Introduction

Loss on ignition (LOI)

Heating of samples to specified temperatures has been used widely to measure organic matter and carbonate content in sediment, soil and coal fly ash samples (Bengtsson and Enell 1986; Heiri et al. 2001; Santisteban et al. 2004). The principle that underlies the method is that the sample will lose weight at specific temperatures, reflecting loss of specific components from the material (Heiri et al. 2001). The standard LOI method presented by Konare et al. (2010) used three different temperatures under an air atmosphere: (1) the sample was first heated to 150 °C during which mass loss was attributed to water, (2) the temperature was next raised to 550 °C which combusts organic matter in the sample, yielding ash and carbon dioxide, and (3) the sample was finally heated to 950 °C, the temperature at which carbonates were decomposed to produce carbon dioxide. Weight losses during each heating stage can be used to calculate the amount of moisture, organic matter and carbonate in the sample (Ben-Dor and Banin 1989; Santisteban et al. 2004; Heiri et al. 2001; Virkanen et al. 1997).

LOI is used widely by paleolimnologists, geochemists and sedimentologists and is an inexpensive, “low-tech” method for analysing the composition of sediment and rock samples. There are, however, some drawbacks to the approach, such as the large sample size required, typically 3–5 g. This may consume a substantial proportion of the sample and has the associated disadvantage of requiring more time for mass loss to occur. Other considerations associated with the LOI approach include furnace type, position of samples in the furnace, exposure time at temperature, and the potential for operator error during the weighing process (Heiri et al. 2001; Fan and Brown 2001).

Thermogravimetric analysis (TGA)

TGA is an instrumental method that is used to measure sample mass as a function of time and temperature. One disadvantage of TGA instrumentation is that its relatively high cost, about four times that of LOI. Nevertheless, TGA has a number of advantages, which include: (1) speed of analysis, (2) the fact that a small sample mass can be used, (3) lower incidence of weighing errors because mass determination is automated, and most importantly, (4) the ability to monitor sample mass throughout the analysis, i.e. changes in sample weight can be tracked continuously with temperature increase (Fan and Brown 2001; Charsley et al. 2019).

We compared results from the two thermal analysis techniques (LOI and TGA), applied to a set of sediment samples from cores taken from a now-submerged landscape (Doggerland) (Gaffney et al. 2017). We assessed organic matter and carbonate content in the samples using both analytical approaches, and evaluated the comparability of data from the two techniques using statistical methods. Assessment of moisture composition was not an aim of this study, as the samples had been subjected to different storage, transport and air-drying histories. Complete drying of the samples is, however, an integral part of both methodologies, and is required to provide accurate measures of organic matter and carbonate content.

Materials and methods

Field sampling

Multiple sediment cores were collected on 14 September 2016 under the auspices of the Europe Lost Frontiers Project at the University of Bradford, which is studying prehistoric land now submerged under the North Sea, known as Doggerland (Gaffney et al. 2017, 2020). The cores were kept in a cold room storage environment maintained at 2 °C ± 0.5. Two bulk samples were taken from cores ELF002 and ELF009. The sample from core ELF002 was a reddish-brown, silty fine sand, which was expected to contain low organic matter content and intermediate carbonate content. The sample from core ELF009 was a very dark brown, amorphous dense peat, which was expected to contain a high percentage of organic matter. Ten subsamples were taken from each bulk sample, yielding 20 subsamples to be measured by each technique. Core names, locations, date of collection, and seafloor depth at the collection site are shown in Table 1.

Table 1 Core names, locations and depths

Laboratory analyses

Prior to analysis, the two bulk samples (~ 500 g each) were air dried at room temperature for 48 h to remove the bulk water, making them easier to manipulate, homogenise and subsample. Then they were ground to fine powders in a mortar and pestle.

LOI

The method of Konare et al. (2010) was applied to 20 samples (10 samples each from cores ELF002 and ELF009). Initial sample weights were 3.00 ± 0.01 g. A high-temperature Carbolite furnace, which can reach 1100 °C was utilized for this analysis. Samples were placed in pre-weighed crucibles and put into the furnace at 150 °C for 2 h, after which they were removed from the furnace and allowed to cool to room temperature in a desiccator. Once cool, they were re-weighed, primarily to ensure that the samples were completely dried prior the determination of organic and carbonate content. The process was repeated at 550 °C and 950 °C, and percentages of organic matter and lost CO2 were determined applying Eqs. 1 and 2, respectively. The results of the latter were expressed as total carbonate in the dry sample (by multiplying CO2 by 1.36 to express as CO3).

$$Organic \% = \frac{(weight\,at\,150^\circ{\rm C} - weight\,at\,550^\circ{\rm C} )}{ weight\,at\,150^\circ{\rm C} } \times 100 $$
(1)
$$Lost \, CO_{2} \% =\frac{(weight\,at\,550\mathrm{^\circ{\rm C} } - weight\,at\,950\mathrm{^\circ{\rm C} })}{weight\,at\,150\mathrm{^\circ{\rm C} }} \times 100 $$
(2)

TGA

Each sample (40 ± 10 mg) was individually placed on a high-temperature platinum pan and analysed using a TGA Q5000IR (TA Instruments, USA). The balance housing was kept at 40 °C under a nitrogen atmosphere while the sample chamber was purged with air at a flow rate of 25 mL/min. The heating rate was 20 °C/min from room temperature to 150 °C followed by an isothermal period of 30 min to remove all moisture. The second heating (20 °C/min) was to 550 °C, where the temperature was again held for 30 min to ensure complete combustion of all organic matter and no further weight loss. The final heating, again at 20 °C /min, was to 950 °C with an isothermal period of 30 min to ensure that all carbonates in the sample were thermally degraded (Konare et al. 2010; Heiri et al. 2001; Bengtsson and Enell 1986).

Results

To obtain the LOI results, sample weights were accurately recorded to 0.0001 g prior to and after each heating stage. These weights were used to calculate the results by applying them to Eqs. 1 and 2. The TGA results were processed using Universal Analysis software (TA Instruments, UA, v5.5.24) to calculate the loss in mass between the isothermal stages of the method, as detailed in the methods section. Examples of the TGA curves obtained are shown in Figs. 1 and 2. LOI and TGA results for core ELF002 and ELF009 sediments are shown in Tables 2 and 3, respectively.

Fig. 1
figure 1

Representative TGA curve from sample ELF002, illustrating three transitions

Fig. 2
figure 2

Representative TGA curve from sample ELF009, illustrating four transitions

Table 2 Organic matter and carbonate content in core ELF002 sediment, determined by TGA and LOI methods
Table 3 Organic matter and carbonate content in core ELF009 sediment, determined by TGA and LOI methods

Discussion

Comparison of LOI and TGA

For each sediment component analysed (i.e. organic matter and carbonate content), an unpaired t-test with 18 degrees of freedom (df = 18) was applied with 95% confidence level (p = 0.05), to compare differences between means from application of each method to each core. Our null hypothesis was that “there is no statistically significant difference between mean values produced by the TGA and LOI methods” (Whitley and Bell 2002; Miller and Miller 2010).

With respect to organic matter content measured by the LOI and TGA methods, the calculated t-values for means are less than the critical t-values (Tables 2, 3), hence the null hypothesis could not be rejected. This indicates that the LOI and TGA techniques yielded indistinguishable organic matter results.

Mean carbonate content in the sediment samples from ELF002, measured by the two techniques, returned a t-value lower than the critical t-value (Table 2). As a consequence, the null hypothesis could not be rejected, and indicates that for this sediment type, the LOI and TGA methods yield indistinguishable carbonate results. For sample ELF009, the calculated t-value, 2.97, is higher than the critical t-value of 2.10 (Table 3), and we thus rejected the null hypothesis. This indicated that the two approaches produced results that differ, perhaps related to the lower content of carbonate in the ELF009 samples.

Repeatability (within-run precision)

In both samples ELF002 and ELF009, for percent organic matter and carbonate content, both methods showed consistent precision and repeatability, i.e. very low standard deviations (SD ≤ 0.55), and small differences in means (DM ≤ 0.50) (Tables 2, 3).

Multistage decomposition of samples from cores ELF002 and ELF009

During TGA analysis, all 10 samples from core ELF002 produced a curve with three clear transitions, associated with loss of free moisture, organic matter and carbonate (Fig. 1). In contrast, all 10 samples from core ELF009 produced a multistage decomposition in the second heating stage, from 150 to 550 °C (Fig. 2). This multistage response was divided into parts A & B (Fig. 2), in which there are two weight losses from two different decompositions. The temperature range for part A was 173–367 °C, and the temperature range for part B was 387–560 °C. The temperature range of the organic matter ignition from ELF002 was between 384 and 563 °C. This matches well with the temperatures in part B of core ELF009 and represents the ignition of organic matter. Therefore, part A in ELF009 represents loss of a different sample component, which occurred at temperatures between those responsible for loss of free moisture and organic matter. One possibility is mineral dehydration, which requires temperatures between 100 and 380 °C (Hoogsteen et al. 2015; Lopes et al. 2001). This range includes parts of the first (ambient to 150 °C) and second LOI heating stages (150–550 °C). It is therefore possible that water bound in minerals contributed to the LOI values attributed to moisture and organic matter, and explains why the LOI and TGA methods yielded statistically indistinguishable results for this sample.

Conclusions

Principles that underlie the two thermal methods (LOI and TGA) for analysis of sediment composition are similar. Both involve measurement of changes in sample weight as a function of temperature and duration (Abbot 2005; Çaylı and Küsefoğlu 2008). Two sediment samples from the North Sea were used to compare LOI and TGA results—ELF002, a sample with relatively low organic content, and ELF009, a sample with relatively high organic content. The experiment yielded the following conclusions:

  1. 1.

    TGA uses an instrument that carries out the analysis automatically, with less chance for human measurement error and less time required. The ignition process can be monitored in real time and results can be presented in a TGA curve (Figs. 1, 2).

  2. 2.

    TGA is an appropriate instrument for small samples and provides stable results with samples as small as 40 mg.

  3. 3.

    Although much cheaper, the LOI method provides information only for specific preselected temperatures and takes much more time.

  4. 4.

    TGA and LOI both yield consistent repeatable results for determination of organic matter and carbonate content.

  5. 5.

    TGA provides details about the temperatures at which mass is lost, which are not possible with LOI. This may enable the user to identify other temperature-associated changes in the sediment, such as mineral dehydration.

In summary, we conclude that TGA can be used instead of LOI as a method to study sediment samples, and it is highly recommended for determination of organic matter and carbonate content in sediments. TGA is the preferred method in cases for which sample quantity is limited.