Prediction of LC retention of steroids using solvatochromic parameters

Authors

  • D. Barrón
    • Department of Analytical ChemistryUniversity of Barcelona
  • J. A. Pascual
    • Department of Pharmacology and Toxicology, Institut Municipal d’Investigació Médica (IMIM)Autonomous University of Barcelona (U.A.B.)
  • J. Segura
    • Department of Pharmacology and Toxicology, Institut Municipal d’Investigació Médica (IMIM)Autonomous University of Barcelona (U.A.B.)
  • J. Barbosa
    • Department of Analytical ChemistryUniversity of Barcelona
Originals

DOI: 10.1007/BF02688088

Cite this article as:
Barrón, D., Pascual, J.A., Segura, J. et al. Chromatographia (1995) 41: 573. doi:10.1007/BF02688088

Summary

The proportion of organic modifier and pH of the hydroorganic mobile phase were optimized to separate a series of nine important steroids by LC. It has been shown that linear plots of logk′ against Reichardt’sETN parameter for the mobile phase is observed over the whole range of compositions of practical importance. Data from these plots summarize an approach for optimization of resolution from only two retention measurements for each compound. pH is shown to be irrelevant. This behaviour could be extrapolated to other steroids and drastically reduce the number of retention values needed for mobile phase optimization for screening analysis of steroids in doping investigations.

Key Words

Column liquid chromatographyMobile phase optimizationSteroidsSolvatochromic parametersRetention prediction