Summary
OBJECTIVE: To evaluate clinical images from a prototype ultrahigh resolution (UHR) combined coronal optical coherence tomography/confocal scanning ophthalmoscope (OCT/SLO) and to compare them to standard-resolution OCT/SLO images on the same patients. DESIGN: Cross-sectional pilot-study. PARTICIPANTS: Sixty-six eyes of 42 patients with various macular pathologies, such as age-related macular degeneration, macular edema, macular hole, central serous retinopathy, epiretinal membrane and posterior vitreous traction syndrome. METHODS: Each subject was first scanned with a standard-resolution OCT/SLO that has an axial resolution of ∼10 micron. Immediately following, patients were scanned with the prototype UHR OCT/SLO device. The UHR system employs a compact super luminescent diode (SLD) with a 150 nm bandwidth centered at 890 nm, which allows imaging of the retina with an axial resolution of 3 microns. Both coronal and longitudinal OCT scans were acquired with each system, and compared side-by-side. Scan quality was assessed for the observer's ability to visualize the vitreo-retinal interface and retinal layers – in particular of the outer retina/RPE/choroidal interface, increased discrimination of pathological changes, and better signal intensity. MAIN OUTCOME MEASURES: Ultrahigh and standard-resolution coronal and longitudinal OCT/SLO images of macular pathologies. RESULTS: In the side-by-side comparison with the commercial standard-resolution OCT/SLO images, the scans in the Ultrahigh resolution OCT/SLO images were superior in 85% of cases. Relatively poor quality images were attributed to lower signal-to-noise ratio, limited focusing, or media opacities. Several images that had a better signal intensity in the standard-resolution OCT/SLO system were found to show more retinal detail in the UHR system. In general, intraretinal layers in the UHR OCT/SLO images were better delineated in both coronal and longitudinal scans. Enhanced details were also seen in the outer retina/RPE/choroidal complex. The UHR OCT/SLO system produced better definition of morphological changes in several macular pathologies. CONCLUSIONS: Broadband SLD-based UHR OCT/SLO offers a compact, efficient, and economic enhancement to the currently available clinical OCT imaging systems. UHR OCT/SLO imaging enhanced the quality of the OCT C-scans, facilitated appreciation of vitreo-retinal pathologies, and improved sensitivity to small changes in the retina, and the outer retina/RPE/choroidal interface.
Similar content being viewed by others
References
Huang D, Swanson EA, Lin CP, et al (1991) Optical coherence tomography. Science 254: 1178–1181
Podoleanu AG, Dobre GM, Seeger M, et al (1998) Low coherence interferometry for en-face imaging of the retina. Lasers and Light 8: 187–192
Hee MR, Izatt JA, Swanson EA, et al (1995) Optical coherence tomography of the human retina. Arch Ophthalmol 113: 325–332
Schuman JS, Puliafito CA, Fujimoto JG (2004) Optical Coherence Tomography of Ocular Diseases (2nd edn). Thorofare (USA): SLACK Inc.
Drexler W, Morgner U, Ghanta RK, et al (2001) Ultrahigh-resolution ophthalmic optical coherence tomography. Nature Medicine 7: 502–507
Drexler W, Sattmann H, Hermann B, et al (2003) Enhanced visualization of macular pathology with the use of ultrahigh-resolution optical coherence tomography. Arch Ophthalmol 121: 695–706
Nassif N, Cense B, Park BH, et al (2004) In vivo human retinal imaging by ultrahigh-speed spectral domain optical coherence tomography. Opt Lett 29: 480–482
Leitgeb RA, Drexler W, Unterhuber A, et al (2004) Ultrahigh resolution Fourier domain optical coherence tomography. Optics Express 12: 2156–2165
Schmidt-Erfurth U, Leitgeb RA, Michels S, et al (2005) Three-dimensional ultrahigh-resolution optical coherence tomography of macular diseases. Invest Ophthalmol Vis Sci 46: 3393–3402
Podoleanu AG, Seeger M, Dobre GM, et al (1998) Transversal and longitudinal images from the retina of the living eye using low coherence reflectometry. J Biomed Opt 3: 12–20
Podoleanu AG, Dobre GM, Cucu RG, et al (2004) Combined multiplanar optical coherence tomography and confocal scanning ophthalmoscopy. J Biomed Opt 9: 86–93
Hitzenberger CK, Trost P, Lo P, Zhou Q (2003) Three-dimensional imaging of the human retina by high-speed optical coherence tomography. Opt Express 11: 2753–2761
Vabre L, Dubois A, Boccara AC (2002) Thermal-light full-field optical coherence tomography. Opt Lett 27: 530–532
van Velthoven ME, Verbraak FD, Yannuzzi LA, et al (2006) Imaging the Retina by en-face Optical Coherence Tomography. Retina 26: 129–136
Unterhuber A, Povazay B, Bizheva K, et al (2004) Advances in broad bandwidth light sources for ultrahigh resolution optical coherence tomography. Phys Med Biol 49: 1235–1246
Ko TH, Adler DC, Fujimoto JG, et al (2004) Ultrahigh resolution optical coherence tomography imaging with a broadband superluminescent diode light source. Optics Express 12: 2112–2119
American National Standards Institute. American National Standard for Safe Use of Lasers. ANSI, Z 136-I. 2000. New York. Ref Type: Generic
Sander B, Larsen M, Thrane L, et al (2005) Enhanced optical coherence tomography imaging by multiple scan averaging. Br J Ophthalmol 89: 207–212
Paunescu LA, Ko TH, Duker JS, et al (2006) Idiopathic juxtafoveal retinal telangiectasis: new findings by ultrahigh-resolution optical coherence tomography. Ophthalmology 113: 48–57
Ergun E, Hermann B, Wirtitsch M, et al (2005) Assessment of central visual function in Stargardt's disease/fundus flavimaculatus with ultrahigh-resolution optical coherence tomography. Invest Ophthalmol Vis Sci 46: 310–316
Ko TH, Fujimoto JG, Duker JS, et al (2004) Comparison of ultrahigh- and standard-resolution optical coherence tomography for imaging macular hole pathology and repair. Ophthalmology 111: 2033–2043
Ko TH, Fujimoto JG, Schuman JS, et al (2005) Comparison of ultrahigh- and standard-resolution optical coherence tomography for imaging macular pathology. Ophthalmology 112: 1922
Wirtitsch MG, Ergun E, Hermann B, et al (2005) Ultrahigh resolution optical coherence tomography in macular dystrophy. Am J Ophthalmol 140: 976–983
Witkin AJ, Duker JS, Ko TH, et al (2005) Ultrahigh resolution optical coherence tomography of birdshot retinochoroidopathy. Br J Ophthalmol 89: 1660–1661
van Velthoven ME, de Vos K, Verbraak FD, et al (2005) Overlay of conventional angiographic and en-face OCT images enhances their interpretation. BMC Ophthalmol 5: 12
van Velthoven ME, Verbraak FD, Garcia PM, et al (2005) Evaluation of central serous retinopathy with en face optical coherence tomography. Br J Ophthalmol 89: 1483–1488
Pircher M, Gotzinger E, Leitgeb R, et al (2004) Imaging of polarization properties of human retina in vivo with phase resolved transversal PS-OCT. Optics Express 12: 5940–5951
Paunescu LA, Schuman JS, Price LL, et al (2004) Reproducibility of Nerve Fiber Thickness, Macular Thickness, and Optic Nerve Head Measurements Using StratusOCT. Invest Ophthalmol Vis Sci 45: 1716–1724
Gurses-Ozden R, Teng C, Vessani R, et al (2004) Macular and Retinal Nerve Fiber Layer Thickness Measurement Reproducibility Using Optical Coherence Tomography (OCT-3). J Glaucoma 13: 238–244
Hee MR (2005) Artifacts in optical coherence tomography topographic maps. Am J Ophthalmol 139: 154–155
Cucu RG, Podoleanu AG, Rogers JA, Pedro J, Rosen RB (2006) Combined confocal/en face T-scan-based ultrahigh-resolution optical coherence tomography in vivo retinal imaging. Opt Lett 31 (11): 1684–1686
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Rosen, R., van Velthoven, M., Garcia, P. et al. Ultrahigh-Resolution Combined Coronal Optical Coherence Tomography Confocal Scanning Ophthalmoscope (OCT/SLO): A pilot study. Spektrum Augenheilkd. 21, 17–28 (2007). https://doi.org/10.1007/s00717-007-0182-4
Issue Date: