European Journal of Nuclear Medicine and Molecular Imaging

, Volume 40, Issue 10, pp 1507–1515

Evaluation of strategies towards harmonization of FDG PET/CT studies in multicentre trials: comparison of scanner validation phantoms and data analysis procedures

  • Nikolaos E. Makris
  • Marc C. Huisman
  • Paul E. Kinahan
  • Adriaan A. Lammertsma
  • Ronald Boellaard
Original Article

DOI: 10.1007/s00259-013-2465-0

Cite this article as:
Makris, N.E., Huisman, M.C., Kinahan, P.E. et al. Eur J Nucl Med Mol Imaging (2013) 40: 1507. doi:10.1007/s00259-013-2465-0

Abstract

Purpose

PET quantification based on standardized uptake values (SUV) is hampered by several factors, in particular by variability in PET acquisition settings and data analysis methods. Quantitative PET/CT studies acquired during a multicentre trial require harmonization of imaging procedures to maximize study power. The aims of this study were to determine which phantoms are most suitable for detecting differences in image quality and quantification, and which methods for defining volumes of interest (VOI) are least sensitive to these differences.

Methods

The most common accreditation phantoms used in oncology FDG PET/CT trials were scanned on the same scanner. These phantoms were those used by the Society of Nuclear Medicine Clinical Trials Network (SNM-CTN), the European Association of Nuclear Medicine/National Electrical Manufacturers Association (EANM/NEMA) and the American College of Radiology (ACR). In addition, tumour SUVs were derived from ten oncology whole-body examinations performed on the same PET/CT system. Both phantom and clinical data were reconstructed using different numbers of iterations, subsets and time-of-flight kernel widths. Subsequently, different VOI methods (VOIA50%, VOImax, VOI3Dpeak, VOI2Dpeak) were applied to assess the impact of changes in image reconstruction settings on SUV and recovery coefficients (RC).

Results

All phantoms demonstrated sensitivity for detecting changes in SUV and RC measures in response to changes in image reconstruction settings and VOI analysis methods. The SNM-CTN and EANM/NEMA phantoms showed almost equal sensitivity in detecting RC differences with changes in image characteristics. Phantom and clinical data demonstrated that the VOI analysis methods VOIA50% and VOImax gave SUV and RC values with large variability in relation to image characteristics, whereas VOI3Dpeak and VOI2Dpeak were less sensitive to these differences.

Conclusion

All three phantoms may be used to harmonize parameters for data acquisition, processing and analysis. However, the SNM-CTN and EANM/NEMA phantoms are the most sensitive to parameter changes and are suitable for harmonizing SUV quantification based on 3D VOIs, such as VOIA50% and VOI3Dpeak, and VOImax. Variability in SUV quantification after harmonization could be further minimized using VOI3Dpeak analysis, which was least sensitive to residual variability in image quality and quantification.

Keywords

PET/CTQuantificationStandardizationHarmonizationAccreditation

Copyright information

© Springer-Verlag Berlin Heidelberg 2013

Authors and Affiliations

  • Nikolaos E. Makris
    • 1
  • Marc C. Huisman
    • 1
  • Paul E. Kinahan
    • 2
  • Adriaan A. Lammertsma
    • 1
  • Ronald Boellaard
    • 1
  1. 1.Department of Radiology & Nuclear MedicineVU University Medical CentreAmsterdamThe Netherlands
  2. 2.Imaging Research Laboratory, Department of RadiologyUniversity of WashingtonSeattleUSA