Abstract
Objectives
To compare native aortic (AV) and mitral valve (MV) image quality on limited-dose retrospectively ECG-gated CTA of the thoracoabdominal aorta reconstructed with iterative reconstruction (IR) and filtered back projection (FBP).
Methods
Fifty patients underwent routine care retrospectively ECG-gated thoracoabdominal limited-dose 256-slice CTA. At 30 % (systole) and 75 % (diastole) of the R-R interval AV and MV were reconstructed using FBP and IR. Objective image quality [density and noise (SD of density measurement)] was measured. Two independent observers scored subjective valve image quality using four-point Likert scales.
Results
IR significantly decreased image noise, but did not alter the aorta and interventricular septum density. Interobserver variability was moderate to good. Valve image quality was scored at least moderate in most cases. IR scored one or two Likert scale points higher than FBP in 10 (first observer) and 27 (second observer) scores. Conversely, IR scored one Likert scale point lower than FBP in 1 (first observer) and 4 (second observer) scores.
Conclusions
Limited-dose retrospectively ECG-gated thoracoabdominal CTA enables moderate to excellent evaluation of AV and MV in most patients, in addition to the primary diagnostic question. Image quality is further improved by IR.
Key Points
• Thoracoabdominal computed tomographic angiography demonstrates both the aortic and mitral valves.
• Iterative reconstruction decreases image noise and increases the SNR/CNR of thoracoabdominal CTA.
• Limited-dose retrospectively gated CTA enables moderate–excellent left-sided valvular evaluation.
• Iterative CT image reconstruction improves subjective native aortic and mitral valvular evaluation.
Similar content being viewed by others
Abbreviations
- AV:
-
Aortic valve
- CNR:
-
Contrast-to-noise ratio
- CTA:
-
Computed tomography angiography
- CTDIvol :
-
Volume CT dose index
- DLP:
-
Dose–length product
- ED:
-
Effective dose
- FBP:
-
Filtered back projection
- HU:
-
Hounsfield units
- IR:
-
Iterative reconstruction
- MV:
-
Mitral valve
- SD:
-
Standard deviation
- SNR:
-
Signal-to-noise ratio
References
Iung B, Baron G, Butchart EG et al (2003) A prospective survey of patients with valvular heart disease in Europe: the Euro Heart Survey on Valvular Heart Disease. Eur Heart J 24:1231–1243
Vahanian A, Baumgartner H, Bax J et al (2007) Guidelines on the management of valvular heart disease: the Task Force on the Management of Valvular Heart Disease of the European Society of Cardiology. Eur Heart J 28:230–268
de Heer LM, Habets J, Chamuleau SA et al (2012) Multidetector row computed tomography assessment of the native aortic and mitral valve: a call for routine assessment of left-sided heart valves during coronary computed tomography. Cardiol Rev. doi:10.1097/CRD.0b013e318250eaaa
Budde RP, Huo F, Cramer MJ et al (2010) Simultaneous aortic and coronary assessment in abdominal aortic aneurysm patients by thoraco-abdominal 64-detector-row CT angiography: estimate of the impact on preoperative management: a pilot study. Eur J Vasc Endovasc Surg 40:196–201
Fleischmann D, Boas FE (2011) Computed tomography—old ideas and new technology. Eur Radiol 21:510–517
Scheffel H, Stolzmann P, Schlett CL et al (2011) Coronary artery plaques: Cardiac CT with model-based and adaptive-statistical iterative reconstruction technique. Eur J Radiol 81:e363–e369
Funama Y, Taguchi K, Utsunomiya D et al (2011) Combination of a low-tube-voltage technique with hybrid iterative reconstruction (iDose) algorithm at coronary computed tomographic angiography. J Comput Assist Tomogr 35:480–485
Moscariello A, Takx RA, Schoepf UJ et al (2011) Coronary CT angiography: image quality, diagnostic accuracy, and potential for radiation dose reduction using a novel iterative image reconstruction technique-comparison with traditional filtered back projection. Eur Radiol 21:2130–2138
Park EA, Lee W, Kim KW et al (2011) Iterative reconstruction of dual-source coronary CT angiography: assessment of image quality and radiation dose. Int J Cardiovasc Imaging. doi:10.1007/s10554-011-0004-2
Renker M, Nance JW Jr, Schoepf UJ et al (2011) Evaluation of heavily calcified vessels with coronary CT angiography: comparison of iterative and filtered back projection image reconstruction. Radiology 260:390–399
Utsunomiya D, Weigold WG, Weissman G, Taylor AJ (2011) Effect of hybrid iterative reconstruction technique on quantitative and qualitative image analysis at 256-slice prospective gating cardiac CT. Eur Radiol 22:1287–1294
Habets J, Symersky P, de Mol BA, Mali WP, Leiner T, Budde RP (2011) A novel iterative reconstruction algorithm allows reduced dose multidetector-row CT imaging of mechanical prosthetic heart valves. Int J Cardiovasc Imaging. doi:10.1007/s10554-011-9954-7
Winklehner A, Karlo C, Puippe G et al (2011) Raw data-based iterative reconstruction in body CTA: evaluation of radiation dose saving potential. Eur Radiol 21:2521–2526
Cornfeld D, Israel G, Detroy E, Bokhari J, Mojibian H (2011) Impact of Adaptive Statistical Iterative Reconstruction (ASIR) on radiation dose and image quality in aortic dissection studies: a qualitative and quantitative analysis. AJR Am J Roentgenol 196:W336–W340
Deak PD, Smal Y, Kalender WA (2010) Multisection CT protocols: sex- and age-specific conversion factors used to determine effective dose from dose-length product. Radiology 257:158–166
Noel PB, Fingerle AA, Renger B, Munzel D, Rummeny EJ, Dobritz M (2011) Initial performance characterization of a clinical noise-suppressing reconstruction algorithm for MDCT. AJR Am J Roentgenol 197:1404–1409
Philips Healthcare (2011) iDose4 iterative reconstruction technique. Philips Healthcare Whitepaper. Available via https://www.healthcare.philips.com/pwc_hc/main/shared/Assets/Documents/ct/idose_white_paper_452296267841.pdf. Accessed June 2012
Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174
Goetti R, Baumuller S, Feuchtner G et al (2010) High-pitch dual-source CT angiography of the thoracic and abdominal aorta: is simultaneous coronary artery assessment possible? AJR Am J Roentgenol 194:938–944
Hausleiter J, Meyer T, Hadamitzky M et al (2006) Radiation dose estimates from cardiac multislice computed tomography in daily practice: impact of different scanning protocols on effective dose estimates. Circulation 113:1305–1310
Hausleiter J, Meyer T, Hermann F et al (2009) Estimated radiation dose associated with cardiac CT angiography. JAMA 301:500–507
Cademartiri F, La GL, Runza G et al (2007) Influence of convolution filtering on coronary plaque attenuation values: observations in an ex vivo model of multislice computed tomography coronary angiography. Eur Radiol 17:1842–1849
Seifarth H, Raupach R, Schaller S et al (2005) Assessment of coronary artery stents using 16-slice MDCT angiography: evaluation of a dedicated reconstruction kernel and a noise reduction filter. Eur Radiol 15:721–726
Acknowledgements
We would like to thank Robert Valkenburg, Anne Manders, Jamal Moumni and Karin Thijn for technical assistance.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Willemink, M.J., Habets, J., de Jong, P.A. et al. Iterative reconstruction improves evaluation of native aortic and mitral valves by retrospectively ECG-gated thoracoabdominal CTA. Eur Radiol 23, 968–974 (2013). https://doi.org/10.1007/s00330-012-2673-1
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00330-012-2673-1