Abstract
Few studies on treatment with inhaled nitric oxide (NOi) have been carried out in small laboratory animals yet, since commercially available dosing devices are not appropriate in this setting for technical or financial reasons. The aim of our study was to establish and validate a simple, cost-effective system for the application of NOi in small animals. The system mixes NOi with constant-flow inspiratory gas. A gas blender allows for a mixture of nitrogen, oxygen, and NO dissolved in nitrogen. A formula using the desired inspiratory oxygen fraction and the desired concentration of NOi as independent variables derives a somewhat higher inspiratory oxygen fraction, which is preset using an oximeter. Then the flow of NO in nitrogen is started, lowering the inspiratory oxygen fraction to the initially desired value, thereby adding NOi in the desired concentration. The method was validated by 153 adjustments, covering a variety of oxygen fractions and concentrations of NOi. NOi was measured by chemiluminescence as reference method. A close correlation (R=0.994) was found, and the regression line was close to the line of identity with y=−0.0994+1.048x. No systematic errors could be identified. We conclude that the method described may serve as a simple, cost-effective way to administer NOi to small animals.
Article PDF
Similar content being viewed by others
References
Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet I: 307–310
Imanaka H, Hess D, Kirmse M, Bigatello LM, Kacmarek RM, Steudel W, Hurford WE (1997) Inaccuracies of nitric oxide delivery systems during adult mechanical ventilation. Anesthesiology 86: 676–688
Lindberg L, Rydgren G, Larsson A, Olsson SG, Nordstrom L (1997) A delivery system for inhalation of nitric oxide evaluated with chemiluminescence, electrochemical fuel cells, and capnography. Crit Care Med 25: 190–196
Luhr O, NathorstWestfelt U, Lundin S, Wickerts CJ, Stiernstrom H, Berggren L, Aardal S, Johansson LA, Stenqvist O, Rudberg U, Lindh A, Bindslev L, Martling CR, Hornbaek V, Frostell C (1997) A retrospective analysis of nitric oxide inhalation in patients with severe acute lung injury in Sweden and Norway 1991–1994. Acta Anaesthesiol Scand 41: 1238–1246
Moutafis M, Hatahet Z, Castelain MH, Renaudin MH, Monnot A, Fischler M (1995) Validation of a simple method assessing nitric oxide and nitrogen dioxide concentrations. Intensive Care Med 21: 537–541
Skimming JW, Blanch PB, Banner MJ (1997) Behavior of nitric oxide infused at constant flow rates directly into a breathing circuit during controlled mechanical ventilation [see comments]. Crit Care Med 25: 1410–1416
Skimming JW, Cassin S, Blanch PB (1995) Nitric oxide administration using constant-flow ventilation. Chest 108: 1065–1072
Tibballs J, Hochmann M, Carter B, Osborne A (1993) An appraisal of techniques for administration of gaseous nitric oxide. Anaesth Intensive Care 21: 844–847
Young JD, Dyar OJ (1996) Delivery and monitoring of inhaled nitric oxide. Intensive Care Med 22: 77–86
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Ney, L., Kemming, G., Kuebler, W.M. et al. Dosage of inhaled nitric oxide: a simple method for experimental studies. Res. Exp. Med. 199, 121–128 (1999). https://doi.org/10.1007/s004330050117
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s004330050117