Background

A wide variety of orthodontic resins are currently being used clinically for the bonding of orthodontic brackets, orthodontic bands and fixation of lingual appliances. Among the options available on the dental market, resin composite adhesives are a good choice for orthodontic bonding because these materials have good mechanical and aesthetic properties and low failure rates [1]. Orthodontic resin-based adhesives with photo-activation properties are a widely accepted group of composites in the profession because of their advantages such as high early bond strength, better aesthetic performance and controlled working time [2, 3].

Recently, flowable or low viscosity composites have been applied for orthodontic use by many clinicians. However, there are few studies available on this type of composite, and its characteristics and interactions with the enamel surfaces and the oral cavity. Flowable composites are produced by retaining the same small particle sizes as for traditional hybrid composites and reducing the filler content, which allows a decrease in the mixture viscosity [4].

Studying the degree of conversion (DC) of orthodontic resins and its relationship with material viscosity is important to determine if different brands of resin composites show different leaching abilities for unpolymerized components, and thus present different potentials for toxicity to cells in the mouth [5]. Studies show that components released by dental resin composites, including diluents such as triethylene glycol dimethacrilate (TEGDMA), some additives, plasticizers (dicyclohexyl phthalate; bis (2-ethylhexyl) phthalate), can make collagen and protein found in the oral cavity and throughout the human body less resistant to enzymatic degradation. In addition, the resin composite co-monomer, TEGDMA, can cause gene mutation in vitro [57]. Thus, materials that have a higher degree of dissolution and/or chemical degradation do not seem to be the best choice for a clinician. Therefore, mechanical and physical properties of new materials need to be studied, to validate their safety when used in the oral environment.

Thus, the purpose of this study is to investigate the DC of five commercially available light-cured orthodontic resins that present different viscosities. The results may determine a relationship between the viscosity and the DC of the materials.

Methods

Five brands of light-cured orthodontic bonding systems with different viscosities were selected for this study. Commercial names, viscosities and composition indicated by the manufacturers of these materials are shown in Table 1. A total of 25 specimens were produced, 5 for each brand of composite used in this experiment (n = 5).

Table 1 Material identification

The specimens, 5 mm in diameter and 2 mm in length, were made at room temperature using a silicon mold. After placing a small amount of material into the mold, a Mylar strip was placed on top of the unpolymerized material and a 4-mm-thick glass slide was pressed on top of the silicon mold. This process was repeated for all brands of resin.

Each specimen was polymerized for 20 s using a LED light-curing unit (VALO, Ultradent), The light intensity was 1000 mW/cm2, the amount of time and energy provide for light curing were above the minimum required by the manufactures. A glass slide (1 mm) was used to standardize the distance between the light source and each specimen, and also to produce specimens with a flat surface. Immediately after polymerization, the samples were subjected to Fourier transform infrared (FTIR) spectroscopy analysis.

Non-cured resin specimens were also subjected to FTIR spectroscopy, to serve as non-cured references. Each sample was placed directly on top of the device’s crystal for prompt analysis. The degree of polymerization was monitored by the FTIR spectroscopic analysis in transmission mode, and then converted to absorbance mode, using the device’s software. The FTIR spectrometer (Nicolet 6700, Thermo Scientific) ran under the following conditions: 4000–400 cm−1 wavenumber range 0.09 cm−1 resolution and 32 scans.

The DC was obtained using the results from the spectroscopic analysis. The DC was calculated using the formula shown in Fig. 1. The equation correlates the absorbance value of the aliphatic C=C peak approximately at 1638 cm−1 with the aromatic C–C peak at approximately 1609 cm−1, for the samples before and after polymerization. Statistical analysis was performed using a one-way analysis of variance (ANOVA) and Tukey’s post hoc test at a significance level of 5 %.

Fig. 1
figure 1

Equation to determine the degree of conversion

Results

There were significant differences in the degree of polymerization for the resins in this study (p < 0.05). The DC was calculated using the FTIR analysis results from each sample, and then determining the percentage for each group of samples. The mean values (MV) and their standard deviations (SD) are presented in Table 2 and illustrated in Fig. 2.

Table 2 DC mean values and standard deviation
Fig. 2
figure 2

Mean and standard deviation for degree of conversion (%) of 5 orthodontic resins

Based on the mean values of conversion recorded for each material, the materials with a lower viscosity showed a higher degree of polymerization than those with higher viscosity.

TR had the lowest mean DC value (24.6 %), which was significantly different than the other materials (p < 0.05). The DC results were 39, 44.3 and 52.5 % for OB, VF and OS, respectively. There were no significant differences between the mean values of OB and VF, or between VF and OS. However, mean values for OB and OS were significantly different. OF showed the highest mean DC value that was significantly different from all the other systems tested, except OS.

Discussion

This study assessed the quantity of residual non-reagent monomers right after the polymerization of orthodontic resins that are available on the dental market. It was found a correlation between the DC and the degree of viscosity for the materials tested.

Among other techniques, FTIR is an efficient and widely used to verify the DC [5, 8, 9]. However, this method has limitations. The FTIR does not allow assessment of the depth of cure; that is, differences in the polymerization level between different parts of the sample are ignored [10]. In this study, the DC % values reported are averages from the top layers of each orthodontic adhesive tested.

The DC or degree of polymerization of the aliphatic C=C in a given orthodontic adhesive is an important factor because it affects the physical and mechanical properties of these materials such as hardness, tensile strength, compressive strength, dissolution (solubility), degradation and biocompatibility [11, 12]. The conversion of monomers is determined by the percentage of the remaining methacrylate groups after curing, when compared to the quantity present in the uncured material [9].

Orthodontic bonding resins, in their polymerized form, are not inert materials [13]. Even when following the manufacturers’ recommendations and instructions regarding polymerization, quantities of uncured components can be found and can be leached into the oral environment with the help of saliva. Thompson et al. [13] used UV spectrophotometry to analyse unpolymerized material extracted from cured orthodontic bonding resins in various aqueous solutions. Their findings show that orthodontic resins release considerable quantities of uncured components and that appropriate precautions should be taken when handling these materials.

The chemical composition of orthodontic bonding adhesives is similar to that of dental composites and sealants [1, 13]. These materials usually contain a mixture of high and low molecular weight methacrylate monomer derivatives such as BIS-GMA and TEGDMA, respectively, and also polymerization inhibitors, catalysts and inorganic particles [14]. These components are exposed to oral fluids and they contact oral tissues throughout orthodontic treatment. Although they are widely accepted in modern dentistry, the composition of these materials presents cytotoxic properties that can be hazardous to patients’ health [13, 1517]. Studies show that monomer leaching from resins can occur at two time points: during the resin setting period and when the resin is degraded [18]. Monomer leaching at the first time point is related to the degree of conversion.

The data obtained in this study is in agreement with published findings, which indicate that different materials have different percentages of remaining methacrylate groups after light curing (i.e. different DC values) [14]. The differences found in conversion values could be attributed to variation in resins composition, in terms of monomeric systems, type and quantity of load particles, concentration of diluents and initiators [19]. The percentage of chemical conversion in the orthodontic resins analyzed decreased within a range of 24.6–53 %, which is in contrast to the range of 55–75 % that was reported in other studies [2022]. Ferracane and Greener [27] studied unfilled Bis-GMA-based dental resins that had a DC ranging from 55 to 72 %. Data related to the TR system, which showed the lowest DC % value, was similar to the findings of other studies that tested the same resin for cytotoxicity and DC [19, 23]. Two resins from Opal Orthodontics, the OF and OS, showed the highest DC % values, but there was no statistical differences among the resins. These two are low viscosity resins, and the OS is a fluoride-releasing orthodontic sealant, which is intended to protect the dental surface that will receive the bracket. To date, there are only a few studies on the Opal Orthodontics’ resins [24]; one of these studies was by Hess et al. [25], who studied the decalcification phenomenon of the dental structure during orthodontic treatment.

Correlating the DC % values and the viscosity of the materials investigated, the highest DC mean values were found in the low viscosity samples, suggesting a relationship between the DC and viscosity of materials. The filler content and the viscosity of resins may interfere in the monomer conversion, because they can limit the mobility of monomers and propagation of the polymerization reaction [26]. The lower viscosity of resins allows better monomeric mobility and distribution of free radicals inside the material, which can enhance the polymerization process leading to a greater monomer conversion [27]. Because of this principle, low viscosity composites may allow enhanced diffusion of reactive groups and promote the curing reaction, resulting in a higher DC [27]. In addition, lower viscosity composites have been shown to adequately bond to enamel without requiring an intermediary resin, and they also have the ability to infiltrate acid-etched enamel to form a strong bond [28, 29].

The findings of this study showed the degree of polymerization of some commercially available orthodontic resins and determined an association with the material viscosity. However, further laboratory and clinical studies are recommended to analyze other properties of orthodontic resins, especially the flowable type.

Conclusions

Among the tested resins, the low viscosity resins had a higher DC, which can indicate a lower toxicity. This may be because of a potentially low residual monomer elution, resulting in a lower health risk and a better choice of material to be used in clinical practice.