In vivo high-resolution MR imaging of the carpal tunnel at 8.0 tesla

  • Article
  • Published:
Skeletal Radiology Aims and scope Submit manuscript

Abstract

Objective. To determine the feasibility of acquiring in vivo images of the human carpal tunnel at 8 tesla (T).

Design. The wrist of an asymptomatic volunteer was imaged with an 8 T /80 cm magnet. The subject was imaged prone with the arm over the head and the wrist placed in neutral position in a custom-built dedicated shielded wrist coil. Axial two-dimensional gradient-echo (GRE) images of the wrist were acquired.

Results. Image contrast and resolution at 8 T are excellent. The infrastructure of the median nerve, particularly the interfascicular epineurium and individual fascicles, is better visualized at 8 T than at 1.5 T. The flexor tendons are well delineated from each other and the surrounding soft tissues, and tertiary tendon fiber bundles are resolved. The boundaries of the carpal tunnel are better defined at 8 T.

Conclusion. We have obtained the first high-quality in vivo images of the human carpal tunnel at 8 T. The 8 T images demonstrated better contrast and resolution than those obtained at 1.5 T.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Additional information

Electronic Publication

Rights and permissions

Reprints and permissions

About this article

Cite this article

Farooki, S., Ashman, C.J., Yu, J.S. et al. In vivo high-resolution MR imaging of the carpal tunnel at 8.0 tesla. Skeletal Radiol 31, 445–450 (2002). https://doi.org/10.1007/s00256-002-0506-z

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00256-002-0506-z

Navigation