Analysis of sacro-iliac joint screw fixation: does quality of reduction and screw orientation influence joint stability? A biomechanical study

  • Original Paper
  • Published:
International Orthopaedics Aims and scope Submit manuscript

Abstract

Introduction

Treatment of posterior pelvic ring injuries is frequently associated with pain or/and high mortality rates. Percutaneous sacro-iliac (SI) screw fixation has proved to be one of the methods of choice, providing minimal operative time, blood loss and wound-related morbidity. However, fixation failures due to secondary fracture dislocation or screw backing out have been reported. There is a little knowledge regarding the impact of varying screw orientation and quality of reduction on the fixation strength.

Purpose

The purpose of the present study was biomechanical investigation of joint stability after SI screw fixation and its dependence on quality of reduction and screw orientation.

Methods

Thirty-two artificial hemi-pelvices were assigned to four study groups and simulated SI dislocations were fixed with two SI screws in oblique or transverse screw orientation and anatomical or non-anatomical reduction in group A (oblique/anatomical), B (transverse/anatomical), C (oblique/non-anatomical) and D (transverse/non-anatomical). Mechanical testing was performed under progressively increasing cyclic axial loading until fixation failure. SI joint movements were captured via optical motion tracking. Fixation performance was statistically evaluated at a level of significance p = 0.05.

Results

The highest cycles to failure were observed in group A (14038 ± 1057), followed by B (13909 ± 1217), D (6936 ± 1654) and C (6706 ± 1295). Groups A and B revealed significantly longer endurance than C and D (p ≤ 0.01).

Conclusions

Different screw orientations in the presented model do not influence substantially SI joint stability. However, anatomical reduction is not only mandatory to restore a malalignment, but also to increase the SI screw fixation strength and prevent fixation failures.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5

Similar content being viewed by others

References

  1. Poenaru DV, Popescu M, Anglitoiu B, Popa I, Andrei D, Birsasteanu F (2015) Emergency pelvic stabilization in patients with pelvic posttraumatic instability. Int Orthop 39(5):961–965. doi:10.1007/s00264-015-2727-5

    Article  PubMed  Google Scholar 

  2. Tile M (1988) Pelvic ring fractures: should they be fixed? J Bone Joint Surg Br 70(1):1–12

    CAS  PubMed  Google Scholar 

  3. Berber O, Amis AA, Day AC (2011) Biomechanical testing of a concept of posterior pelvic reconstruction in rotationally and vertically unstable fractures. J Bone Joint Surg Br 93(2):237–244. doi:10.1302/0301-620X.93B2.24567

    Article  CAS  PubMed  Google Scholar 

  4. Leighton RK, Waddell JP, Bray TJ, Hapman MW, Simpson L, Martin RB, Sharkey NA (1991) Biomechanical testing of new and old fixation devices for vertical shear fractures of the pelvis. J Orthop Trauma 5(3):313–317

    Article  CAS  PubMed  Google Scholar 

  5. Shuler TE, Boone DC, Gruen GS, Peitzman AB (1995) Percutaneous iliosacral screw fixation: early treatment for unstable posterior pelvic ring disruptions. J Trauma 38(3):453–458

    Article  CAS  PubMed  Google Scholar 

  6. Keating JF, Werier J, Blachut P, Broekhuyse H, Meek RN, O’Brien PJ (1999) Early fixation of the vertically unstable pelvis: the role of iliosacral screw fixation of the posterior lesion. J Orthop Trauma 13(2):107–113

    Article  CAS  PubMed  Google Scholar 

  7. van Zwienen CM, van den Bosch EW, Hoek van Dijke GA, Snijders CJ, van Vugt AB (2005) Cyclic loading of sacroiliac screws in Tile C pelvic fractures. J Trauma 58(5):1029–1034

    Article  PubMed  Google Scholar 

  8. Routt C Jr, Meier MC, Kregor PJ, Mayo KA (1993) Percutaneous iliosacral screws with the patient supine technique. Oper Tech Orthop 3(1):35–45

    Article  Google Scholar 

  9. Sagi HC, Ordway NR, DiPasquale T (2004) Biomechanical analysis of fixation for vertically unstable sacroiliac dislocations with iliosacral screws and symphyseal plating. J Orthop Trauma 18(3):138–143

    Article  CAS  PubMed  Google Scholar 

  10. Carlson DA, Scheid DK, Maar DC, Baele JR, Kaehr DM (2000) Safe placement of S1 and S2 iliosacral screws: the “vestibule” concept. J Orthop Trauma 14(4):264–269

    Article  CAS  PubMed  Google Scholar 

  11. Citak M, Hufner T, Geerling J, Kfuri M Jr, Gansslen A, Look V, Kendoff D, Krettek C (2006) Navigated percutaneous pelvic sacroiliac screw fixation: experimental comparison of accuracy between fluoroscopy and Iso-C3D navigation. Comput Aided Surg: Off J Int Soc Comput Aided Surg 11(4):209–213. doi:10.3109/10929080600890015

    CAS  Google Scholar 

  12. Bergmann G, Deuretzbacher G, Heller M, Graichen F, Rohlmann A, Strauss J, Duda GN (2001) Hip contact forces and gait patterns from routine activities. J Biomech 34(7):859–871

    Article  CAS  PubMed  Google Scholar 

  13. Gueorguiev B, Ockert B, Schwieger K, Wahnert D, Lawson-Smith M, Windolf M, Stoffel K (2011) Angular stability potentially permits fewer locking screws compared with conventional locking in intramedullary nailed distal tibia fractures: a biomechanical study. J Orthop Trauma 25(6):340–346. doi:10.1097/BOT.0b013e3182163345

    Article  PubMed  Google Scholar 

  14. Windolf M, Muths R, Braunstein V, Gueorguiev B, Hanni M, Schwieger K (2009) Quantification of cancellous bone-compaction due to DHS Blade insertion and influence upon cut-out resistance. Clin Biomech (Bristol, Avon) 24(1):53–58. doi:10.1016/j.clinbiomech.2008.09.005

    Article  Google Scholar 

  15. Gardner MJ, Morshed S, Nork SE, Ricci WM, Chip Routt ML Jr (2010) Quantification of the upper and second sacral segment safe zones in normal and dysmorphic sacra. J Orthop Trauma 24(10):622–629. doi:10.1097/BOT.0b013e3181cf0404

    Article  PubMed  Google Scholar 

  16. Moed BR, Whiting DR (2010) Locked transsacral screw fixation of bilateral injuries of the posterior pelvic ring: initial clinical series. J Orthop Trauma 24(10):616–621. doi:10.1097/BOT.0b013e3181df97eb

    Article  PubMed  Google Scholar 

  17. Beaule PE, Antoniades J, Matta JM (2006) Trans-sacral fixation for failed posterior fixation of the pelvic ring. Arch Orthop Trauma Surg 126(1):49–52. doi:10.1007/s00402-005-0069-2

    Article  PubMed  Google Scholar 

  18. Gras F, Hillmann S, Rausch S, Klos K, Hofmann GO, Marintschev I (2015) Biomorphometric analysis of ilio-sacro-iliacal corridors for an intra-osseous implant to fix posterior pelvic ring fractures. J Orthop Res: Off Publ Orthop Res Soc 33(2):254–260. doi:10.1002/jor.22754

    Article  Google Scholar 

  19. Denis F, Davis S, Comfort T (1988) Sacral fractures: an important problem. Retrospective analysis of 236 cases. Clin Orthop Relat Res 227:67–81

    CAS  PubMed  Google Scholar 

  20. Mullis BH, Sagi HC (2008) Minimum 1-year follow-up for patients with vertical shear sacroiliac joint dislocations treated with iliosacral screws: does joint ankylosis or anatomic reduction contribute to functional outcome? J Orthop Trauma 22(5):293–298. doi:10.1097/BOT.0b013e31816b6b4e

    Article  PubMed  Google Scholar 

  21. Dujardin FH, Hossenbaccus M, Duparc F, Biga N, Thomine JM (1998) Long-term functional prognosis of posterior injuries in high-energy pelvic disruption. J Orthop Trauma 12(3):145–150, discussion 150–141

    Article  CAS  PubMed  Google Scholar 

  22. McLaren AC, Rorabeck CH, Halpenny J (1990) Long-term pain and disability in relation to residual deformity after displaced pelvic ring fractures. Can J Surg J Can Chir 33(6):492–494

    CAS  Google Scholar 

  23. Nepola JV, Trenhaile SW, Miranda MA, Butterfield SL, Fredericks DC, Riemer BL (1999) Vertical shear injuries: is there a relationship between residual displacement and functional outcome? J Trauma 46(6):1024–1029, discussion 1029–1030

    Article  CAS  PubMed  Google Scholar 

  24. Heiner AD (2008) Structural properties of fourth-generation composite femurs and tibias. J Biomech 41(15):3282–3284. doi:10.1016/j.jbiomech.2008.08.013

    Article  PubMed  Google Scholar 

  25. van Zwienen CM, van den Bosch EW, Snijders CJ, Kleinrensink GJ, van Vugt AB (2004) Biomechanical comparison of sacroiliac screw techniques for unstable pelvic ring fractures. J Orthop Trauma 18(9):589–595

    Article  PubMed  Google Scholar 

  26. Clements JP, Moriaty N, Chesser TJ, Ward AJ, Cunningham JL (2008) Determination of pelvic ring stability: a new technique using a composite hemi-pelvis. Proc Inst Mech Eng H J Eng Med 222(5):611–616

    Article  CAS  Google Scholar 

Download references

Acknowledgments

The authors are not compensated and there are no other institutional subsidies, corporate affiliations, or funding sources supporting this work unless clearly documented and disclosed. This investigation was performed with the assistance of the AO Foundation via the AOTRAUMA Network (Grant No.: AR2013_10).

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Ivan Zderic.

Ethics declarations

Conflict of interest

The authors have no conflicts of interest to disclose.

Additional information

Gaston Camino Willhuber and Ivan Zderic contributed equally to this work.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Camino Willhuber, G., Zderic, I., Gras, F. et al. Analysis of sacro-iliac joint screw fixation: does quality of reduction and screw orientation influence joint stability? A biomechanical study. International Orthopaedics (SICOT) 40, 1537–1543 (2016). https://doi.org/10.1007/s00264-015-3007-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00264-015-3007-0

Keywords

Navigation