Abstract
Purpose
The subcutaneous implantable defibrillator (S-ICD) was developed to avoid complications related to transvenous leads. A trade-off with the S-ICD is the inability to deliver antitachycardia pacing (ATP). Data is scarce about the recurrence and characteristics of ventricular tachyarrhythmias (VTa) during a follow-up in survivors of out-of-hospital cardiac arrest due to ventricular fibrillation (OHCA-VF). The aim of the study is to determine the characteristics of VTa triggering ICD therapy in order to assess whether survivors of OHCA-VF are eligible candidates for the S-ICD.
Methods
All OHCA-VF patients who received a transvenous ICD were identified, 378 patients, age 57 ± 14 years, predominantly male (76%) with ischemic heart disease (58%). Arrhythmic endpoints were appropriate ICD therapies for any ventricular arrhythmia.
Results
Over a median follow-up of 4.5 years, 690 VTa in 91 patients (24%) were terminated by ICD therapy; 70% of patients had < 5 VTa with ICD therapy. VTa with cycle length ≤ 300 ms were mainly (82%) treated by shock, while 83% of VTa with cycle length > 300 ms were treated by ATP. The presence of a remote myocardial infarction (OR 2.07; 95% CI 1.08–3.97) and LVEF ≤ 0.35 (OR 2.09; 95% CI 1.09–4.00) were significantly associated with the occurrence of VTa with cycle length > 300 ms.
Conclusion
S-ICD implantation may be reasonable in survivors of OHCA-VF who present without a remote myocardial infarction and LVEF > 35%.
Similar content being viewed by others
1 Introduction
Placement of an implantable cardioverter-defibrillator (ICD) is a class 1A indication for survivors of cardiac arrest due to ventricular fibrillation (VF). These recommendations are based on the results of randomized clinical trials [1,2,3,4]. ICD implantation in the setting of secondary prevention has been labeled as “appropriate use” in the recent Appropriate Use Criteria Task Force report [5]. Despite the effectiveness of the ICD in terminating life-threatening ventricular arrhythmias, contemporary transvenous ICD systems have been associated with acute and chronic complications due to the use of transvenous leads [6, 7]. Based on this, an entirely subcutaneous implantable defibrillator (S-ICD) has been developed as an alternative to the transvenous ICD system [8]. The S-ICD-system holds the promise of life-saving defibrillation [9,10,11]. However, a trade-off with the S-ICD is the inability to deliver antitachycardia pacing (ATP). The question arises “who are candidates for the S-ICD system?”; one group is those patients who are young and are, therefore, at greater lifelong risk of complications from transvenous systems. Another opinion is that the S-ICD should be limited to patients for primary prevention of sudden death. Is secondary prevention of sudden death a contra-indication for the S-ICD? Given the inability to deliver ATP, the S-ICD is contra-indicated in secondary prevention patients with a history of monomorphic ventricular tachycardia (VT). But little is known about the recurrence and characteristics of ventricular arrhythmias during follow-up in survivors of cardiac arrest due to VF (OHCA-VF).
The aim of the study is to determine characteristics of ventricular arrhythmias triggering ICD therapy in order to assess whether survivors of OHCA-VF are eligible candidates for the S-ICD.
2 Methods
2.1 Study population
Patients for this retrospective observational cohort study were obtained from the prospectively collected registry of all patients who underwent ICD implantation at the Erasmus Medical Center, Rotterdam, The Netherlands. For the purpose of the study, patients with ischemic or nonischemic heart disease who survived OHCA-VF were included. Patients with age < 18 years or inherited arrhythmia disorders were excluded from analysis. The study period for inclusion was from January 2000 to June 2015. The administrative censoring date for analyses was December 2015 for all patients alive until that date.
Data on baseline clinical characteristics, implantation procedures, and scheduled or unscheduled follow-up visits were prospectively collected in the Erasmus MC ICD registry. Information on clinical variables was acquired at the time of ICD implantation. For all patients, renal function was assessed by estimating the baseline glomerular filtration rate (eGFR) using the abbreviated Modification of Diet in Renal Disease (MDRD) Study equation [12]. Impaired renal function was defined as an eGFR < 60 ml/min/1.73m2 according to practice guidelines [13]. The elapsed time from most recent myocardial infarction to ICD implantation was dichotomized at 18 months based on the results of the MADIT II study [14]. A remote myocardial infarction was defined as the time between most recent myocardial infarction and ICD implantation of ≥ 18 months. The programming of the VT zone of the ICD was set to 330–350 ms and the VF zone was set to 250–300 ms. Over the years, the programming of devices has changed, current programming strategies apply higher rate cutoffs and longer duration for arrhythmia detection. In order to identify possible trends, we defined 3 groups according to the implant year (1, 2000–2004; 2, 2005–2009; 3, 2010–2015).
2.2 The follow-up and ICD therapy event analysis
The follow-up started at the time of ICD implantation. Device interrogation was performed on regular visits scheduled every 3- to 6-month basis, and after symptomatic events. At each visit, arrhythmic events with stored electrograms (EGMs) were retrieved from the device’s memory. Two investigators analyzed the stored EGMs with therapy delivery to classify the arrhythmia and assess the appropriateness of device classification and therapy. Ventricular tachyarrhythmias were defined as events with a sudden increase in rate combined with a change in ventricular near-field and far-field EGM morphology from the baseline rhythm. If an atrial EGM was present, the presence of atrioventricular dissociation (ventricular rate > atrial rate) was used to diagnose ventricular tachyarrhythmia. The differentiation of polymorphic from monomorphic ventricular tachyarrhythmias was based on regularity in the rhythm and in the morphology of the near-field and far-field EGMs. Polymorphic ventricular tachyarrhythmias had irregularity in the rhythm and changing morphology in near-field and far-field EGMs. Appropriate ICD therapy was defined as ATP or shock delivered for a ventricular tachyarrhythmia.
Survival status of patients was retrieved from the civil registry. As required by Dutch law, all deceased inhabitants must be registered at the civil registry within 3 days. For analysis of the endpoints, the maximum follow-up was set at 10 years. For patients whose implant date was more than 10 years before December 2015, events occurring after 10 years were not used in the analysis.
2.3 Statistical analysis
Normality of distribution as assessed using the Shapiro-Wilk test. Continuous variables are presented as mean ± SD or as median with 25th and 75th percentiles, where appropriate. Data were compared by the ANOVA or Kruskal-Wallis test, as appropriate. Categorical data are expressed as percentages and compared with Fisher’s exact test. Cumulative mortality and event rates of appropriate ICD therapy were calculated according to the Kaplan-Meier method and were compared with the log-rank test. Univariate predictors associated with appropriate ICD therapy were included in a multivariate logistic regression model. Odds ratios (OR’s) are presented with corresponding 95% confidence intervals (CI’s). Statistical analysis was performed using STATA version 12 SE for Windows (StataCorp, College Station, TX) and SPSS version 24 (IBM Corp., Somers, NY). Statistical significance was defined as P < 0.05 (two-tailed).
3 Results
3.1 Study population
The study cohort consisted of 378 patients. The population was predominantly male (76%) with a mean age of 57 ± 14 years. The majority of patients had ischemic heart disease (58%) and the mean LVEF was 40 ± 16%, with 154 patients (41%) having a LVEF ≤ 35%. Further baseline characteristics of the study population are presented in Table 1. The median follow-up was 4.5 years (2.0 to 7.6 years), during which 91 patients (24%) received appropriate device therapy (ATP or shock) and 66 patients (18%) died.
3.2 The first appropriate ICD intervention
The cumulative incidences of appropriate device therapy were 18%, 25%, and 34%, at 2, 4, and 10 years, respectively. As shown in Fig. 1, the risk for appropriate ICD therapy was highest during the first 2 years after implantation and persisted during the long-term follow-up. The median interval to the first any appropriate ICD intervention was 1.1 years (1.6–6.0 years). The first appropriate ICD intervention was shock delivery as initial therapy in 48 patients and ATP in 43 patients. The mean cycle length of shocked ventricular tachyarrhythmias was 231 ± 41 ms, and the mean cycle length of ventricular tachyarrhythmias treated by ATP was 307 ± 34 ms (P < 0.001).
In Fig. 2, the temporal trend of appropriate ICD therapy by the implant period is presented. At the 3-year follow-up, cumulative appropriate ICD therapy is higher in patients implanted before 2005 than those implanted later (39.7% versus 24.1% and 14.7%; both P’s < 0.05). The mean cycle length of ventricular tachyarrhythmias triggering the first appropriate ICD therapy was not different between the different implant periods (P = 0.70).
3.3 Ventricular tachyarrhythmias triggering ICD intervention
In total, 690 episodes with appropriate ICD intervention (91 patients) were adjudicated. The number of episodes ranged from 1 to 95 per patient. The majority of patients (70%) experienced less than 5 episodes with appropriate ICD interventions. Table 2 shows characteristics of the episodes with appropriate ICD therapy. Four hundred ninety-five episodes for which ATP was delivered as the first therapy were recorded in 59 patients (range 1 to 72 episodes per patient). Subsequent shock delivery was necessary in 132 (27%) of these episodes in 26 patients (range 1 to 70 episodes per patient). Shock as initial therapy was delivered in 195 episodes (62 patients; range 1 to 19 per patient).
Figure 3 presents the distribution of cycle lengths of ventricular tachyarrhythmias for which ICD therapy was delivered. The mean cycle length of shocked arrhythmias was shorter compared to those treated with ATP (229 ± 39 ms versus 326 ± 52 ms; P < 0.001). When considering a cutoff value of 300 ms for cycle length, the majority of arrhythmias with cycle length ≤ 300 ms (82%) were treated by shock, either as initial therapy or after unsuccessful ATP. In contrast, arrhythmias with cycle length > 300 ms were mainly (83%) treated by ATP only. The association between clinical covariates and ventricular arrhythmias with cycle length > 300 ms in univariate and multivariate logistic regression analyses is shown in Table 3. The presence of a remote myocardial infarction (OR 2.08; 95% CI 1.08–3.97) and LVEF ≤ 0.35 (OR 2.09; 95% CI 1.09–4.00) were significantly associated with the occurrence of ventricular tachyarrhythmias with cycle length > 300 ms which are mainly terminated by ATP. When considering the implant period, ventricular arrhythmias with cycle length > 300 ms were less likely to occur in patients implanted after 2010 (OR 0.29; 95% CI 0.12–0.69).
3.4 Mortality
During follow-up, 66 patients (18%) died. At 10-year follow-up, the observed cumulative mortality was 35%. The median interval from device implantation to death was 2.7 years (1.6–6.0 years). In univariate analysis, no association between any appropriate device therapy (ATP or shock) and mortality was observed (P = 0.45). When considering appropriate ICD shocks, a similar result was found (P = 0.41).
4 Discussion
In the current study, we show the rate distribution of ventricular tachyarrhythmias and type of delivered ICD therapy, in patients who survived OHCA-VF. The main findings are (1) the cumulative incidence of appropriate ICD therapy at 10 years was 34%, (2) the majority of patients (70%) experienced less than 5 ventricular tachyarrhythmias treated with ICD therapy, (3) ventricular tachyarrhythmias with cycle length ≤ 300 ms are mainly terminated by shock, and (4) the presence of LVEF ≤ 0.35 and a remote myocardial infarction are associated with slower ventricular tachyarrhythmias (cycle length > 300 ms).
Although prophylaxis of ICD therapy has been widely studied, there is relatively little data on the outcome of patients with a secondary prevention indication. The incidence of appropriate ICD therapy in a secondary prevention population has been studied in a few studies. Freedberg et al. [15] observed appropriate ICD therapy in 62% of patients after the 2-year follow-up. In the Leiden registry, cumulative incidences of appropriate ICD therapy were 52% and 61%, at 5 and 10 years, respectively [16]. More recently, Schaer et al. [17] reported a cumulative incidence of appropriate ICD therapy of 65% at 10 years. The incidence of appropriate ICD therapy in our study (33.5% at 10 years) is lower compared to the previous studies. The higher incidences in previous studies can be explained by the heterogeneous group of patients with both VF and VT as index arrhythmia. A post hoc analysis of the Antiarrhythmics Versus Implantable Defibrillators (AVID) trial compared the incidence of ICD therapy according to the index arrhythmia [18]. After the 3-year follow-up, patients with VT as index arrhythmia were more likely to experience appropriate ICD therapy than those with VF as index arrhythmia (75.5% versus 47.4%). In a recent cohort analysis of 239 patients with a median follow-up of 7.8 years, Boule et al. [19], patients presenting with VF were less likely to require appropriate ICD therapy compared to those presenting with VT (sub-hazard ratio 0.62). Patients who were implanted after an aborted cardiac arrest due to VF have a lower risk of appropriate ICD therapy during the follow-up compared to those with VT as index arrhythmia.
Studies reporting on the type of ventricular arrhythmias during follow-up in secondary prevention patients are scarce. Post hoc analysis of the Pacing Fast VT Reduces Shock ThErapies (PainFREE Rx II) trial and the Inhibition of Unnecessary RV Pacing with AV Search Hysteresis in ICDs (INTRINSIC RV) trial demonstrated that the rate distribution of ventricular arrhythmias was similar between the primary and secondary prevention patients [20, 21]. However, the majority of secondary prevention patients had VT as index arrhythmia in both trials. In our study, only patients who survived OHCA-VF were included to assess the eligibility for the S-ICD. Patients with known recurrent monomorphic VT thought to be amenable for ATP termination are considered not eligible for the S-ICD. The presence of monomorphic VTs is more common late after myocardial infarction due to reentry around scar [22]. In the present study, slower ventricular arrhythmias (rate < 200 bpm) were associated with the presence of LVEF ≤ 35% and a remote myocardial infarction.
The armamentarium of devices that defibrillate in order to prevent SCD has expanded in recent years. As a consequence, ICDs have been used in patients with a variety of clinical needs, including those with documented monomorphic VTs, survivors of OHCA-VF, patients with requirements for bradycardia pacing or cardiac resynchronization with a concomitant indication for an ICD, and patients at increased risk for SCD. Given the short- and long-term complications of ICDs, it is important to select the right device carefully for each individual patient. In clinical practice, many physicians think that a device without ATP to be inferior. However, the perceived requirement of ATP must be balanced with the benefit of S-ICD therapy whose risk of any VT is low for the first years after implantation. Considering recurrent ventricular arrhythmias in S-ICD patients, a recent analysis of the EFFORTLESS registry demonstrated that only 2.2% of patients experienced ICD therapy for more than 1 episode of monomorphic VT [23]. In addition, only 0.5% of patients had the S-ICD removed for conceived need for ATP. Taken all together, the implantation of a S-ICD might be reasonable in survivors of OHCA-VF with LVEF > 35% without the presence of a remote myocardial infarction.
5 Limitations
Our study has several limitations. First, the study is a retrospective, single-center study. However, data in this ICD registry is prospectively collected. Another possible limitation is that the risk of appropriate ICD therapy depends on ICD programming. Over the years, programming of higher rate cutoffs for detection and longer detection duration to reduce unnecessary ICD interventions has been recognized both in the primary and secondary preventions [24, 25]. We accounted for this by defining 3 groups according to date of implant. According to this, the reported overall incidence of appropriate ICD therapy in the present study is probably overestimated rather than underestimated. Last, the negative impact of shocks on mortality has been consistently seen in trials with the primary prevention patients. We found no association between shocks and mortality. However, our study was not designed to evaluate the association between shocks and mortality. The impact of shocks on mortality in patients with a secondary prevention indication merits further investigation.
6 Conclusion
The findings of this study suggest that the implantation of the S-ICD may be reasonable in survivors of OHCA-VF with LVEF > 35% without a remote myocardial infarction.
References
The Antiarrhythmics versus Implantable Defibrillators (AVID) Investigators. A comparison of antiarrhythmic-drug therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmias. N Engl J Med. 1997;337:1576–83.
Connolly SJ, Gent M, Roberts RS, Dorian P, Roy D, Sheldon RS, et al. Canadian implantable defibrillator study (CIDS): a randomized trial of the implantable cardioverter defibrillator against amiodarone. Circulation. 2000;101:1297–302.
Connolly SJ, Hallstrom AP, Cappato R, Schron EB, Kuck KH, Zipes DP, et al. Meta-analysis of the implantable cardioverter defibrillator secondary prevention trials. AVID, CASH and CIDS studies. Antiarrhythmics vs Implantable Defibrillator study. Cardiac Arrest Study Hamburg. Canadian Implantable Defibrillator Study. Eur Heart J. 2000;21:2071–8.
Kuck KH, Cappato R, Siebels J, Ruppel R. Randomized comparison of antiarrhythmic drug therapy with implantable defibrillators in patients resuscitated from cardiac arrest: the Cardiac Arrest Study Hamburg (CASH). Circulation. 2000;102:748–54.
Russo AM, Stainback RF, Bailey SR, Epstein AE, Heidenreich PA, Jessup M, et al. ACCF/HRS/AHA/ASE/HFSA/SCAI/SCCT/SCMR 2013 appropriate use criteria for implantable cardioverter-defibrillators and cardiac resynchronization therapy: a report of the American College of Cardiology Foundation appropriate use criteria task force, Heart Rhythm Society, American Heart Association, American Society of Echocardiography, Heart Failure Society of America, Society for Cardiovascular Angiography and Interventions, Society of Cardiovascular Computed Tomography, and Society for Cardiovascular Magnetic Resonance. J Am Coll Cardiol. 2013;61:1318–68.
Kleemann T, Becker T, Doenges K, Vater M, Senges J, Schneider S, et al. Annual rate of transvenous defibrillation lead defects in implantable cardioverter-defibrillators over a period of >10 years. Circulation. 2007;115:2474–80.
Borleffs CJ, van Erven L, van Bommel RJ, van der Velde ET, van der Wall EE, Bax JJ, et al. Risk of failure of transvenous implantable cardioverter-defibrillator leads. Circ Arrhythm Electrophysiol. 2009;2:411–6.
Bardy GH, Smith WM, Hood MA, Crozier IG, Melton IC, Jordaens L, et al. An entirely subcutaneous implantable cardioverter-defibrillator. N Engl J Med. 2010;363:36–44.
Weiss R, Knight BP, Gold MR, Leon AR, Herre JM, Hood M, et al. Safety and efficacy of a totally subcutaneous implantable-cardioverter defibrillator. Circulation. 2013;128:944–53.
Lambiase PD, Barr C, Theuns DA, Knops R, Neuzil P, Johansen JB, et al. Worldwide experience with a totally subcutaneous implantable defibrillator: early results from the EFFORTLESS S-ICD Registry. Eur Heart J. 2014;35:1657–65.
Burke MC, Gold MR, Knight BP, Barr CS, Theuns DA, Boersma LV, et al. Safety and efficacy of the totally subcutaneous implantable defibrillator: 2-year results from a pooled analysis of the IDE study and EFFORTLESS registry. J Am Coll Cardiol. 2015;65:1605–15.
Levey AS, Bosch JP, Lewis JB, Greene T, Rogers N, Roth D. A more accurate method to estimate glomerular filtration rate from serum creatinine: a new prediction equation. Modification of Diet in Renal Disease Study Group. Ann Intern Med. 1999;130:461–70.
Foundation NK. K/DOQI clinical practice guidelines for chronic kidney disease: evaluation, classification, and stratification. Am J Kidney Dis. 2002;39:S1–266.
Wilber DJ, Zareba W, Hall WJ, Brown MW, Lin AC, Andrews ML, et al. Time dependence of mortality risk and defibrillator benefit after myocardial infarction. Circulation. 2004;109:1082–4.
Freedberg NA, Hill JN, Fogel RI, Prystowsky EN. Recurrence of symptomatic ventricular arrhythmias in patients with implantable cardioverter defibrillator after the first device therapy: implications for antiarrhythmic therapy and driving restrictions. CARE Group. J Am Coll Cardiol. 2001;37:1910–5.
Borleffs CJ, van Erven L, Schotman M, Boersma E, Kies P, van der Burg AE, et al. Recurrence of ventricular arrhythmias in ischaemic secondary prevention implantable cardioverter defibrillator recipients: long-term follow-up of the Leiden out-of-hospital cardiac arrest study (LOHCAT). Eur Heart J. 2009;30:1621–6.
Schaer B, Kuhne M, Reichlin T, Osswald S, Sticherling C. Incidence of and predictors for appropriate implantable cardioverter-defibrillator therapy in patients with a secondary preventive implantable cardioverter-defibrillator indication. Europace. 2016;18:227–31.
Raitt MH, Klein RC, Wyse DG, Wilkoff BL, Beckman K, Epstein AE, et al. Comparison of arrhythmia recurrence in patients presenting with ventricular fibrillation versus ventricular tachycardia in the Antiarrhythmics Versus Implantable Defibrillators (AVID) trial. Am J Cardiol. 2003;91:812–6.
Boule S, Semichon M, Guedon-Moreau L, Drumez E, Kouakam C, Marquie C, et al. Long-term outcome of implantable cardioverter-defibrillator implantation in secondary prevention of sudden cardiac death. Arch Cardiovasc Dis. 2016;109:517–26.
Sweeney MO, Wathen MS, Volosin K, Abdalla I, DeGroot PJ, Otterness MF, et al. Appropriate and inappropriate ventricular therapies, quality of life, and mortality among primary and secondary prevention implantable cardioverter defibrillator patients: results from the Pacing Fast VT REduces Shock ThErapies (PainFREE Rx II) trial. Circulation. 2005;111:2898–905.
Sullivan RM, Russo AM, Berg KC, Stolen KQ, Seth M, Perschbacher D, et al. Arrhythmia rate distribution and tachyarrhythmia therapy in an ICD population: results from the INTRINSIC RV trial. Heart Rhythm. 2012;9:351–8.
Crawford T, Cowger J, Desjardins B, Kim HM, Good E, Jongnarangsin K, et al. Determinants of postinfarction ventricular tachycardia. Circ Arrhythm Electrophysiol. 2010;3:624–31.
Boersma L, Barr C, Knops R, Theuns D, Eckardt L, Neuzil P, et al. Implant and midterm outcomes of the subcutaneous implantable cardioverter-defibrillator registry: the EFFORTLESS Study. J Am Coll Cardiol. 2017;70:830–41.
Kloppe A, Proclemer A, Arenal A, Lunati M, Martinez Ferrer JB, Hersi A, et al. Efficacy of long detection interval implantable cardioverter-defibrillator settings in secondary prevention population: data from the Avoid Delivering Therapies for Nonsustained Arrhythmias in ICD Patients III (ADVANCE III) trial. Circulation. 2014;130:308–14.
Sterns LD, Meine M, Kurita T, Meijer A, Auricchio A, Ando K, et al. Extended detection time to reduce shocks is safe in secondary prevention patients: the secondary prevention substudy of PainFree SST. Heart Rhythm. 2016;13:1489–96.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
The study protocol was approved by the IRB of the Erasmus MC (MEC-2015-619). The ethics committee waived the need for written informed consent, since this retrospective study was not subjected to the Dutch Medical Research Involving Human Subjects Act. The study was carried out according to the ethical principles for medical research involving human subjects established by the Declaration of Helsinki. The privacy of all patients and the confidentiality of their personal information were protected.
Conflict of interest
Dr. Theuns has received research grants from Biotronik and Boston Scientific and consulting fees from Boston Scientific. Dr. Yap has been an ad hoc consultant for Medtronic and Boston Scientific and has been a speaker for Medtronic. Drs. Bhagwandien, Szili-Torok, and Zijlstra have nothing to disclose.
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
About this article
Cite this article
Theuns, D.A.M.J., Bhagwandien, R.E., Szili-Torok, T. et al. Evaluation of recurrent ventricular tachyarrhythmias in patients who survived out-of-hospital cardiac arrest due to ventricular fibrillation: eligibility for subcutaneous implantable defibrillator therapy. J Interv Card Electrophysiol 55, 317–323 (2019). https://doi.org/10.1007/s10840-018-0490-4
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10840-018-0490-4