Abstract
Purpose
Total knee arthroplasty (TKA) designs evolve continuously to improve patient outcomes. However, incidences of radiolucent lines (RLL) in the latest TKA system have recently been reported, raising concerns. The purpose of the current study was to compare radiographic outcomes of this new TKA implant to its predecessor design.
Methods
A group of 100 patients undergoing TKA using the newer design (Attune) was matched by age and gender to 191 patients with the classic design (LCS). All patients underwent computer-navigated primary TKA by the same surgeon using the same technique. Radiographs were taken before discharge, and 2 and 12 months postoperatively. Radiographic analysis was performed independently by three assessors, using the Modern Knee Society Radiographic Evaluation System and Methodology (MKSRES).
Results
At 12 months postoperatively, the incidence of RLL did not statistically differ between the two implants (14%, n = 14 Attune vs. 8% n = 17 LCS, n.s.). The posterior femoral flange was most commonly affected (12%; n = 12 Attune vs. 7.9%; n = 15 LCS, n.s.) followed by the anterior flange (1%; n = 1 Attune vs. 3.1%; n = 6 LCS, n.s.). The tibial baseplate was only affected in 1% (n = 1) of the Attune and 2.6% (n = 5) of the LCS (n.s.).
Conclusion
At 12 months follow-up we found no significant difference in RLL between the two implants. Both Attune and LCS TKA systems showed RLL predominantly at the posterior femoral flange. The reasons for the RLL remain a matter of speculation; however, shortcomings in surgical and cementing techniques seem to be more important than implant-related factors.
Level of evidence
III
Similar content being viewed by others
Abbreviations
- AP:
-
Anterior–posterior
- CR:
-
Cruciate retaining
- Fz:
-
Femoral zone
- ICC:
-
Interclass-correlation coefficient
- MKSRESM:
-
Modern Knee Society Radiographic Evaluation System and Methodology
- RLL:
-
Radiolucent lines
- RP:
-
Rotating platform
- SD:
-
Standard deviation
- TKA:
-
Total knee arthroplasty
- Tz:
-
Tibial zone
References
No authors listed (2017) Hip, knee & shoulder arthroplasty. In: National Joint Replacement Registry. Australian Orthopaedic Association. Annual Report 2017. https://aoanjrr.sahmri.com/annual-reports-2017. Accessed 1 Nov 2019
No authors listed (2017) National Joint Registry for England, Wales, Northern Ireland and the Isle of Man. In: 14th annual report, 2017. https://reports.njrcentre.org.uk/downloads. Accessed 1 Nov 2019
Bonutti PM, Khlopas A, Chughtai M, Cole C, Gwam CU et al (2017) Unusually high rate of early failure of tibial component in ATTUNE total knee arthroplasty system at implant-cement interface. J Knee Surg 30(5):435–439
Staats K, Wannmacher T, Weihs V, Koller U, Kubista B et al (2019) Modern cemented total knee arthroplasty design shows a higher incidence of radiolucent lines compared to its predecessor. Knee Surg Sports Traumatol Arthrosc 27(4):1148–1155
Sharkey PF, Lichstein PM, Shen C, Tokarski AT, Parvizi J (2013) Why are total knee arthroplasties failing today-has anything changed after 10 years? J Arthroplast 29(9):1774–1778
Ranawat CS, White PB, West S, Ranawat AS (2017) Clinical and radiographic results of attune and PFC sigma knee designs at 2-year follow-up: a prospective matched-pair analysis. J Arthroplast 32(2):431–436
Behrend H, Zdravkovic V, Bösch M, Hochreiter B (2019) No difference in joint awareness after TKA: a matched-pair analysis of a classic implant and its evolutional design. Knee Surg Sports Traumatol Arthrosc 27(7):2124–2129
Meneghini RM, Mont MA, Backstein DB, Bourne RB, Dennis DA et al (2015) Development of a modern knee society radiographic evaluation system and methodology for total knee arthroplasty. J Arthroplast 30(12):2311–2314
Holinka M, Gallo J, Pavlicny R (2017) Occurrence and progression of early radiolucent lines around total knee arthroplasty. Acta Chir Orthop Traumatol Cech 84(5):347–354
Markolf KL, Amstutz HC (1976) Penetration and flow of acrylic bone cement. Clin Orthop Relat Res 121:99–102
Bitsch RG, Obermeyer BJ, Rieger JS, Jager S, Schmalzried TP et al (2013) What is the upper limit of cement penetration for different femoral hip resurfacing components? J Arthroplast 28(4):654–662
Labutti RS, Bayers-Thering M, Krackow KA (2003) Enhancing femoral cement fixation in total knee arthroplasty. J Arthroplast 18(8):979–983
Han HS, Lee MC (2017) Cementing technique affects the rate of femoral component loosening after high flexion total knee arthroplasty. Knee 24(6):1435–1441
Lutz MJ, Pincus PF, Whitehouse SL, Halliday BR (2009) The effect of cement gun and cement syringe use on the tibial cement mantle in total knee arthroplasty. J Arthroplast 24(3):461–467
Schwarze M, Schonhoff M, Beckmann NA, Eckert JA, Bitsch RG et al (2019) Femoral cementation in knee arthroplasty-a comparison of three cementing techniques in a sawbone model using the ATTUNE knee. J Knee Surg. https://doi.org/10.1055/s-0039-1694795
Mann KA, Mocarski R, Damron LA, Allen MJ, Ayers DC (2001) Mixed-mode failure response of the cement-bone interface. J Orthop Res 19(6):1153–1161
Cerquiglini A, Henckel J, Hothi H, Allen P, Lewis J et al (2019) Analysis of the Attune tibial tray backside: a comparative retrieval study. Bone Joint Res 8(3):136–145
Mutsuzaki H, Watanabe A, Kinugasa T, Ikeda K (2018) Radiolucent lines are decreased at 3 years following total knee arthroplasty using trabecular metal tibial components. J Int Med Res 46(5):1919–1927
Gulati A, Chau R, Pandit HG, Gray H, Price AJ et al (2009) The incidence of physiological radiolucency following Oxford unicompartmental knee replacement and its relationship to outcome. J Bone Joint Surg Br 91(7):896–902
Berry DJ (2004) Recognizing and identifying osteolysis around total knee arthroplasty. Instr Course Lect 53:261–264
Engh GA, Ammeen DJ (2001) Periprosthetic osteolysis with total knee arthroplasty. Instr Course Lect 50:391–398
Funding
No external funding was used.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Ethical approval
Approval was obtained from the local ethics committee (Ethikkommission Ostschweiz, Project ID 2019-00927).
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Behrend, H., Hochreiter, B., Potocnik, P. et al. No difference in radiolucent lines after TKA: a matched-pair analysis of the classic implant and its evolutional design. Knee Surg Sports Traumatol Arthrosc 28, 3962–3968 (2020). https://doi.org/10.1007/s00167-020-05894-w
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00167-020-05894-w