Introduction

Complex sleep apnea syndrome (CompSAS) describes the development of central apneas in a patient with obstructive sleep apnea upon treatment restoring the airway patency [15]. Due to its all-inclusive definition (predominance of obstructive apneas during diagnostic testing, appearance of central apneas during PAP therapy) complex sleep apnea likely represents a common phenotypic manifestation of a group of conditions. Mechanisms that are postulated in the development of CompSAS include elevated chemoreceptor sensitivity [6], decreased arousal threshold [7], prolonged circulation time [8], and use of opioid medications [9]. This multitude of pathways leading to CompSAS, and the improvement of some of these mechanisms with time on treatment [10] likely results in the dynamic phenotype, with the improvement or disappearance CompSAS activity in some patients [1115], and its appearance de novo in others [14].

As a consequence of this variable etiologic background, several clinical risk factors for CompSAS among patients with OSA have been identified. CompSAS phenotype has been described in patients with congestive heart failure and Cheyne–Stokes breathing pattern [16], with up to 18 % of CHF patients in a recent study exhibiting CompSAS activity [17]. The presence of arterial hypertension and coronary artery disease has been associated with CompSAS activity in other studies [14, 16].

It is, however, apparent from all the discussed studies and our clinical practice, that many patients with CompSAS do not have any identified “risk factors” for this condition prior to the polysomnography (PSG) evaluation. Delayed recognition of residual central sleep apnea delays institution of the effective treatment [18], and use of ineffective treatment is associated with higher risk of residual symptoms [19] and may worsen cardiovascular prognosis at least in some patient populations [20]. Using our database of CompSAS patients, we therefore aimed to estimate the prevalence of recognized risk factors among patients with CompSAS and to define other clinical and polysomnographic characteristics of these patients.

Material and methods

The study was approved by the Institutional Review Board of the NorthShore University HealthSystem. All patients were referred to a single sleep center, and evaluated between 11/1/2006 and 06/30/2011 by one of the team of ten sleep board-certified or board-eligible physicians. Patients underwent a diagnostic and therapeutic polysomnogram. If severe sleep apnea was observed during the diagnostic study (apnea–hypopnea index, AHI > 30/h), over at least 2 h of recorded sleep, and sufficient time remained (>3 h of expected bed time), the patient had a CPAP titration during the same night (“split-night protocol”). All studies were performed in compliance with American Academy of Sleep Medicine (AASM) standards and were scored according to the AASM scoring criteria. Airflow and respiratory effort were monitored using nasal pressure transducer and respiratory impedance plethysmography during the diagnostic study, and using the flow channel from the CPAP or BPAP plus respiratory impedance plethysmography during the positive pressure titration studies. Hypopneas were defined as an abnormal respiratory event lasting at least 10 s with at least a 30 % reduction in thoracoabdominal movement or airflow as compared to baseline, and with at least a 4 % oxygen desaturation [21].

Following the evaluation, patients were treated with PAP modalities. According to the local policies, upon the appearance of central apnea activity, bilevel PAP in the spontaneous mode (BPAP-S) was tried with expiratory PAP set at the level of CPAP that controlled obstructive events, and inspiratory PAP at the level of EPAP + 4–5 cmH2O. Oxygen was supplemented with PAP therapy in cases of persistent hypoxemia. In cases of ineffectiveness of BPAP-S, adaptive servo ventilation was tried, with expiratory pressure based on CPAP titration, and pressure support range left to the discretion of the physician. Final decision on therapy was made by the managing physician based on the results of all PAP studies. There was no formal policy on the use of hypnotics with PAP.

The objective compliance with the PAP device was assessed at the first visit at 4–6 weeks of its use. Patients who did not come for their follow-up appointment were contacted over the telephone regarding their usage. All patients who could not be contacted or could not supply the compliance card were treated as non-compliers.

Patients with complex sleep apnea were included in the analysis. “Complex sleep apnea” was defined as development of central sleep apnea or with central apnea index of more than 5/hour or prominent and disruptive Cheyne–Stokes breathing pattern in a patient with obstructive sleep apnea [2]. Individual demographic and clinical variables were collected. Patients were recorded to have (1) congestive heart failure, if they had an echocardiographic documentation within 6 months of their polysomnogram of systolic or diastolic heart failure, (2) coronary artery disease, if they had a history of myocardial infarction, or a revascularization procedure, (3) hypertension, if they were on therapy for it, (4) diabetes mellitus, if they were treated for it, (5) hypothyroidism, if they were treated with thyroxin supplementation, (6) asthma or (7) chronic obstructive pulmonary disease, if they carried either diagnosis made by their primary physician and were treated, (8) stroke if they had a history of it, (9) atrial fibrillation if they had this arrhythmia in their medical history (patients did not have to be in atrial fibrillation at the time of their sleep evaluation), (10) neuromuscular weakness, (11) Parkinson’s disease, or (12) cognitive impairment, if they were diagnosed and treated for those by the neurologist, (13) depression, if they were taking antidepressants at the time of sleep consultation, and (14) end-stage renal disease if they were on renal replacement therapy. Patients were noted to take opioids, benzodiazepines, or gabapentin if they were currently taking respective medications at the time of sleep consultation.

Laboratory and auxiliary testing (pulmonary function tests, echocardiography) data were included in the analysis if they were collected within 3 months of the sleep evaluation. Serum bicarbonate level, arterial blood gases, left ventricular ejection fraction (LVEF), estimated right ventricular systolic pressure (RVSP), forced expiratory flow in one second (FEV1) and diffusing capacity of the lung for carbon monoxide (DLCO) were collected wherever available.

Continuous data were presented as medians and interquartile ranges (IQR) and analyzed using ANOVA. Discrete data were analyzed using Fisher’s exact test. Correlations were analyzed with Pearson’s correlation. Comparisons for serial variables were performed using a paired t test.

Results

One hundred and fifty patients (117 (78 %) males, 33 (22 %) females) with CompSAS were identified. Mean (IQR) age of the population was 62.0 (53–72) years. At the time of sleep consultation they were found to be overweight with the body mass index (BMI) of 32.0 (28.4–35.8) kg/m2 and had excessive daytime sleepiness with Epworth Sleepiness Scale score of 11.0 (8.0–14.0). Significant comorbidities are presented in Table 1.

Table 1 Comorbidities of patients with CompSAS included in the analysis. N = 150

Thirty patients (20 %) had been treated for OSA prior to the evaluation. Of those, 16 were on positive airway pressure therapy (10 continuous positive airway pressure (CPAP), 3 bilevel positive airway pressure-spontaneous (BPAP-S), 2 autoadjusting positive airway pressure (APAP), and 1 BPAP-S with oxygen), 1 was treated with uvulopalatopharyngoplasty, and 13 had been on CPAP, but had stopped this therapy prior to current evaluation.

The sleep evaluation consisted of diagnostic polysomnography (61 patients—full night, 89—split night), with mean apnea–hypopnea index (AHI) of 50.8 ± 29.7/h. The results of therapeutic testing are presented in Table 2. Primary positive airway pressure testing consisted of CPAP (148 patients, 98.7 %) or BPAP-S testing (two patients, one patient with prior BPAP-S use and one patient with concurrent hypoventilation). Eventually, 70 patients (46.7 %) tried BPAP-S at some point of PAP titration, typically during a multi-modality titration study [18, 22]. Of note, similar central apnea indices on CPAP were noted in patients studied in the full-night protocol (21.7 ± 21.6) and split-night protocol (26.0 ± 21.0, p = 0.24).

Table 2 Polysomnographic variables during diagnostic testing and therapeutic testing in patients with CompSAS. For CPAP and SV nights, presented data reflect “at best pressure” results. Stage- and position-dependent AHI were calculated over whole night. For adaptive servo ventilation, values represent expiratory/minimal inspiratory/maximal inspiratory pressures

Seven patients refused further evaluations and were lost to follow-up. One hundred thirteen patients underwent a titration with an adaptive servo ventilation device. Ultimately, 107 patients were prescribed an adaptive servo ventilation device, 21 patients were prescribed CPAP, 8 patients were treated with CPAP and a hypnotic, 4 patients were treated with APAP, 2 patients were treated with bilevel PAP in a spontaneous-timed mode (BPAP-ST), and one patient was treated with BPAP-S.

We then calculated the prevalence of three risk factors that have been associated with CompSAS phenotype—hypertension, coronary artery disease and congestive heart failure. Fifty-three patients (35.3 %) did not have any of the three risk factors examined, while 97 patients (64.7 %) had at least one (Fig. 1). Since one of the postulated mechanisms of CompSAS is hypocapnia leading to repeated crossing of the apneic threshold, we examined the data of 91 patients with both serum bicarbonate and LV ejection fraction. Six patients had serum bicarbonate of <23 mM/L and 16 patients had EF of <55 % (Fig. 2).

Fig. 1
figure 1

Prevalence of hypertension (HTN), coronary artery disease (CAD), and congestive heart failure (CHF) among 150 patients with complex sleep apnea

Fig. 2
figure 2

Serum bicarbonate level [in millimoles per liter] and left ventricular ejection fraction [LVEF, %] in 91 patients with CompSAS. Lines represent limits of normal values (23–29 mM/L for serum bicarbonate and 55 % for LVEF)

At follow-up visit, 110 patients were using their therapy, and were significantly less sleepy than on the initial evaluation (ESS score 7.6 ± 5.0, p < 0.001). Download of the compliance device showed an average nightly use of 4.5 ± 2.9 h, on 70 % ± 40 % of the nights.

Discussion

Since the first description of complex sleep apnea by Gilmartin et al [1, 22], the sheer existence, definition, and treatment of this syndrome is debated [23, 24]. In our opinion, clinical utility of “CompSAS” designation lies in differences in acute and subacute response to PAP treatments of patients with “typical” OSA and CompSAS patients: while CPAP is highly effective in OSA, its acute effectiveness in much lower in CompSAS [25]. This low acute effectiveness of CPAP (documented again in our patients) translates into poorer subjective benefit of CompSAS patients than OSA patients when they are treated with CPAP [19]; such a poor early subjective benefit has been associated with poorer long term use among OSA patients [26]. Additionally, in a substantial proportion of patients with CompSAS, long-term effects of therapy with CPAP are also suboptimal [12, 27]. Therefore, accepting the limitations of current definition of CompSAS, which is purely based on polysomnographic variables, we attempted to define the broad clinical characteristics of patients with CompSAS.

Our study represents the largest group of patients with CompSAS described clinically to date. Cassel et al. described 82 patients with CompSAS in their prospective observational study [14], while Westhof et al. described 56 patients with central apnea activity, but included only patients with normal B-type natriuretic peptide levels [15]. Finally, Javaheri et al. examined 84 patients with central apnea activity on CPAP, but included in this group those with central sleep apnea [13]. Hypertension was present in over half of our patients, which confirms the findings of Westhof et al. and Cassel et al., who found hypertension in 72 and 71 % of their patients with CompSAS, respectively [14, 15]. Our dataset contained significantly more patients with CompSAS and heart failure than in populations described by other authors [1316]. Coronary artery disease, diabetes, and depression were all present in more than 20 % patients of our population. Our laboratory data analysis showed changes typical of comorbidities: patients treated with opioids had hypercapnia, while patients with CHF tended to have lower LVEF. Though CompSAS activity has been reported in opioid users, this patient population has not been identified as a risk factor for CompSAS [2830].

In our patient population, over one third of patients did not have any recognized risk factors for CompSAS and this diagnosis was unexpected prior to PSG testing. This suggests that different mechanisms are responsible for appearance of CompSAS phenotype in different patients; polysomnography is likely to be the one of the modalities used to establish the exact mechanism in an individual patient but it is unlikely to be the only one. Measurement of chemoresponsiveness, arousal threshold or controller gain may improve the accuracy of this diagnosis, but are still experimental and not in daily clinical use [31]. Breath-to-breath variability prior to sleep onset was found to be higher in patients with mixed and central apnea tendency (and presumably higher chemoresponsiveness) that in patients with pure obstructive sleep apnea [32].

Treatment of CompSAS is another area of debate, and various PAP modalities have been used [25, 33]. Since central component of CompSAS resolves with time in some patients, CPAP has been proposed as therapy, especially in those patients who do not have cardiac or pulmonary comorbidities [11, 13]. In our sample, adaptive servo ventilation was used most frequently. This reflects both the PAP titration policies and practice style at NorthShore University HealthSystem and is backed up by recent data showing an advantage of ASV over CPAP [27, 34].

The major limitation of the study is its retrospective design. This, specifically, may have introduced a type II error in our analyses pertaining to the physiological variables we gathered in our patients. However, given the low prevalence of CompSAS (1–15 % based on most studies) [2, 13, 14], the setup of a large prospective study systematically examining a similar scope of clinical data would be difficult. The design of our study did not allow for direct comparison of prevalence of comorbidities between CompSAS and OSA patients. In various studies reported in the literature, higher age, male gender, individuals with lower BMI, higher baseline AHI and CAI, coronary artery disease, hypertension and congestive heart failure were more frequently associated with higher prevalence of CompSAS than matched patients with OSA (Table 3), with the significant differences among the studies. We speculate that these differences were due to different ethnic backgrounds of populations studied, which included Americans [2, 12, 13, 19], Australians [16], Japanese [35, 36], and European Caucasians [14] and varying referral patterns in different sleep labs. For the purpose of our analysis, patients diagnosed with CompSAS based on a split-night study were combined with patients in whom the diagnosis was established on the basis of a full-night study, which might be a potential source of bias. Given the descriptive character of this study, we do not think that this bias is significant; also, there were no statistically significant differences in all polysomnographic variables tested between patients diagnosed with CompSAS in a full-night vs split-night protocol.

Table 3 Clinical characteristics of patients with CompSAS (as compared to pts with OSA) in published series

In summary, we have documented the clinical heterogeneity in the largest sample of consecutive patients with complex sleep apnea examined to date. Further research in CompSAS should be focused on better delineation of the interplay of pathophysiological chains involved which will hopefully translate into more accurate prediction of CompSAS activity and better tailoring of therapy in this group of patients.