Evaluation of a new generation quadrupole-based ICP-MS for uranium isotopic measurements in environmental samples

Mass Spectrometric Methods for the Determination of Ultra Trace Actinides

DOI: 10.1007/s10967-008-0534-9

Cite this article as:
Pointurier, F., Hubert, A., Baglan, N. et al. J Radioanal Nucl Chem (2008) 276: 505. doi:10.1007/s10967-008-0534-9


We compare the analytical performance of a modern quadrupole-based ICP-MS (“X-Series”, Thermo-Electron, Winsford, UK) with a single-collector double-focusing sector-field ICP-MS (“Axiom”, VG Elemental, Winsford, UK) for uranium isotopic measurements in environmental samples. We focus on the precision and accuracy obtained with both instruments for the 235U/238U isotopic ratios and on the abundance sensitivity that is a key parameter for low 236U/238U isotopic ratios measurements. We observe that isotopic measurements are more precise accurate with the “X-Series” than with the “Axiom”. Besides, we demonstrate that the “Axiom’s” higher abundance sensitivity limits its capability to measure 236U/238U ratio below a few ppm.