Development and Evaluation of a Novel Curved Biopsy Device for CT-Guided Biopsy of Lesions Unreachable Using Standard Straight Needle Trajectories

  • Technical Note
  • Published:
CardioVascular and Interventional Radiology Aims and scope Submit manuscript

Abstract

Purpose

To evaluate the feasibility of a novel curved CT-guided biopsy needle prototype with shape memory to access otherwise not accessible biopsy targets.

Methods and Materials

A biopsy needle curved by 90° with specific radius was designed. It was manufactured using nitinol to acquire shape memory, encased in a straight guiding trocar to be driven out for access of otherwise inaccessible targets. Fifty CT-guided punctures were conducted in a biopsy phantom and 10 CT-guided punctures in a swine corpse. Biposies from porcine liver and muscle tissue were separately gained using the biopsy device, and histological examination was performed subsequently.

Results

Mean time for placement of the trocar and deployment of the inner biopsy needle was ~205 ± 69 and ~93 ± 58 s, respectively, with a mean of ~4.5 ± 1.3 steps to reach adequate biopsy position. Mean distance from the tip of the needle to the target was ~0.7 ± 0.8 mm. CT-guided punctures in the swine corpse took relatively longer and required more biopsy steps (~574 ± 107 and ~380 ± 148 s, 8 ± 2.6 steps). Histology demonstrated appropriate tissue samples in nine out of ten cases (90%).

Conclusions

Targets that were otherwise inaccessible via standard straight needle trajectories could be successfully reached with the curved biopsy needle prototype. Shape memory and preformed size with specific radius of the curved needle simplify the target accessibility with a low risk of injuring adjacent structures.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5

References

  1. Gupta S, Madoff DC, Ahrar K, Morello FA, Wallace MJ, Murthy R, et al. CT-guided needle biopsy of deep pelvic lesions by extraperitoneal approach through iliopsoas muscle. Cardiovasc Intervent Radiol. 2003;26:534–8.

    Article  PubMed  Google Scholar 

  2. Gupta S, Ahrar K, Morello FA, Wallace MJ, Madoff DC, Hicks ME. Using a coaxial technique with a curved inner needle for CT-guided fine-needle aspiration biopsy. Am J Roentgenol. 2002;179:109–12.

    Article  Google Scholar 

  3. Chojniak R, Isberner R, Viana L, Yu L, Aita A, Soares F. Computed tomography guided needle biopsy: experience from 1,300 procedures. Sao Paulo Med J. 2006;124:10–4.

    Article  PubMed  Google Scholar 

  4. Warnock NG. Technical note. Curved needle technique for the avoidance of interposed structures in CT-guided percutaneous fine needle biopsy. J Comput Assist Tomogr. 1996;20:826–8.

    Article  CAS  PubMed  Google Scholar 

  5. Levy WJ, Oro JJ. Curved biopsy needle for stereotactic surgery: a technical note. Neurosurgery. 1984;15:82–5.

    Article  PubMed  Google Scholar 

  6. Hawkins IF, Caridi JG. Curved stylus for redirection of the fine-needle guide. Radiology. 1984;151:530.

    Article  PubMed  Google Scholar 

  7. Carrasco CH, Wallace S, Charnsangavej C. Aspiration biopsy: use of a curved needle. Radiology. 1985;155:254.

    Article  CAS  PubMed  Google Scholar 

  8. Singh AK, Leeman J, Shankar S, Ferrucci JT. Core biopsy with curved needle technique. Am J Roentgenol. 2008;191:1745–50.

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Maximilian Franz Schulze-Hagen.

Ethics declarations

Conflict of interest

On behalf of all authors, the corresponding author states that there is no conflict of interest.

Ethical Approval

All applicable institutional and/or national guidelines for the care and use of animals were followed.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Schulze-Hagen, M.F., Pfeffer, J., Zimmermann, M. et al. Development and Evaluation of a Novel Curved Biopsy Device for CT-Guided Biopsy of Lesions Unreachable Using Standard Straight Needle Trajectories. Cardiovasc Intervent Radiol 40, 924–929 (2017). https://doi.org/10.1007/s00270-017-1597-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00270-017-1597-1

Keywords

Navigation