber angle assessment techniques. Surv Ophthalmol 53:250–273

    Article  PubMed  Google Scholar 

  6. Hernández-Camarena JC, Chirinos-Saldaña P, Ramirez-Miranda A, de la Mota A, Jimenez-Corona A, Graue-Hernindez EO (2014) Repeatability, reproducibility, and agreement between three different Scheimpflug systems in measuring corneal and anterior segment biometry. J Refract Surg 30:616–621

    Article  PubMed  Google Scholar 

  7. Wang Q, Ding X, Savini G, Chen H, Feng Y, Pan C, Hua Y, Huang J (2015) Anterior chamber depth measurements using Scheimpflug imaging and optical coherence tomography: repeatability, reproducibility, and agreement. J Cataract Refract Surg 41:178–185

    Article  PubMed  Google Scholar 

  8. Masou M, Livny E, Bahar I (2015) Repeatability and intrasession reproducibility obtained by the Sirius anterior segment analysis system. Eye Contact Lens 41:107–110

    Article  Google Scholar 

  9. Bedei A, Appolloni I, Madesani A, Pietrelli A, Franceschi S, Barabesi L (2012) Repeatability and agreement of 2 Scheimpflug analyzers in measuring the central corneal thickness and anterior chamber angle, volume, and depth. Eur J Ophthalmol 22(Suppl 7):S29–S32

    Article  PubMed  Google Scholar 

  10. Milla M, Piñero DP, Amparo F, Alió JL (2011) Pachymetric measurements with a new Scheimpflug photography-based system: intraobserver repeatability and agreement with optical coherence tomography pachymetry. J Cataract Refract Surg 37:310–316

    Article  PubMed  Google Scholar 

  11. Savini G, Barboni P, Carbonelli M, Hoffer KJ (2011) Repeatability of automatic measurements by a new Scheimpflug camera combined with Placido topography. J Cataract Refract Surg 37:1809–1816

    Article  PubMed  Google Scholar 

  12. Huang J, Savini G, Hu L, Hoffer KJ, Lu W, Feng Y, Yang F, Hu X, Wang Q (2013) Precision of a new Scheimpflug and Placido-disk analyzer in measuring corneal thickness and agreement with ultrasound pachymetry. J Cataract Refract Surg 39:219–224

    Article  PubMed  Google Scholar 

  13. Prakash G, Srivastava D, Choudhuri S (2015) A novel Hartman Shack-based topography system: repeatability and agreement for corneal power with Scheimpflug + Placido topographer and rotating prism auto-keratorefractor. Int Ophthalmol Mar 28 [Epub ahead of print].

  14. Maresca, Zeri F, Palumbo P, Calosssi A (2014) Agreement and reliability in measuring central corneal thickness with a rotating Scheimpflug-Placido system and ultrasound pachymetry. Cont Lens Anterior Eye 37:442–446

    Article  PubMed  Google Scholar 

  15. de la Lopez Fuente, Sanchez-Cano A, Segura F, Fuentes-Broto L, Pinilla I (2014) Repeatability of ocular measurements with a dual-Scheimpflug analyzer in healthy eyes. Biomed Res Int 2014:808646

    Google Scholar 

  16. Piñero DP (2015) Technologies for anatomical and geometric characterization of the corneal structure and anterior segment: a review. Semin Ophthalmol 30:161–170

    Article  PubMed  Google Scholar 

  17. Bosem ME, Morsman D, Lusky M, Weinreb RN (1992) Reproducibility of quantitative anterior chamber angle measurements with Scheimpflug video imaging. J Glaucoma 1:254–257

    Article  PubMed  CAS  Google Scholar 

  18. Tun TA, Baskaran M, Perera SA, Chan AS, Cheng CY, Htoon HM, Sakata LM, Cheung CY, Aung T (2014) Sectoral variations of iridocorneal angle width and iris volume in Chinese Singaporeans: a swept-source optical coherence tomography study. Graefes Arch Clin Exp Ophthalmol 252:1127–1132

    Article  PubMed  Google Scholar 

  19. Moss SE, Klein R, Klein BE (2000) Arcus senilis and mortality in a population with diabetes. Am J Ophthalmol 129:676–678

    Article  PubMed  CAS  Google Scholar 

  20. Sugar A, Gal RL, Beck W, Ruedy KJ, Blanton CL, Feder RS, Hardten DR, Holland EJ, Lass JH, Mannis MJ, O’Keefe MB, Cornea Donor Study Group (2005) Baseline donor characteristics in the Cornea Donor Study. Cornea 24:389–396

    Article  PubMed  Google Scholar 

  21. Maruyama Y, Mori K, Ikeda Y, Ueno M, Kinoshita S (2014) Morphological analysis of age-related iridocorneal angle changes in normal and glaucomatous cases using anterior segment optical coherence tomography. Clin Ophthalmol 8:113–118

    PubMed  PubMed Central  Google Scholar 

  22. Rüfer F, Schröder A, Klettner A, Frimpong-Boateng A, Roider JB, Erb C (2010) Anterior chamber depth and iridocorneal angle in healthy White subjects: effects of age, gender and refraction. Acta Ophthalmol 88:885–890

    Article  PubMed  Google Scholar 

  23. Koretz JF, Cook CA, Kaufman PL (2001) Aging of the human lens: changes in lens shape at zero-diopter accommodation. J Opt Soc Am A Opt Image Sci Vis 18:265–272

    Article  PubMed  CAS  Google Scholar 

Download references

Conflict of interest

All authors certify that they have NO affiliations with or involvement in any organization or entity with any financial interest (such as honoraria; educational grants; participation in speakers’ bureaus; membership, employment, consultancies, stock ownership, or other equity interest; and expert testimony or patent-licensing arrangements), or non-financial interest (such as personal or professional relationships, affiliations, knowledge or beliefs) in the subject matter or materials discussed in this manuscript.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to David P. Piñero.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Ruiz-Belda, C., Piñero, D.P., Ruiz-Fortes, P. et al. Intra-session repeatability of iridocorneal angle measurements provided by a Scheimpflug photography-based system in healthy eyes. Graefes Arch Clin Exp Ophthalmol 254, 169–175 (2016). https://doi.org/10.1007/s00417-015-3105-0

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00417-015-3105-0

Keywords

Navigation