Abstract
Despite continual improvements in ventricular assist device (VAD) therapy, various clinical issues are emerging. Importantly, various types of thromboembolic complications have been reported to date. Recently, we encountered a rare continuous-flow VAD-related thromboembolic event that resulted in acute myocardial infarction. A 26-year-old female who just underwent HeartMate II® VAD implantation suddenly developed widespread anterolateral myocardial infarction on postoperative day 16. Echocardiography and aortography revealed a large thrombus on the left coronary cusp of the aortic valve that almost completely occluded the left coronary ostium. After VAD implantation, her aortic valve did not open, even at relatively low pump speeds; this was thought to be one of the causes for thrombus formation. Continuous suction of blood from the left ventricle and non-pulsatile flow into the ascending aorta resulted in a continuously closed aortic valve and stagnation of blood in the coronary cusp. Furthermore, both small body size (body surface area <1.3 m2) and postoperative right ventricular failure may have exacerbated blood stagnation and thrombus formation in this patient. We should have adjusted the anticoagulation and antiplatelet therapy protocols based on the patient’s condition. She underwent off-pump coronary artery bypass surgery and remained in clinically stable condition afterwards.
Similar content being viewed by others
References
Pagani FD, Miller LW, Russell SD, Aaronson KD, John R, Boyle AJ, Conte JV, Bogaev RC, MacGillivray TE, Naka Y, Mancini D, Massey HT, Chen L, Klodell CT, Aranda JM, Moazami N, Ewald GA, Farrar DJ, Frazier OH. HeartMate II investigators. Extended mechanical circulatory support with a continuous-flow rotary left ventricular assist device. J Am Coll Cardiol. 2009;54:312–21.
Slaughter, Rogers LG, Milano CA, Russell SD, Conte JV, Feldman D, Sun B, Tatooles AJ, Delgado RM 3rd, Long JW, Wozniak TC, Ghumman W, Farrar DJ, Frazier OH. HeartMate II investigators. Advanced heart failure treated with continuous-flow left ventricular assist device. N Eng J Med. 2009;361:2241–51.
John R, Kamdar F, Liao K, Colvin-Adams M, Miller L, Joyce L, Boyle A. Low thromboembolic risk for patients with the Heartmate II left ventricular assist device. J Thorac Cardiovasc Surg. 2008;136:1318–23.
Boyle AJ, Russel SD, Teuteberg JJ, Slaughter MS, Moazami N, Pagani FD, Frazier OH, Heatley G, Farrar DJ, John R. Low thromboembolism and pump thrombosis with the Heartmate II left ventricular assist device: analysis of outpatient anti-coagulation. J Heart Lung Transplant. 2009;28:881–7.
Kirklin JK, Naftel DC, Kormos RL, Stevenson LW, Pagani FD, Miller MA, Baldwin JT, Young JB. Fifth INTERMACS annual report: risk factor analysis from more than 6000 mechanical circulatory support patients. J Heart Lung Transplant. 2013;32:141–56.
Freed BH, Jeevanandam V, Jolly N. Aortic root and valve thrombosis after implantation of a left ventricular assist device. J Invasive Cardiol. 2011;23:E63–5.
Crow S, Chen D, Milano C, Thomas W, Joyce L, Piacentino V, Sharma R, Wu J, Arepally G, Bowles D, Rogers J, Villamizar-Ortiz N. Acquired von Willebrand syndrome in continuous-flow ventricular assist device recipients. Ann Thorac Surg. 2010;90:1263–9.
Slaughter MS, Naka Y, John R, Boyle A, Conte JV, Russel SD, Aaronson KD, Sundareswaran KS, Farrar DJ, Pagani FD. Post-operative heparin may not be required for transitioning patients with HeartMate II left ventricular assist system to long-term warfarin therapy. J Heart Lung Transplant. 2010;29:616–24.
Demirozu ZT, Frazier OH. Aortic valve noncoronary cusp thrombosis after implantation of a nonpulsatile, continuous-flow pump. Tex Heart Inst J. 2012;39:618–20.
Fujita T, Kobayashi J, Hata H, Seguchi O, Sato T, Yanase M, Murata Y, Sunami H, Nakatani S. Aortic valve closure for rapidly deteriorated aortic insufficiency after continuous flow left ventricular assist device implantation. J Artif Organs. 2013;16:98–100.
Conflict of interest
The authors declare that they have no conflicts of interest.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Nakajima, S., Seguchi, O., Murata, Y. et al. Left coronary artery occlusion caused by a large thrombus on the left coronary cusp in a patient with a continuous-flow ventricular assist device. J Artif Organs 17, 197–201 (2014). https://doi.org/10.1007/s10047-014-0758-0
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10047-014-0758-0