Abstract
New miniaturized portable electrocardiogram (ECG) measuring devices may require small interelectrode distance. However, finding a suitable location for a tiny measurement device may prove tedious, as reducing interelectrode distance reduces signal strength. The objective of the study was to define the optimal location for a very closely located (5 cm) bipolar electrode pair. A total of 120 bipolar leads were analyzed from a body surface potential map (BSPM) data with 236 subjects with a normal ECG. The average and standard deviation (SD) of the QRS-complex and the P-wave amplitudes in each electrode location and for each subject were determined. The results showed that deviation in signal amplitude between different subjects is significant. However, judging from average values, the best orientation for a closely located bipolar electrode pair is diagonally on the chest. The best locations for QRS-complex and P-wave detection are around the chest electrodes of the standard precordial leads V2, V3, and V4, and above the chest electrodes of leads V1 and V2, respectively.
Similar content being viewed by others
References
Alive Technologies. http://www.alivetec.com/products.htm. Accessed on 13.11.2007.
Donnelly M. P., D. D. Finlay, C. D. Nugent, and N. D. Black. 2008 Lead selection: old and new methods for locating the most electrocardiogram information. Journal of Electrocardiology 41:257–263. doi:10.1016/j.jelectrocard.2008.02.004
Dunn O. J. and V. A. Clark 1974 Applied statistics: analysis of variance and regression. John Wiley & Sons, Inc., New York
Fernandez M. and R. Pallas-Areny. 2000 Ag-AgCl electrode noise in high-resolution ECG measurements. Biomedical Instrumentation & Technology 34:125–130.
Gondran C., E. Siebert, S. Yacoub, and E. Novakov. 1996 Noise of surface bio-potential electrodes based on NASICON ceramic and Ag-AgCl. Medical & Biological Engineering & Computing 34:460–466. doi:10.1007/BF02523851
Huigen E., A. Peper, and C. A. Grimbergen. 2002 Investigation into the origin of the noise of surface electrodes. Medical & Biological Engineering & Computing 40:332–338. doi:10.1007/BF02344216
Kornreich F., T. J. Montague, M. Kavadis, P. Rautaharju, M. B. Horácek, and B. Taccardi. 1989 Multigroup diagnosis of body surface potential maps. Journal of Electrocardiology 22(Suppl):169–178. doi:10.1016/S0022-0736(07)80119-1
Lux R. L. and R. Greg. 2004 New leads for P-wave detection and arrhytmia classification. Journal of Electrocardiology 37(Suppl):80. doi:10.1016/j.jelectrocard.2004.08.029
Puurtinen, M., J. Hyttinen, and J. Malmivuo. Optimizing bipolar electrode location for wireless ECG measurement—analysis of ECG signal strength and deviation between individuals. In: Proceedings of the IEEE EMBS, Minneapolis, USA, 2005, pp. 236–239.
Riistama, J., J. Väisänen, S. Heinisuo, J. Lekkala, and J. Kaihilahti. Evaluation of an implantable ECG monitoring device in vitro and in vivo. In: Proceedings of the IEEE EMBS, Lyon, France, 2007, pp. 5703–5706.
Ryynänen O., J. Hyttinen, P. Laarne, and J. Malmivuo. 2004 Effect of measurement noise on the spatial resolution of EEG. Biomedizinische Technik 48:94–97.
Acknowledgments
The authors would like to express their gratitude to Professor Friedrich Kornreich, MD, for providing the valuable clinical data. The authors are equally indebted to Mr. Robert MacGilleon for proofreading the manuscript.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Puurtinen, M., Viik, J. & Hyttinen, J. Best Electrode Locations for a Small Bipolar ECG Device: Signal Strength Analysis of Clinical Data. Ann Biomed Eng 37, 331–336 (2009). https://doi.org/10.1007/s10439-008-9604-y
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10439-008-9604-y