Abstract
Objective
In vitro testing of archwires in a multibracket model may provide estimates of force–moment (F/M) systems applied to individual teeth in a realistic geometry. Such investigations have mostly been performed by continuous wire deflection, leading to frictional forces biasing the pure deflection forces. Aim of this study was to quantify this bias and the pure deflection forces for leveling archwires.
Materials and methods
Three nickel–titanium (NiTi) and two multistranded wires were tested in a three-bracket model simulating vertical movement of an upper incisor with a typical interbracket distance of 8 mm (intercenter). To determine pure deflection forces, the middle bracket was first leveled incrementally from its vertical malposition to neutral position with repeated wire insertion at each step (so-called “static leveling mode”). For comparison, forces at the middle bracket were also determined during dynamic leveling with or without ligation of the wire at the lateral brackets by either elastic, tight or loose steel ligatures.
Results
The dynamic mode resulted in significantly lower mean leveling forces for all the tested wires (ANOVA [analysis of variance], p < 0.01) compared to the static mode. Expressed in numbers, dynamic wire unloading resulted in mean force underestimation of 53 ± 9% (loose steel ligatures), 56 ± 11% (elastic ligatures) or 91 ± 29% (tight steel ligatures).
Conclusions
Orthodontic tooth movement is quasi-static. This concerns the initial hyalinization phase in particular. Thus, especially static testing of archwires provides valid reference data for the peak forces exerted directly after clinical insertion of a leveling wire. In dynamic wire testing, significant underestimation of actual forces exerted on individual teeth may occur due to experimental friction, which might considerably differ from that occurring during clinical therapy. This aspect has to be taken into account in the interpretation of published stiffness values for orthodontic wires, and in the selection of the appropriate archwire for leveling of the present tooth malposition, respectively.
Zusammenfassung
Ziel
In-vitro-Untersuchungen von kieferorthopädischen Drähten in einem Multibracket-Modell dienen der Abschätzung der auf die einzelnen Zähne applizierten Kraft-Drehmoment(F/M-)Systeme in einer realitätsnahen Geometrie. Bisher wurden solche Untersuchungen zumeist als kontinuierlich zunehmende Drahtdeflektionen durchgeführt. Auftretende Reibungskräfte führen dazu, dass die dabei gemessenen Kraftwerte nicht die reinen Deflektionskräfte der Drähte darstellen. Ziel dieser Studie war es, diesen Bias zu bestimmen und somit die reinen Deflektionskräfte von Nivellierungsdrähten zu quantifizieren.
Materialien und Methoden
Drei Nickel-Titan(NiTi)-Drähte und zwei mehrsträngige Drähte wurden in einem Drei-Bracket-Modell untersucht. Simuliert wurde dabei eine vertikale Bewegung eines oberen Schneidezahnes mit einer typischen Interbracketdistanz von 8 mm (zwischen den Bracketmitten). Zur Bestimmung der reinen Deflektionskräfte ohne Reibung wurde das mittlere Bracket zunächst schrittweise von seiner vertikalen Fehlstellung in die neutrale Position nivelliert und bei jedem Schritt der Draht neu inseriert (sog. statischer Nivellierungsmodus). Zum Vergleich wurden die Kräfte am mittleren Bracket auch während einer kontinuierlichen, d. h. dynamischen Nivellierung ohne bzw. mit Befestigen des Drahtes an den seitlichen Brackets (entweder bei Verwendung elastischer, fest angezogener oder loser Stahlligaturen) bestimmt.
Ergebnisse
Der dynamische Modus ergab im Vergleich zum statischen Modus für alle getesteten Drähte im Durchschnitt signifikant geringere Kräfte (ANOVA [„analysis of variance“], p < 0,01). Quantitativ ausgedrückt lag die durchschnittliche Kraftunterschätzung bei dynamischer Drahtentlastung bei 53 ± 9 % (lose Stahlligaturen), 56 ± 11 % (elastische Ligaturen) bzw. 91 ± 29 % (fest angezogene Stahlligaturen).
Schlussfolgerungen
Die orthodontische Zahnbewegung ist quasi statisch. Dies betrifft insbesondere die initiale Hyalinisationsphase. Die statische Untersuchung kieferorthopädischer Drähte liefert zumindest für diese Phase valide Referenzdaten bezüglich der direkt nach Insertion eines Nivellierungsdrahtes ausgeübten Kräfte. Aufgrund der experimentellen Reibung, die sich erheblich von der klinisch auftretenden Reibung unterscheiden kann, führen dynamische Drahtuntersuchungen dazu, dass die tatsächlich direkt nach Drahtinsertion auf die einzelnen Zähne ausgeübten Kräfte signifikant unterschätzt werden. Dieser Aspekt muss bei der Interpretation der veröffentlichten Kraft- und Steifigkeitswerte für kieferorthopädische Drähte bzw. bei der darauf basierenden Auswahl eines geeigneten Nivellierungsdrahtes für die vorhandene Zahnfehlstellung berücksichtigt werden.
Similar content being viewed by others
References
Alobeid A, Dirk C, Reimann S et al (2017) Mechanical properties of different esthetic and conventional orthodontic wires in bending tests: an in vitro study. J Orofac Orthop 78(3):241–252
Armstrong D, Kharbanda OP, Petocz P et al (2006) Root resorption after orthodontic treatment. Aust Orthod J 22(2):153–160
Badawi HM, Toogood RW, Carey JPR et al (2009) Three-dimensional orthodontic force measurements. Am J Orthod Dentofacial Orthop 136(4):518–528
Bantleon HP, Droschl H, Stern G (1989) Different applications of various wire alloys in fixed appliance technic. Inf Orthod Kieferorthop 21(2):173–183
Bartzela TN, Senn C, Wichelhaus A (2007) Load-deflection characteristics of superelastic nickel-titanium wires. Angle Orthod 77(6):991–998
Braun S, Bluestein M, Moore BK et al (1999) Friction in perspective. Am J Orthod Dentofacial Orthop 115(6):619–627
Drescher D, Bourauel C, Thier M (1991) Application of the orthodontic measurement and simulation system (OMSS) in orthodontics. Eur J Orthod 13(3):169–178
Franchi L, Baccetti T, Camporesi M et al (2009) Forces released by nonconventional bracket or ligature systems during alignment of buccally displaced teeth. Am J Orthod Dentofacial Orthop 136(3):316
Fuck L‑M, Drescher D (2006) Force systems in the initial phase of orthodontic treatment—a comparison of different leveling arch wires. J Orofac Orthop 67(1):6–18
Fuck L‑M, Wilmes B, Gürler G et al (2007) Frictional behavior of self ligating brackets in comparison to conventional brackets. Inf Orthod Kieferorthop 39(1):6–17
Kasuya S, Nagasaka S, Hanyuda A et al (2007) The effect of ligation on the load deflection characteristics of nickel titanium orthodontic wire. Eur J Orthod 29(6):578–582
Krishnan V (2007) Orthodontic pain: from causes to management—a review. Eur J Orthod 29(2):170–179
Kurol J, Owman-Moll P (1998) Hyalinization and root resorption during early orthodontic tooth movement in adolescents. Angle Orthod 68(2):161–165
Lapatki BG, Paul O (2007) Smart brackets for 3D-force-moment measurements in orthodontic research and therapy—developmental status and prospects. J Orofac Orthop 68(5):377–396
Lindhe J, Karring T, Araujo M (2003) Anatomy of the periodontium. In: Lindhe J, Karring T, Lang NP (eds) Clinical periodontology and implant dentistry, 4th edn. Blackwell Munksgaard, Oxford, pp 3–49
Moyers RE, van der Linden FPGM, Riolo ML, Mc Namara JA (1976) Tooth Size. In: Moyers RE (ed) Standards of human occlusal development. Center for Human Growth and Development, Univ. of Michigan, Ann Arbor
Nakano H, Satoh K, Norris R et al (1999) Mechanical properties of several nickel-titanium alloy wires in three-point bending tests. Am J Orthod Dentofacial Orthop 115(4):390–395
Nucera R, Gatto E, Borsellino C et al (2014) Influence of bracket-slot design on the forces released by superelastic nickel-titanium alignment wires in different deflection configurations. Angle Orthod 84(3):541–547
Oltjen JM, Duncanson MG, Ghosh J et al (1997) Stiffness-deflection behavior of selected orthodontic wires. Angle Orthod 67(3):209–218
Pesce RE, Uribe F, Janakiraman N et al (2014) Evaluation of rotational control and forces generated during first-order archwire deflections: a comparison of self-ligating and conventional brackets. Eur J Orthod 36(3):245–254
Proffit WR (2007) Biological Basis of Orthodontic Therapy. In: Proffit WR, Fields HW, Sarver DM (eds) Contemporary orthodontics, 4th edn. Elsevier Mosby, St. Louis, pp 331–358
Risinger RK, Proffit WR (1996) Continuous overnight observation of human premolar eruption. Arch Oral Biol 41(8–9):779–789
Schumacher HA, Bourauel C, Drescher D (1992) Deactivation and effectiveness of orthodontic levelling arches—a dynamic analysis of the force systems. J Orofac Orthop 53(5):273–285
Segner D, Ibe D (1995) Properties of superelastic wires and their relevance to orthodontic treatment. Eur J Orthod 17(5):395–402
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
K. Naziris, N.E. Piro, R. Jäger, F. Schmidt, F. Elkholy and B.G. Lapatki declare that they have no competing interests.
Rights and permissions
About this article
Cite this article
Naziris, K., Piro, N.E., Jäger, R. et al. Experimental friction and deflection forces of orthodontic leveling archwires in three-bracket model experiments. J Orofac Orthop 80, 223–235 (2019). https://doi.org/10.1007/s00056-019-00187-5
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00056-019-00187-5