Abstract
Purpose
In retrospective clinical studies digital variance angiography (DVA) provided higher contrast-to-noise ratio and better image quality than digital subtraction angiography (DSA). Our aim was to verify the clinical usefulness and benefits of DVA in carbon dioxide (CO2)-assisted lower limb interventions.
Materials and Methods
A workstation running the DVA software was integrated into a Siemens Artis Zee with Pure angiography system, and this new image processing technology was used in four patients (3 male, 1 female, age: 76.2 ± 4.2 years) with peripheral artery disease (PAD, Rutherford 2–3) and impaired renal function (average eGFR 25.5 ± 11.2 ml/min/1.73 m2). The DSA and DVA images of 46 CO2-assisted runs were visually evaluated by five experts in single-image evaluation using a 5-grade Likert scale and in paired comparisons.
Results
DVA images received significantly higher score (3.84 ± 0.10) than DSA images (3.31 ± 0.10, p < 0.001). Raters preferred DVA images in terms of diagnostic value and usefulness for therapeutic decisions in 85.2% and 83.9% of all comparisons, respectively. These benefits were achieved at lower frame rates (1–3 FPS) than usually recommended for CO2 angiography (4–6 FPS). No adverse events were recorded during or after the procedures.
Conclusions
Our initial experience shows that DVA might facilitate the correct diagnostic and therapeutic decisions, and potentially help to reduce radiation exposure in lower limb CO2 angiography. Although the dose management capabilities of DVA have to be validated in further clinical studies, this technology might be a useful new tool in the operating room and contributes to the safety and efficacy of CO2-enhanced endovascular interventions.
Level of Evidence
Level IV.
Introduction
Carbon dioxide (CO2) gas was introduced in medical imaging as a negative contrast agent decades ago [1]. Unlike iodinated contrast media (ICM), CO2 is a non-nephrotoxic and non-allergic contrast material, therefore the most important indications for its use are impaired renal function and ICM allergy [2]. According to a recent large-scale clinical study (AMACING), the occurrence of contrast-induced nephropathy (CIN) increases to fivefold in patients with eGFR < 30 ml/min/1.73 m2 [3], indicating a very high ICM-related risk for this subset of patients. The use of CO2 significantly decreased (− 65%) the incidence of CIN [4], and proved to be safe and effective in a 5-year follow-up study [5], thus CO2 might be an effective tool to prevent ICM-related side effects.
However, as CO2 is a negative contrast agent, obtaining good quality images with CO2 is a more challenging task, which requires a high-resolution DSA system with image stacking and CO2 specific acquisition modes to increase photon flux and enhance negative contrast visualization [2]. The stacking process requires higher frame rates (4–7.5 FPS), but even under these conditions, the additional use of ICM might be necessary sometimes, resulting in increased radiation exposure and risk for ICM-related adverse events. In special cases, the modified delivery of CO2 can improve the image quality [6], but a more general solution might be an improved image processing method.
The newly developed digital variance angiography (DVA) is based on the principles of kinetic imaging [7] and provides better image quality than the traditionally used digital subtraction angiography (DSA) [8]. In a recent multicentre study assessing the performance of DVA in lower limb CO2 angiography, the contrast-to-noise ratio (CNR) of DVA was 3.5–4.5 times higher than that of DSA, and the visual evaluations clearly showed the superiority of DVA [9]. This retrospective image processing study provided strong proof-of-concept validation but was not able to verify the usefulness and benefits of DVA in clinical practice. For this latter purpose, we have installed the image processing software in our operating room to investigate how the use of DVA can help CO2-enhanced lower limb interventions.
Materials and Methods
Patients
We present a case series of 4 patients who underwent elective CO2-enhanced lower limb angiography between May and July 2019 at the Heart and Vascular Center of Semmelweis University, Budapest, Hungary. The inclusion criteria were symptomatic peripheral artery disease (PAD, Rutherford 2–3) with severe renal impairment (eGFR < 30 ml/min/1.73 m2) or with moderate renal impairment (eGFR < 60 ml/min/1.73 m2) and earlier CIN. Every patient signed a written informed consent before the diagnostic or interventional procedure.
Angiography Procedures
Transradial or transbrachial access and a 5F Pigtail-catheter at the level of aortic bifurcation was used for conventional angiography, whereas an antegrade femoral access was applied for femoro-popliteal interventions. An Angiodroid automated injector (Angiodroid SRL, Bologna, Italy) was used for CO2 delivery. In some cases, lower volumes (3–30 ml) of ICM (Ultravist 370, Bayer Schering Pharma, Berlin, Germany) were also used for diagnostic purposes. Images were recorded with a modified CO2 protocol (Siemens Evenflow, 1–3 FPS instead of the official preset 7.5). The use of the DVA method did not require additional radiation or contrast media load.
Imaging System
DVA technology is available in a CE marked, platform-independent stand-alone software, the Kinepict Medical Imaging Tool v3.1 (KMIT, Kinepict Health Ltd, Budapest, Hungary), which was installed on a desktop computer. This workstation was integrated into a Siemens Artis Zee with Pure angiography suite containing a 30 × 40 cm detector and a Syngo workstation (XWP VD11B Service Pack 2, Siemens Healthcare, Munich, Germany). The KMIT workstation received the unsubtracted image series from the Syngo workstation via a DICOM port and the local area network, and sent back the DVA images automatically to the operating room monitor within 1–2 s, where the interventional radiologist could observe the DSA and DVA images simultaneously (Fig. 1), and used them for guiding the intervention.
Visual Evaluation
The DSA and DVA images were saved for a retrospective visual evaluation. DSA images were carefully postprocessed (stacking, brightness/contrast optimization) to obtain the best available DSA quality and were compared with the DVA images generated automatically by the KMIT software. Five evaluators with 5+ years interventional radiology or endovascular surgery experience rated the image quality of all individual images in a blinded and randomized manner using a 5-grade Likert scale:
1:poor-5:outstanding (see Supplementary material for details). The mean ± SEM and, because of the non-Gaussian distribution of data, the median with interquartile ranges were calculated. Data were analyzed by the Wilcoxon signed rank test. The evaluators also compared the DSA-DVA image pairs in another blinded and randomized survey, where they had to select which image is more suitable for A: the identification of vascular pathology (diagnostic value); and B: achieving correct therapeutic decisions. To describe agreement between raters, percent agreement, Fleiss’ kappa and p values were calculated.
Results
Case Presentations
All presented PAD patients (3 males, 1 female, age: 76.2 ± 4.2, creatinine (Cr) 260 ± 93 μmol/l, eGFR 25.5 ± 11.2 ml/min/1.73 m2) were referred by a vascular surgeon. None of the patients received hemodialysis care, and no adverse events were recorded during or after the procedures. The cases include one diagnostic lower limb angiography and three interventions (for details and images see Figs. 2, 3, 4, 5 and the Supplementary material). There was a general agreement among the interventional radiologists performing the procedures, that DVA images showed more details and better helped the interventional decisions. To objectivate this opinion, the image set of the 4 patients (altogether 46 DSA-DVA image pairs) were retrospectively evaluated by 5 experienced experts.
Visual Evaluation
In single-image evaluation, DVA images received significantly higher score (mean ± SEM: 3.84 ± 0.10) than DSA images (3.31 ± 0.10, Wilkinson signed rank p < 0.001). The data did not follow a normal distribution, therefore the median and interquartile range were also determined for DVA (4.00, 3.45-4.40) and DSA (3.6, 3.00-3.80) (Fig. 6A). To compare the DSA and DVA scores of individual images, a DVA-DSA plot was prepared (Fig. 6B). In 84.8% of images DVA received higher score, in 6.5% the score was equal, and DSA received higher score only in 8.7%.
In paired comparisons, the evaluators preferred DVA images in terms of diagnostic value and usefulness for therapeutic decisions in 85.2% and 83.9% of all comparisons, respectively. The percent agreement, Fleiss K and p were 79.6%, 0.21, p < 0.001 and 77.8%, 0,18, p < 0.001, respectively.
Discussion
DSA records a contrast-enhanced image series and subtracts one of these images (the mask) from the other frames. These subtracted images compose the DSA video and their appropriate integration yields the DSA image [10]. In contrast, DVA uses all frames of a series and calculates standard deviation for each pixel [7]. This algorithm enhances the functional motion-related information (i.e., the flow of contrast agents) but suppresses the noise, therefore the CNR and consequently the image quality is greatly improved.
In line with this, the image comparison showed that the identification of vascular pathology and the therapeutic decisions were much easier on the basis of DVA images. The real-time DVA images received significantly higher Likert score than the postprocessed DSA images, and the available postprocessing tools (brightness/contrast adjustment, pixel shift) make the technology even more versatile. It should be emphasized that the image quality was improved at a lower frame rate (1–3 FPS) than the factory preset 7.5 FPS or the generally recommended 4–6 FPS [2], which provides opportunity for substantial radiation exposure reduction in CO2 angiography.
Conclusions
Our initial experience shows that DVA might facilitate the correct diagnostic and therapeutic decisions and potentially help to reduce radiation exposure in lower limb CO2 angiography. Although the dose management capabilities of DVA have to be validated in further clinical studies, our preliminary results clearly demonstrate that the platform-independent KMIT software might be a useful new tool in the operating room, and contributes to the safety and efficacy of CO2-enhanced endovascular interventions.
References
Seeger JM, Self S, Harward TR, Flynn TC, Hawkins IF Jr. Carbon dioxide gas as an arterial contrast agent. Ann Surg. 1993;217:688–97.
Sharafuddin MJ, Marjan AE. Current status of carbon dioxide angiography. J Vasc Surg. 2017;66:618–37.
Nijssen EC, Nelemans PJ, Rennenberg RJ, van Ommen V, Wildberger JE. Evaluation of safety guidelines on the use of iodinated contrast material: conundrum continued. Invest Radiol. 2018;53:616–22.
Ghumman SS, Weinerman J, Khan A, Cheema MS, Garcia M, Levin D, Suri R, Prasad A. Contrast induced-acute kidney injury following peripheral angiography with carbon dioxide versus iodinated contrast media: a meta-analysis and systematic review of current literature. Catheter Cardiovasc Interv. 2017;90:437–48.
Chacko S, Joseph G, Thomson V, George P, George O, Danda D. Carbon dioxide angiography-guided renal-related interventions in patients with Takayasu arteritis and renal insufficiency. Cardiovasc Intervent Radiol. 2018;41:998–1007.
Kishino M, Nakaminato S, Kitazume Y, Miyasaka N, Kudo T, Saida Y, Tateishi U. Balloon-occluded carbon dioxide gas angiography for internal iliac arteriography and intervention. Cardiovasc Intervent Radiol. 2018;41:1106–11.
Szigeti K, Máthé D, Osváth S. Motion based X-ray imaging modality. IEEE Trans Med Imaging. 2014;33:2031–8.
Gyano M, Gog I, Orias VI, Ruzsa Z, Nemes B, Csobay-Novak C, Oláh Z, Nagy Z, Merkely B, Szigeti K, Osváth S, Sótonyi P. kinetic imaging in lower extremity arteriography: comparison to digital subtraction angiography. Radiology. 2019;290:246–53.
Orias VI, Gyano M, Gog I, Szollosi D, Veres DS, Nagy Z, Csobay-Novák C, Olah Z, Kiss JP, Osváth S, Szigeti K, Ruzsa Z, Sótonyi P. Digital variance angiography as a paradigm shift in carbon dioxide angiography. Invest Radiol. 2019;54:428–36.
Jeans WD, Stout P. The development and use of digital subtraction angiography. Brit J Radiol. 1990;63:161–8.
Acknowledgements
Open access funding provided by Semmelweis University (SE). MB received a scholarship (“Kiegészítő Kutatási Kiválósági Ösztöndíj”) from Semmelweis University (EFOP-3.6.3-VEKOP-16-2017-00009).
Funding
This study was supported by the National Research, Development and Innovation Office of Hungary (NKFIA; NVKP-16-1-2016-0017 National Heart Program).
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
MG, MB, IG are part time, JPK is full time employee of Kinepict Health Ltd. KS and SO are the owners and founders of Kinepict Health Ltd.
Consent for Publication
Consent for publication was obtained for every individual person’s data included in the study.
Ethical Approval
All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.
Informed Consent
Informed consent was obtained from all individual participants included in the study.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Electronic supplementary material
Below is the link to the electronic supplementary material.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Gyánó, M., Csobay-Novák, C., Berczeli, M. et al. Initial Operating Room Experience with Digital Variance Angiography in Carbon Dioxide-Assisted Lower Limb Interventions: A Pilot Study. Cardiovasc Intervent Radiol 43, 1226–1231 (2020). https://doi.org/10.1007/s00270-020-02530-5
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00270-020-02530-5