Abstract
Purpose
Metallic dental implants cause severe streaking artifacts in computed tomography (CT) data, which affect the accuracy of dose calculations in radiation therapy. The aim of this study was to investigate the benefit of the metal artifact reduction algorithm iterative metal artifact reduction (iMAR) in terms of correct representation of Hounsfield units (HU) and dose calculation accuracy.
Materials and methods
Heterogeneous phantoms consisting of different types of tissue equivalent material surrounding metallic dental implants were designed. Artifact-containing CT data of the phantoms were corrected using iMAR. Corrected and uncorrected CT data were compared to synthetic CT data to evaluate accuracy of HU reproduction. Intensity-modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT) plans were calculated in Oncentra v4.3 on corrected and uncorrected CT data and compared to Gafchromic™ EBT3 films to assess accuracy of dose calculation.
Results
The use of iMAR increased the accuracy of HU reproduction. The average deviation of HU decreased from 1006 HU to 408 HU in areas including metal and from 283 HU to 33 HU in tissue areas excluding metal. Dose calculation accuracy could be significantly improved for all phantoms and plans: The mean passing rate for gamma evaluation with 3 % dose tolerance and 3 mm distance to agreement increased from 90.6 % to 96.2 % if artifacts were corrected by iMAR.
Conclusion
The application of iMAR allows metal artifacts to be removed to a great extent which leads to a significant increase in dose calculation accuracy.
Zusammenfassung
Zielsetzung
Metallische Implantate verursachen streifenförmige Artefakte in CT-Bildern, welche die Dosisberechnung beeinflussen. In dieser Studie soll der Nutzen des iterativen Metall-Artefakt-Reduktions-Algorithmus iMAR hinsichtlich der Wiedergabetreue von Hounsfield-Werten (HU) und der Genauigkeit von Dosisberechnungen untersucht werden.
Material und Methoden
Es wurden heterogene Phantome aus verschiedenen Arten gewebeäquivalenten Materials mit Zahnimplantaten entworfen. Von den Phantomen wurden CT-Scans angefertigt und mittels des iMAR-Algorithmus korrigiert. Um die Wiedergabetreue der Hounsfield-Einheiten zu testen, wurden die korrigierten und nichtkorrigierten CT-Daten gegen synthetische CT-Daten verglichen. IMRT- und VMAT-Pläne wurden in Oncentra v4.3 berechnet und zur Beurteilung der Genauigkeit der Dosisberechnung mit Gafchromic™-EBT3-Filmmessungen verglichen.
Ergebnisse
Die Anwendung von iMAR erhöht die Wiedergabetreue der Hounsfield-Einheiten. Die durchschnittliche Abweichung zu den synthetischen Daten fällt in Gewebebereichen ohne Metall von 283 HU auf 33 HU und in Metallbereichen von 1006 HU auf 408 HU. Die Genauigkeit der Dosisberechnungen konnte für alle Phantome und Pläne signifikant verbessert werden. Die mittlere Akzeptanzrate der Gammaevaluation mit 3 % Dosistoleranz und 3 mm Ortstoleranz steigt von 90,6 auf 96,2%, wenn Artefakte mittels iMAR korrigiert werden.
Schlussfolgerung
Mit Hilfe von iMAR konnten die Artefakte zu großen Teilen entfernt werden, was zu einer signifikanten Erhöhung der Dosisberechnungsgenauigkeit führt.
Similar content being viewed by others
References
Maerz M, Koelbl O, Dobler B (2015) Influence of metallic dental implants and metal artefacts on dose calculation accuracy. Strahlenther Onkol 191:234–241
Glide-Hurst C, Chen D, Zhong H et al (2013) Changes realized from extended bit-depth and metal artifact reduction in CT. Med Phys 40:061711
Li H, Noel C, Chen H et al (2012) Clinical evaluation of a commercial orthopedic metal artifact reduction tool for CT simulations in radiation therapy. Med Phys 39:7507–7517
Mail N, Albarakati Y, Ahmad Khan M et al (2013) The impacts of dental filling materials on RapidArc treatment planning and dose delivery: challenges and solution. Med Phys 40:081714
Paudel MR, Mackenzie M, Fallone BG et al (2013) Evaluation of normalized metal artifact reduction (NMAR) in kVCT using MVCT prior images for radiotherapy treatment planning. Med Phys 40:081701
Paudel MR, Mackenzie M, Fallone BG et al (2014) Clinical evaluation of normalized metal artifact reduction in kVCT using MVCT prior images (MVCT-NMAR) for radiation therapy treatment planning. Int J Radiat Oncol Biol Phys 89:682–689
Spadea MF, Verburg J, Baroni G et al (2013) Dosimetric assessment of a novel metal artifact reduction method in CT images. J Appl Clin Med Phys 14:4027
Spadea MF, Verburg JM, Baroni G et al (2014) The impact of low-Z and high-Z metal implants in IMRT: a Monte Carlo study of dose inaccuracies in commercial dose algorithms. Med Phys 41:011702
Axente M, Paidi A, Von Eyben R et al (2015) Clinical evaluation of the iterative metal artifact reduction algorithm for CT simulation in radiotherapy. Med Phys 42:1170–1183
Morsbach F, Bickelhaupt S, Wanner GA et al (2013) Reduction of metal artifacts from hip prostheses on CT images of the pelvis: value of iterative reconstructions. Radiology 268:237–244
Morsbach F, Wurnig M, Kunz DM et al (2013) Metal artefact reduction from dental hardware in carotid CT angiography using iterative reconstructions. Eur Radiol 23:2687–2694
Subhas N, Primak AN, Obuchowski NA et al (2014) Iterative metal artifact reduction: evaluation and optimization of technique. Skeletal Radiol 43:1729–1735
Yohannes I, Kolditz D, Langner O et al (2012) A formulation of tissue- and water-equivalent materials using the stoichiometric analysis method for CT-number calibration in radiotherapy treatment planning. Phys Med Biol 57:1173–1190
Meyer E, Raupach R, Lell M et al (2010) Normalized metal artifact reduction (NMAR) in computed tomography. Med Phys 37:5482–5493
Meyer E, Raupach R, Lell M et al (2012) Frequency split metal artifact reduction (FSMAR) in computed tomography. Med Phys 39:1904–1916
Alvarez-Moret J, Pohl F, Koelbl O et al (2010) Evaluation of volumetric modulated arc therapy (VMAT) with Oncentra MasterPlan(R) for the treatment of head and neck cancer. Radiat Oncol 5:110
Dobler B, Pohl F, Bogner L et al (2007) Comparison of direct machine parameter optimization versus fluence optimization with sequential sequencing in IMRT of hypopharyngeal carcinoma. Radiat Oncol 2:33
Dobler B, Weidner K, Koelbl O (2010) Application of volumetric modulated arc therapy (VMAT) in a dual-vendor environment. Radiat Oncol 5:95
International Commission on Radiological Protection (1975) Report of the task group on reference man. ICRP 23
International Commission on Radiation Units and Measurements (1989) Tissue substitutes in radiation dosimetry and measurement. ICRU 44
Crijns W, Maes F, Van Der Heide UA et al (2013) Calibrating page sized Gafchromic EBT3 films. Med Phys 40:012102
Siebers JV, Keall PJ, Nahum AE et al (2000) Converting absorbed dose to medium to absorbed dose to water for Monte Carlo based photon beam dose calculations. Phys Med Biol 45:983–995
Low DA, Harms WB, Mutic S et al (1998) A technique for the quantitative evaluation of dose distributions. Med Phys 25:656–661
Ezzell GA, Burmeister JW, Dogan N et al (2009) IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119. Med Phys 36:5359–5373
Hayashi N, Watanabe Y, Malmin R et al (2012) Evaluation of triple channel correction acquisition method for radiochromic film dosimetry. J Radiat Res 53:930–935
Micke A, Lewis DF, Yu X (2011) Multichannel film dosimetry with nonuniformity correction. Med Phys 38:2523–2534
Acknowledgments
The work was supported by the Wilhelm Sander Foundation. We want to thank Walter Schäffer for providing the teeth implant material. We want to thank Elmar Lang for support in applying for third-party funding.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The department of M. Maerz, P. Mittermair, O. Koelbl and B. Dobler has a research cooperation with Elekta GmbH Hamburg.
M. Maerz, P. Mittermair, O. Koelbl and B. Dobler state that there are no conflicts of interest. A. Krauss is employee of Siemens Healthcare GmbH.
The accompanying manuscript does not include studies on humans or animals.
Rights and permissions
About this article
Cite this article
Maerz, M., Mittermair, P., Krauss, A. et al. Iterative metal artifact reduction improves dose calculation accuracy. Strahlenther Onkol 192, 403–413 (2016). https://doi.org/10.1007/s00066-016-0958-z
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00066-016-0958-z
Keywords
- Metal artifact reduction
- Metal artifacts
- MAR
- Dental implants
- Dose calculation
- Gafchromic
- Radiation therapy