Abstract
Background and Purpose:
To compare plan quality criteria and dosimetric accuracy of step-and-shoot intensity-modulated radiotherapy (ss-IMRT) and volumetric modulated arc radiotherapy (VMAT) using two different gantry rotation times.
Patients and Methods:
This retrospective planning study based on 20 patients was comprised of 10 prostate cancer (PC) and 10 head and neck (HN) cancer cases. Each plan contained two target volumes: a primary planning target volume (PTV) and a boost volume. For each patient, one ss-IMRT plan and two VMAT plans at 90 s (VMAT90) and 120 s (VMAT120) per arc were generated with the Pinnacle© planning system. Two arcs were provided for the PTV plans and a single arc for boost volumes. Dosimetric verification of the plans was performed using a 2D ionization chamber array placed in a full scatter phantom.
Results:
VMAT reduced delivery time and monitor units for both treatment sites compared to IMRT. VMAT120 vs. VMAT90 increased delivery time and monitor units in PC plans without improving plan quality. For HN cases, VMAT120 provided comparable organs at risk sparing and better target coverage and conformity than VMAT90. In the VMAT plan verification, an average of 97.1% of the detector points passed the 3 mm, 3% γ criterion, while in IMRT verification it was 98.8%.
Conclusion:
VMAT90, VMAT120, and IMRT achieved comparable treatment plans. Slower gantry movement in VMAT120 plans only improves dosimetric quality for highly complex targets.
Zusammenfassung
Ziel:
Vergleich von Planqualität und dosimetrischer Genauigkeit von „step-and-shoot“ intensitätsmodulierter Strahlentherapie (ss-IMRT) und volumetrisch modulierter Rotationstherapie (VMAT) bei zwei unterschiedlichen Gantry-Rotationszeiten.
Patienten und Methoden:
Diese retrospektive Planungsstudie basierte auf 20 Patienten: 10 mit Prostatatumor (PC) und 10 mit Kopf-Hals-Tumor (HN). Jeder Plan enthielt 2 Zielvolumina: das primäre Planungszielvolumen (PTV) und ein Boostvolumen. Pro Patient wurden je ein ss-IMRT- und zwei VMAT-Pläne mit je 90 (VMAT90) und 120 (VMAT120) Sekunden pro Rotation im Pinnacle©-Planungssystem generiert. VMAT-Pläne für primäre Zielvolumina wurden mit 2 Vollrotationen erstellt, Pläne für Boostvolumina mit 1 Rotation. Für die dosimetrische Verifikation der Bestrahlungspläne wurden eine 2D Ionisationskammer-Matrix und ein oktagonales Festkörperphantom verwendet.
Ergebnisse:
Im Vergleich zur IMRT wurden durch VMAT-Pläne die Bestrahlungszeit und die Anzahl der Monitoreinheiten für beide Behandlungslokalisationen reduziert (Tabellen 2 und 3). VMAT120 erhöhte gegenüber VMAT90 sowohl die Bestrahlungszeit als auch die Monitoreinheiten für PC-Pläne, ohne eine Verbesserung der Planqualität zu erzielen. Für HN-Patienten konnte mit VMAT120 eine vergleichbare Risikoorganschonung sowie eine bessere Zielvolumenabdeckung als mit VMAT90 erzielt werden (siehe Tabellen 1 und 3, sowie Abbildung 2b). VMAT-Pläne erreichten einen Gamma-Index < 1 (3 mm Abstand und 3% Dosis) für 97,1% der Detektorpunkte, IMRT-Pläne erreichten 98,8% der Detektorpunkte.
Schlussfolgerung:
VMAT90, VMAT120 und IMRT erzielten vergleichbare Bestrahlungspläne. Eine langsamere Bewegung der Gantry in VMAT120-Plänen konnte die Planqualität nur für sehr komplexe Zielvolumina verbessern.
Similar content being viewed by others
References
Bertelsen A, Hansen CR, Johansen J et al. Single arc volumetric modulated arc therapy of head and neck cancer. Radiother Oncol 2010;95:142–148.
Bratengeier K. Applications of two-step intensity modulated arc therapy. Strahlenther Onkol 2001;177:394–403.
Bzdusek K, Friberger H, Eriksson K et al. Development and evaluation of an efficient approach to volumetric arc therapy planning. Med Phys 2009;36:2328–2339.
Clivio A, Fogliata A, Franzetti-Pellanda A et al. Volumetric-modulated arc radiotherapy for carcinomas of the anal canal: a treatment planning comparison with fixed field IMRT. Radiother Oncol 2009;92:118–124.
Cotrutz C, Xing L. Segment-based dose optimisation using a genetic algorithm. Phys Med Biol 2003;48:2987–2998.
Cozzi L, Dinshaw KA, Shrivastava SK et al. A treatment planning study comparing volumetric arc modulation with RapidArc and fixed field IMRT for cervix uteri radiotherapy. Radiother Oncol 2008;89:180–191.
Eppinga E, Lagerwaard F, Verbakel W et al. Volumetric modulated arc therapy for advanced pancreatic cancer. Strahlenther Onkol 2010;186:382–387.
Grégoire V, Coche E, Cosnard G et al. Selection and delineation of lymph node target volumes in head and neck conformal radiotherapy. Proposal for standardizing terminology and procedure based on the surgical experience. Radiother Oncol 2000;56:135–150.
Guckenberger M, Richter A, Krieger T et al. Is a single arc sufficient in volumetric- modulated arc therapy (VMAT) for complex-shaped target volumes? Radiother Oncol 2009;93:259–265.
Hall EJ, Wuu CS. Radiation-induced second cancers: the impact of 3D-CRT and IMRT. Int J Radiat Oncol Biol Phys 2003;56:83–88.
Hall EJ. Intensity-modulated radiation therapy, protons, and the risk of second cancers. Int J Radiat Oncol Biol Phys 2006;65:1–7.
Intensity Modulated Radiation Therapy Collaborative Working Group. Intensity-modulated radiotherapy: current status of interest. Int J Radiat Oncol Biol Phys 2001;51:880–914.
Kotte AN, Hofman P, Lagendijk JJ et al. Intrafraction motion of the prostate during external-beam radiation therapy: analysis of 427 patients with implanted fiducial markers. Int J Radiat Oncol Biol Phys 2007;69:419–425.
Lawton CA, Michalski J, El-Naqa I et al. RTOG GU Radiation oncology specialists reach consensus on pelvic lymph node volumes for high risk prostate cancer, Int J Radiat Oncol Biol Phys 2009;74:383–387.
Low DA, Harms WB, Mutic S et al. A technique for the quantitative evaluation of dose distributions. Med Phys 1998;25:656–661.
Michalski JM, Lawton C, El-Naqa I et al. Development of RTOG consensus guidelines for the definition of the clinical target volume for postoperative conformal radiation therapy for prostate cancer. Int J Radiat Oncol Biol Phys 2010;76:361–368.
Moret J, Kölbl O, Bogner L. Quasi-IMAT Study with conventional equipment to show high plan quality with a single gantry arc. Strahlenther Onkol 2009;185:41–48.
Otto K. Volumetric modulated arc therapy: IMRT in a single gantry arc. Med Phys 2008;35:310–317.
Palma D, Vollans E, James K et al. Volumetric modulated arc therapy for delivery of prostate radiotherapy: comparison with intensity-modulated radiotherapy and three-dimensional conformal radiotherapy. Int J Radiat Oncol Biol Phys 2008;72:996–1001.
Shaffer R, Nichol A, Vollans E et al. A comparison of volumetric modulated arc therapy and conventional intensity-modulated radiotherapy for frontal and temporal high-grade gliomas. Int J Radiat Oncol Biol Phys 2010;76:1177–1184.
Shepard DM, Earl MA, Li XA et al. Direct aperture optimisation: a turnkey solution for step-and-shoot IMRT. Med Phys 2002;29:1007–1018.
Van Esch A, Clermont C, Devillers M et al. On-line quality assurance of rotational radiotherapy treatment delivery by means of a 2D ion chamber array and the Octavius phantom. Med Phys 2007;34:3825–3837.
Vanetti E, Clivio A, Nicolini G et al. Volumetric modulated arc radiotherapy for carcinomas of the oro-pharynx, hypo-pharynx and larynx: a treatment planning comparison with fixed field IMRT. Radiother Oncol 2009;92:111–117.
Vandecasteele K, De Neve W, De Gersem W et al. Intensity-modulated arc therapy with simultaneous integrated boost in the treatment of primary irresectable cervical cancer. Strahlenther Onkol 2009;185:799–807.
Verbakel WF, Cuijpers JP, Hoffmans D et al. Volumetric intensity-modulated arc therapy vs. conventional IMRT in head-and-neck cancer: a comparative planning and dosimetric study. Int J Radiat Oncol Biol Phys 2009;74:252–259.
Wiezorek T., Banz N, Schwedas M et al. Dosimetric quality assurance for intensity-modulated radiotherapy. Feasibility study for a filmless approach. Strahlenther Onkol 2005;181:468–474.
Wolff D, Stieler F, Welzel G et al. Volumetric modulated arc therapy (VMAT) vs. serial tomotherapy, step-and-shoot IMRT and 3D-conformal RT for treatment of prostate cancer. Radiother Oncol 2009;93:226–233.
Wolff D, Stieler F, Hermann B et al. Clinical Implementation of Volumetric Intensity-modulated arc therapy (VMAT) with ERGO++. Strahlenther Onkol 2010;186:280–288.
Yu CX. Intensity-modulated arc therapy with dynamic multileaf collimation: an alternative to tomotherapy. Phys Med Biol 1995;40:1435–1449.
Zhang P, Happersett L, Hunt M et al. Volumetric modulated arc therapy: Planning and evaluation for prostate cancer cases. Int J Radiat Oncol Biol Phys 2010;76:1456–1462.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Pasler, M., Wirtz, H. & Lutterbach, J. Impact of Gantry Rotation Time on Plan Quality and Dosimetric Verification – Volumetric Modulated Arc Therapy (VMAT) vs. Intensity Modulated Radiotherapy (IMRT). Strahlenther Onkol 187, 812–819 (2011). https://doi.org/10.1007/s00066-011-2263-1
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00066-011-2263-1