Abstract
Purpose
To apply for detecting low-concentration gadolinium contrast agent (Gd), such as those in the perilymph and cerebrospinal fluid (CSF) at 4 h after intravenous administrations (IV), we tried to clarify the degree of increased signal intensity (SI) of low-concentration Gd by longer repetition time (TR) in heavily T2-weighed 3D-FLAIR.
Materials and methods
A phantom was designed using Gd diluted with saline at 0.250–0.008 mM and obtained images by varying the TR from 9000 to 16,000. We measured the SI of five slices for each concentration and compared the mean SI (SImean) values. Normalized SI (SInorm) values, defined as the SImean value of each TR divided by the SImean value of 9000-ms TR for each concentration were compared.
Results
Longer TR increased the SImean values in all solutions. In the 0.250 mM solution, the SImean value of 16,000-ms TR was 1.07-times greater than that of 9000-ms TR. In contrast, the SImean value of 16,000-ms TR was 1.67-times greater than that of 9000-ms TR in 0.008 mM solution.
Conclusion
SI increased with longer TR and lower Gd concentration. Thus, longer TR was useful in detecting low-concentration Gd, such as those in the perilymph and CSF at 4 h after IV.
Similar content being viewed by others
References
Naganawa S, Kawai H, Sone M, Nakashima T. Increased sensitivity to low concentration gadolinium contrast by optimized heavily T2-weighted 3D-FLAIR to visualize endolymphatic space. Magn Reson Med Sci. 2010;9:73–80.
Naganawa S, Suzuki K, Yamazaki M, Sakurai Y. Serial scans in healthy volunteers following intravenous administration of gadoteridol: time course of contrast enhancement in various cranial fluid spaces. Magn Reson Med Sci. 2014;13:7–13.
Naganawa S, Nakane T, Kawai H, Taoka T. Gd-based contrast enhancement of the perivascular spaces in the basal ganglia. Magn Reson Med Sci. 2017;16:61–5.
Deike-Hofmann K, Reuter J, Haase R, Paech D, Gnirs R, Bickelhaupt S, et al. Glymphatic pathway of gadolinium-based contrast agents through the brain: overlooked and misinterpreted. Investig Radiol. 2018. https://doi.org/10.1097/RLI.0000000000000533.
Iliff JJ, Lee H, Yu M, Feng T, Logan J, Nedergaard M, et al. Brain-wide pathway for waste clearance captured by contrast-enhanced MRI. J Clin Investig. 2013;123:1299–309.
Iliff JJ, Nedergaard M. Is there a cerebral lymphatic system? Stroke. 2013;44:S93–S9595.
Yang L, Kress BT, Weber HJ, Thiyagarajan M, Wang B, Deane R, et al. Evaluating glymphatic pathway function utilizing clinically relevant intrathecal infusion of CSF tracer. J Transl Med. 2013;11:107.
Iliff JJ, Chen MJ, Plog BA, Zeppenfeld DM, Soltero M, Yang L, et al. Impairment of glymphatic pathway function promotes tau pathology after traumatic brain injury. J Neurosci. 2014;34:16180–93.
Kress BT, Iliff JJ, Xia M, Wang M, Wei HS, Zeppenfeld D, et al. Impairment of paravascular clearance pathways in the aging brain. Ann Neurol. 2014;76:845–61.
Naganawa S, Kawai H, Taoka T, Sone M. Improved HYDROPS: imaging of endolymphatic hydrops after intravenous administration of gadolinium. Magn Reson Med Sci. 2017;16:357–61.
Uematsu H, Dougherty L, Takahashi M, Butler NS, Song HK, Ohno Y, et al. A direct comparison of signal behavior between 4.0 and 1.5 T: a phantom study. Eur J Radiol. 2003;45:154–9.
Rohrer M, Bauer H, Mintorovitch J, Requardt M, Weinmann HJ. Comparison of magnetic properties of MRI contrast media solutions at different magnetic field strengths. Investig Radiol. 2005;40:715–24.
Rydberg JN, Riederer SJ, Rydberg CH, Jack CR. Contrast optimization of fluid-attenuated inversion recovery (FLAIR) imaging. Magn Reson Med. 1995;34:868–77.
Kita M, Sato M, Kawano K, Kometani K, Tanaka H, Oda H, et al. Online tool for calculating null points in various inversion recovery sequences. Magn Reson Imaging. 2013;31:1631–9.
Wong EC, Liu TT, Luh WM, Frank LR, Buxton RB. T(1) and T(2) selective method for improved SNR in CSF-attenuated imaging: T(2)-FLAIR. Magn Reson Med. 2001;45:529–32.
Funding
None.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Ethical statement
This article does not contain any studies with human participants or animals performed by any of the authors.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
About this article
Cite this article
Kato, Y., Bokura, K., Taoka, T. et al. Increased signal intensity of low-concentration gadolinium contrast agent by longer repetition time in heavily T2-weighted-3D-FLAIR. Jpn J Radiol 37, 431–435 (2019). https://doi.org/10.1007/s11604-019-00828-0
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11604-019-00828-0