Abstract
Purpose
A fluorescence-based technique for the detection of parathyroid glands (PGs) intraoperatively was previously reported. The technique was based on the phenomenon in which PGs emit autofluorescence when exposed to near-infrared light and we undertook an evaluation to consider the pathological accuracy of the method.
Methods
The study comprised 17 patients (18 specimens) who underwent thyroid surgery at Kushiro City General Hospital between November 2018 and June 2019. We searched for PGs intraoperatively using a fluorescence spectroscopy system and evaluated the pathological accuracy of the system. We statistically evaluated the clinical factors associated with the accuracy of the system, including age, gender, body mass index, laterality, disease state, renal function, and comorbidity.
Results
Eighteen specimens were evaluated pathologically, with 13 specimens confirmed as PGs. These were evaluated as “true positive,” giving a positive predictive value of 72.2% (13/18). Among the false-negative cases, one specimen was a metastatic lymph node in a patient with papillary thyroid carcinoma. There was a significant difference in the true-positive rates between malignant (25%) and benign (85.7%) disease (P = 0.044).
Conclusion
We consider that this technique is useful, however, we have to exercise care in malignant cases as the true-positive rate may be low.
Similar content being viewed by others
References
Paras C, Keller M, White L, Phay J, Mahadevan-Jansen A (2011) Near-infrared autofluorescence for the detection of parathyroid glands. J Biomed Opt 16(6):067012
McWade MA, Sanders ME, Broome JT, Solórzano CC, Mahadevan-Jansen A (2016) Establishing the clinical utility of autofluorescence spectroscopy for parathyroid detection. Surgery 159(1):193–202
Kahramangil B, Dip F, Benmiloud F et al (2018) Detection of parathyroid autofluorescence using near-infrared imaging: a multicenter analysis of concordance between different surgeons. Ann Surg Oncol 25(4):957–962
McWade MA, Thomas G, Nguyen JQ et al (2019) Enhancing parathyroid gland visualization using a near infrared fluorescence-based overlay imaging system. J Am Coll Surg 225(5):730–743
DiMarco A, Chotalia R, Bloxham R et al (2019) Autofluorescence in parathyroidectomy: signal intensity correlates with serum calcium and parathyroid hormone but routine clinical use is not justified. World J Surg 43(6):1532–1537
Idogawa H, Sakashita T, Homma A (2020) A novel study for fluorescence patterns of the parathyroid glands during surgery using a fluorescence spectroscopy system. Eur Arch Otorhinolaryngol 277(5):1525–1529
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Ethical approval
All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee (include name of committee + reference number) and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standards. Approval for this study was obtained from the Institutional Review Board at Kushiro City General Hospital (H30-24).
Informed consent
Informed consent was obtained from all individual participants included in the study.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Idogawa, H., Sakashita, T., Yagi, T. et al. Pathological evaluation of the accuracy of a fluorescence spectroscopy system for detecting parathyroid glands. Eur Arch Otorhinolaryngol 277, 3145–3147 (2020). https://doi.org/10.1007/s00405-020-06011-w
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00405-020-06011-w