Abstract
Alumina ceramic-on-ceramic total hip arthroplasty (THA) has been widely used due to its advantages such as low wear, scratch resistance, wettable surface and relatively low biological reactivity of the wear particles. Nevertheless, this material in THA still persists to be one of the major concerns about the risk of fracture, due to its brittleness. Many authors have reported a fracture of the ceramic head but few reported a fracture of the ceramic acetabular insert. In order to reduce the rigidity of the ceramic-on-ceramic coupling and prevent an impingement between the rim of the ceramic liner and the metal neck of the femoral stem, a modular acetabular component with a sandwich insertion (alumina/polyethylene/titanium) was proposed. We report the fracture of the ceramic acetabular liner of such a ceramic sandwich cup due to a slightly retroverted position of the cup that causes an impingement between the femoral stem and the rim of the insert. The fracture occurred 3 years after the operation without trauma. At revision the entire cup was replaced using a polyethylene liner without inner ceramic liner.
Similar content being viewed by others
References
Arenas A, Tejero A, Garbayo A, Arias C, Barbadillo J, Lecumberri J (1999) Ceramic femoral head fractures in total hip replacement. Int Orthop 23(6):351–352
Barrack RL, Burak C, Skinner HB (2004) Concerns about ceramics in THA. Clin Orthop Relat Res 429:73–79
Diwanji SR, Seon JK, Song EK, Yoon TR (2007) Fracture of the ABC® ceramic liner: a report of three cases. Clin Orthop Relat Res 453:97–102
Ha YC, Kim SY, Kim HJ, Yoo JJ, Koo KH (2007) Ceramic liner fracture after cementless alumina-on-alumina total hip arthroplasty. Clin Orthop Relat Res 458:106–110
Hasegawa M, Sudo A, Hirata H, Uchida A (2003) Ceramic acetabular liner fracture in total hip arthroplasty with a ceramic sandwich cup. J Arthroplasty 18(5):658–661
Murphy SB, Ecker TM, Tannast M (2006) Two- to 9-year clinical results of alumina ceramic-on-ceramic THA. Clin Orthop 1–6
Pulliam IT, Trousdale RT (1997) Fracture of a ceramic femoral head after a revision operation. A case report. J Bone Joint Surg Am 79A(1):118–121
Toran MM, Cuenca J, Martinez AA, Herrera A, Thomas JV (2006) Fracture of a ceramic femoral head after ceramic-on-ceramic total hip arthroplasty. J Arthroplasty 21(7):1072–1073
Walter A (1992) On the material and the tribology of alumina–alumina couplings for hip joint prostheses. Clin Orthop 282:31–46