Reflection photoplethysmography of arterial-blood-volume pulses

  • Published:
Medical and Biological Engineering and Computing Aims and scope Submit manuscript

Abstract

Reflection photoplethysmography allows noninvasive recording of arterial-blood-volume pulses. The method is based on monitoring backscattered infrared light and has promising diagnostic value: volume pulse shapes depend on cardiac performance and arterial elasticity. The poor repooducibility of photoplethysmograms at repeated trials on the same subject impairs their usefulness, however. The actual formation of photoplethysmograms was invastigated using agar blocks with blood-filled vessels embedded in them. Two optical processes were observed: (a) Attenuation by the vessel of backscattered and back-reflected light, here vessel extension leads to ‘less’ light at the sensor;(b) Light reflection by the vessel wall proximal to the sensor, and vessel extension gives rise to ‘more’ light at the sensor, this counteracts the attenuation. The combined effects of (a) and (b) make the recorded pulse shape dependent on the transducer/artery position relationship. Volume-pulse transducers that monitor the displacement of arterial walls mechanically (e.g. piezoelectric crystals) are liable to produce similar artefacts. The prospects of minimising these artefacts, by using improved transducer designs, are promising.

Sommaire

La photopléthysmographie permet de faire des enregistrements non-envahissants des impulsions volumétriques du sang artériel. La méthode est basée sur la surveillance d'une lumière infra-rouge diffusée en arrière et a une valeur diagnostique prometteuse: les formes des impulsions volumétriques dépendent de la performance cardiaque et de l'élasticité artérielle. Toutefois la faible reproductibilité des photopléthysmogrammes lors des essais répétés au même sujet réduit leur valuer. La formation des photopléthysmogrammes eux-mêmes a été étudiée à l'aide de blocs d'agar incorporant des vaisseaux remplis de sang. On a observé deux procédés optiques: (a) l'atténuation par le vaisseau de la lumière diffusée en arrière et réfléchie vers l'arrière: dans ce cas l'extension du vaisseau donne ‘moins’ de lumière au niveau du capteur; (b) le réfléchissement de la lumière par la paroi du vaisseau à proximité du capteur, et l'extension du vaisseau donne ‘plus’ de lumiére au capteur, ce qui neutralise l'atténuation. Les effets conjugués de (a) et (b) font que la forme d'impulsion enregistrée dépend de la position du capteur relative à l'artère. Les capteurs des impulsions volumétriques qui surveillent mécaniquement (par exemple, par cristaux piézo-électriques) le déplacement des parois artérielles sont susceptibles de produire des artéfacts similaires. Les possibilités de minimiser ces artéfacts, en utilisant des définitions de capteur améliorées, sont plus prometteuses pour la photopléthysmographie.

Zusammenfassung

Durch Reflektionsphotoplethysmographie ist eine nichtagressive Registrierung der arteriellen Blutvolumenimpulse möglich. Dieses Verfahren basiert auf der Überwachung von rückgestreutem infrarotem Licht und hat einen vielversprechenden diagnostischen Wert: Die Volumenimpulsformen hängen von der Herzleistung und Elastizität der Arterien ab. Die schlechtere Reproduzierbarkeit von Photoplethysmogrammen bei wiederholten Versuchen am gleichen Patienten beeinträchtigt jedoch ihre Nützlichkeit. Die eigentliche Bildung von Photoplethysmogrammen wurde unter Verwendung von Agarblocks mit darin eingebetteten blutgefüllten Gefäßen untersucht. Es wurden zwei optische Vorgänge beobachtet: (a) Dämpfung von rückgestreutem und zurückgeworfenem Licht durch das Gefäß. Hier führt die Gefäßerweiterung zu ‘weniger’ Licht am Fühler. (b) Lichtreflektion durch die Gefäßwand, die dem Fühler am nächsten liegt. Durch Gefäßerweiterung ist ‘mehr’ Licht am Fühler verfügbar, und dies wirkt der Dämpfung entgegen. Die kombinierten Wirkungen von (a) und (b) machen die registrierte Pulsform abhängig vom Meßwandler im Vergleich zur Arterienposition. Volumenpuls-Meßwandler, die die Verlagerung von Arterienwänden mechanisch überwachen (d.h. piezoelektrische Kristalle) erzeugen wahrscheinlich ähnliche Artefakten. Die Aussichten, diese Artefakten auf eine Mindestmaß zu beschränken, indem ein verbesserter Meßwandler eingesetzt wird, sind für die Photoplethysmographie vielversprechender.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Similar content being viewed by others

References

  • Anderson, N. M. andSekej, P. (1967) Reflection and transmission of light by thin films of non-haemolysed blood.Phys. Med. Biol. 12, 185–192.

    Article  Google Scholar 

  • Brecht, K. andBoucke, K. (1953) Zur Abnahme des Arterienpulses am Menschen mit dem Infraton-Mikrophon.Pflugers Arch. 257, 490–494.

    Article  Google Scholar 

  • Barbey, K., Brecht, K. andKutscha, W. (1960) Die Infraton Kapsel, ein neuer mechanoelektrischer Wandler zur Registrierung von arteriellen Volumenpulsen.Ztschr. Kreislauff. 49, 303.

    Google Scholar 

  • Cartwright, C. M. (1930) Infrared transmission of the flesh.J. Opt. Soc. Amer. 20, 81–84.

    Article  Google Scholar 

  • Clairex photoconductive cell application handbook (1971) (10CL971) Mount Vernon, NY 10550.

  • Eliakim, M., Sapoznikov, D. andWeinman, J. (1971) Pluse wave velocity in healthy subjects and in patients with various disease states.Amer. Heart J. 82, 445–457.

    Article  Google Scholar 

  • Fine, S. andWeinman, J. The use of photoconductive cells in photoplethysmography.J. Med. Biol. Eng. (to be published).

  • Gadermann, E. andJungmann, H. (1964)Klinische Arterienpulsschreibung. Ambrosius Barth, Munchen, W. G.

    Google Scholar 

  • Hoeven, G. M. A. andBeneken, J. E. W. (1970) A reliable transducer for the recording of the arterial pulse wave. Progress report 2 on medical physics. TNO, Utrecht, Netherlands.

    Google Scholar 

  • Houssay, B. A. (1955)Human physiology. McGraw-Hill,

  • Kramer, K., Graf, K. andOverbeck, W. (1956) Photoelektrische Reflektometrie an tierischen Organen.Pflugers Arch. 262, 285–300.

    Article  Google Scholar 

  • Pauli, N. E. (1957) Untersuchungen uber das Absorptionsvermogen der Haut im langwelligen Gebiet des Spektrums.Strahlentherapie, 532.

  • Petersen, L. H. (1963) Properties and behaviour of living arterial wall.Physiol. Rev. 42, Suppl. 5, 304.

    Google Scholar 

  • Skan-a-Matic (1970) Skaneteles, NY 13152 POB 68.

  • U.S. Naval Air Development Center Summary (1954) Review of the influence of thermal radiation on human skin. Report MA-27, 11735, Johnsville, Penn.

  • Volker, R. (1965)Cardiac and vascular disorders. C. C. Thomas.

  • Volker, R. (1968)Vasogram adapter zur Diagnostik angiologischer and Kardiologischer ErkranKungen. Neuberger, Penzberg, W. Germany.

    Google Scholar 

  • Wehn, P. S. (1957) Pulsatory activity of peripheral arteries.Scand. J. Clin. Lab. Investig. 9, Suppl 30.

    Google Scholar 

  • Weinman, J. (1967) Photoplethysmography. In:A manual of psychophysiological methods. North Holland Publ. Co., 187–217.

  • Weinman, J. andSapoznikov, D. (1971) Equipment for continuous measurements of pulse wave velocities.Med. & Biol. Eng. 9, 125–138.

    Google Scholar 

  • Weinman, J., Sapoznikov, D. andEliakim, M. (1971) Arterial pulse wave velocity and lieft ventricular tension period in cardiac arrhythmias.Cardiov. Res. 5, 513–523.

    Article  Google Scholar 

  • Weinman, J. (1971) An appraisal of some non-invasive cardiovascular techniques. Workshop no. 2, 9th international conference on medical & biological engineering, Melbourne. Canadian Association for BME.

  • Weinman, J. andFine, S. (1972) Detectivities of photoconductive and siliconp-i-n light sensors in photoplethysmography.T.I.T. J. Life Sci. 2.

  • Wetterer, E. andKenner, Th. (1968)Dynamik des Arterienpulses. Springer-Verlag,241, Fig. 34.8.

Download references

Author information

Authors and Affiliations

Authors

Additional information

Partly supported by the Joint Research Fund of the Hebrew University and Hadassah

Rights and permissions

Reprints and permissions

About this article

Cite this article

Weinman, J., Hayat, A. & Raviv, G. Reflection photoplethysmography of arterial-blood-volume pulses. Med. Biol. Eng. Comput. 15, 22–31 (1977). https://doi.org/10.1007/BF02441571

Download citation

  • Received:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02441571

Keywords

Navigation