Development of a method for the estimation of low amount of plutonium in 200 L waste drums in presence of high amount of β γ activity due to 137Cs and 60Co

Article

DOI: 10.1007/s10967-012-2322-9

Cite this article as:
Singh, S., Sagar, V. & Mhatre, A. J Radioanal Nucl Chem (2013) 297: 149. doi:10.1007/s10967-012-2322-9
  • 125 Downloads

Abstract

The waste drum monitoring system based on HPGe detector was used to study its performance for the estimation of low amounts of plutonium in presence of high activity of 137Cs and 60Co. The counting was carried out by keeping amount of plutonium constant at 100 mg level and varying the count rate for the γ rays of 137Cs and 60Co. Present study has shown that the estimation of low amount of 239Pu in a waste drum can be carried out using 129 keV γ ray in the presence of 137Cs up to an activity level of 16 mCi and in the presence of 60Co up to an activity level of 8 mCi.

Keywords

Waste drum monitoring γ Ray scanning system Plutonium 239Pu 240Pu 241Pu α Active waste HPGe detector 60Co 137Cs Dead time correction 