Abstract
Introduction
Evaluation of early cup movement is an important diagnostic tool to predict the likelihood of long-term implant loosening and clinical failure. The investigated cementless cup is clinically proven over 10 years, but there is a paucity of information that accurately describes the migration characteristics of this component.
Materials and methods
We retrospectively analysed the clinical outcome and migration behaviour of 60 Pinnacle 100 shells after an average 3.8-year follow-up (range 2.1–5.4 years). For migration measurement, EBRA (Einzel-Bild-Röntgen-Analyse) digital software was applied. Clinical assessment was performed using the HHS, the UCLA score and the SF-36 health survey.
Results
The clinical outcome showed excellent results with a mean HHS of 95.4 (SD 7.1) and mean UCLA of 6.9 (SD 1.3). All implants were radiologically stable within the observation period and none of the cups was at risk for aseptical loosening. EBRA analysis revealed a mean total migration of 1.4 mm (SD 0.9) (95 % CI 1.1–1.6) at 3 years. Eight cups migrated more than 1 mm within the first three postoperative months, thereafter the migration curves flattened down.
Conclusion
Surgeons may expect to find a variable amount of early migration when using the Pinnacle cup. To our knowledge, these are the first results, which show an early “impaction” of a cementless cup, followed by subsequent osseointegration. We believe that an appropriate long-term outcome of the investigated cup is ensured. The data of the present investigation will provide clinicians with useful baseline information with which to compare new cup designs.
Similar content being viewed by others
References
NJRfEaW (2011) National Joint Registry for England and Wales 8th Annual Report http://www.njrcentre.org.uk/NjrCentre/Portals/0/Documents/NJR%208th%20Annual%20Report%202011.pdf
Kindsfater KA, Sychterz Terefenko CJ, Gruen TA, Sherman CM (2012) Minimum 5-year results of modular metal-on-metal total hip arthroplasty. J Arthroplasty 27:545–550
Lubbeke A, Katz JN, Perneger TV, Hoffmeyer P (2007) Primary and revision hip arthroplasty: 5-year outcomes and influence of age and comorbidity. J Rheumatol 34:394–400
Gheduzzi S, Miles AW (2007) A review of pre-clinical testing of femoral stem subsidence and comparison with clinical data. Proc Inst Mech Eng H 221:39–46
Kolb A, Grubl A, Schneckener CD, Chiari C, Kaider A, Lass R, Windhager R (2012) Cementless total hip arthroplasty with the rectangular Titanium Zweymuller Stem: a concise follow-up, at a minimum of 20 years, of previous reports. J Bone Jt Surg Am 94:1681–1684
Stefl MD, Callaghan JJ, Liu SS, Pedersen DR, Goetz DD, Johnston RC (2012) Primary cementless acetabular fixation at a minimum of 20 years of follow-up: a concise update of a previous report. J Bone Jt Surg Am 94:234–239. doi:10.2106/JBJS.K.00237
Gotze C, Tschugunow A, Gotze HG, Bottner F, Potzl W, Gosheger G (2006) Long-term results of the metal-cancellous cementless Lubeck total hip arthroplasty: a critical review at 12.8 years. Arch Orthop Trauma Surg 126:28–35. doi:10.1007/s00402-005-0064-7
Parsch D, Jung AW, Thomsen M, Ewerbeck V, Aldinger PR (2008) Good survival of uncemented tapered stems for failed intertrochanteric osteotomy: a mean 16 year follow-up study in 45 patients. Arch Orthop Trauma Surg 128:1081–1085. doi:10.1007/s00402-007-0444-2
Krismer M, Stockl B, Fischer M, Bauer R, Mayrhofer P, Ogon M (1996) Early migration predicts late aseptic failure of hip sockets. J Bone Jt Surg Br 78:422–426
Stocks GW, Freeman MA, Evans SJ (1995) Acetabular cup migration. Prediction of aseptic loosening. J Bone Jt Surg Br 77:853–861
Mjoberg B (1991) Fixation and loosening of hip prostheses. Rev Acta Orthop Scand 62:500–508
Krismer M, Bauer R, Tschupik J, Mayrhofer P (1995) EBRA: a method to measure migration of acetabular components. J Biomech 28:1225–1236 [pii: 0021929094001776]
Stihsen C, Pabinger C, Radl R, Rehak P, Windhager R (2008) Migration of the Duraloc cup after 5 years. Int Orthop 32:791–794. doi:10.1007/s00264-007-0405-y
Stoeckl B, Brabec E, Wanner S, Krismer M, Biedermann R (2005) Radiographic evaluation of the Duraloc cup after 4 years. Int Orthop 29:14–17. doi:10.1007/s00264-004-0600-z
Stockl B, Sandow M, Krismer M, Biedermann R, Wimmer C, Frischhut B (1999) Migration of the Duraloc cup at 2 years. J Bone Jt Surg Br 81:51–53
Zahiri CA, Schmalzried TP, Szuszczewicz ES, Amstutz HC (1998) Assessing activity in joint replacement patients. J Arthroplast 13:890–895 [pii: S0883-5403(98)90195-4]
Harris WH (1969) Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Jt Surg Am 51:737–755
Ware JE Jr, Sherbourne CD (1992) The MOS 36-item short-form health survey (SF-36). I. Conceptual framework and item selection. Med Care 30:473–483
DeLee JG, Charnley J (1976) Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res 121:20–32
Ilchmann T, Kesteris U, Wingstrand H (1998) EBRA improves the accuracy of radiographic analysis of acetabular cup migration. Acta Orthop Scand 69:119–124
Ilchmann T, Franzen H, Mjoberg B, Wingstrand H (1992) Measurement accuracy in acetabular cup migration. A comparison of four radiologic methods versus roentgen stereophotogrammetric analysis. J Arthroplast 7:121–127
Bauer R, Kerschbaumer F, Poisel S, Oberthaler W (1979) The transgluteal approach to the hip joint. Arch Orthop Trauma Surg 95:47–49
Watson–Jones R (1936) Fractures of the neck of the femur. Br J Surg 23:787–808
Bertin KC, Rottinger H (2004) Anterolateral mini-incision hip replacement surgery: a modified Watson–Jones approach. Clin Orthop Relat Res 429:248–255
Campbell D, Mercer G, Nilsson KG, Wells V, Field JR, Callary SA (2009) Early migration characteristics of a hydroxyapatite-coated femoral stem: an RSA study. Int Orthop 35:483–488. doi:10.1007/s00264-009-0913-z
Strom H, Nilsson O, Milbrink J, Mallmin H, Larsson S (2007) The effect of early weight bearing on migration pattern of the uncemented CLS stem in total hip arthroplasty. J Arthroplast 22:1122–1129. doi:10.1016/j.arth.2006.11.015
Powers CC, Ho H, Beykirch SE, Huynh C, Hopper RH Jr, Engh CA Jr, Engh CA (2010) A comparison of a second- and a third-generation modular cup design: is new improved? J Arthroplast 25:514–521. doi:10.1016/j.arth.2009.02.018
Selvik G (1989) Roentgen stereophotogrammetry. A method for the study of the kinematics of the skeletal system. Acta Orthop Scand Suppl 232:1–51
Acknowledgments
The study was designed and performed as an academic investigation without any external or industrial funding sources.
Conflict of interest
The authors declare that they have no conflict of interest.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Stihsen, C., Rath, C., Radl, R. et al. Early migration characteristics of a 180° porous-coated cup with 1-mm press fit. Arch Orthop Trauma Surg 133, 707–712 (2013). https://doi.org/10.1007/s00402-013-1713-x
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00402-013-1713-x