Abstract
Wear and osteolysis continue to be major reasons for revision surgery in THA. Ceramic-on-ceramic bearings eliminate polyethylene wear debris. The newest generation of these bearings incorporate nanosized, yttria-stabilized tetragonal zirconia particles producing an alumina matrix composite. We asked whether this new material would perform as well as a conventional bearing in terms of functional hip scores, radiographic migration and osteolysis, complications and survival. As part of a US FDA investigational device exemption study (G000075), we conducted an initial prospective safety study of 21 alumina matrix composite femoral heads articulating on alumina liners followed by a prospective, randomized study with 44 more of these articulations and 45 zirconia femoral heads on polyethylene liners. The minimum followup for all patients was 26 months (mean, 73 months; range, 26–108 months). Harris hip scores and radiographic findings were similar in the two groups as was survivorship (trial 95% versus control 93%). There were three reoperations in the trial group and three in the control group. A fractured head retrieval showed a 33% monoclinic transformation with an increase in surface roughness from 3 to 5 nm at the main wear zone. While our numbers were insufficient to compare device-related complications, the trial device performed as well as the control device in terms of reoperation, and clinical and radiographic outcome. The alumina matrix composite femoral head on an alumina liner provided high survivorship.
Level of Evidence: Level II, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.
Similar content being viewed by others
References
Akagi M, Nonaka T, Nishisaka F, Mori S, Fukuda K, Hamanishi C. Late dissociation of an alumina-on-alumina bearing modular acetabular component. J Arthroplasty. 2004;19:647–651.
Allain J, Roudot-Thoraval F, Delecrin J, Anract P, Migaud H, Goutallier D. Revision total hip arthroplasty performed after fracture of a ceramic femoral head. J Bone Joint Surg Am. 2003;85:825–830.
Bierbaum BE, Mairus J, Kuesis D, Morrison JC, Ward D. Ceramic-on-ceramic bearings in total hip arthroplasty. Clin Orthop Relat Res. 2002;405:158–163.
Blackwelder WC. “Proving the null hypothesis” in clinical trials. Control Clin Trials. 1982;3:345–353.
Clarke IC, Manaka M, Green DD, Williams P, Pezzotti G, Kim YH, Ries M, Sugano N, Sedel L, Delauney C, Nissan BB, Donaldson T, Gustafson GA. Current status of zirconia used in total hip implants. J Bone Joint Surg Am. 2003;85:73–84.
D’Antonio J, Capello W, Manley M, Naughton M, Sutton K. Alumina ceramic bearings for total hip arthroplasty: five-year results of a prospective randomized study. Clin Orthop Relat Res. 2005;436:164–171.
Dearnley PA. A review of metallic, ceramic and surface-treated metals used for bearing surfaces in human joint replacement. Proc Inst Mech Eng H. 1999;213:107–135.
De Aza AH, Chevalier J, Fantozzi G, Schehl M, Torrecillas R. Crack growth resistance of alumina, zirconia and zirconia toughened alumina ceramics for joint prostheses. Biomaterials. 2002;23:937–945.
DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976;121:20–32.
Frndak PA, Mallory TH, Lombardi AV Jr. Translateral surgical approach to the hip: the abductor muscle “split.” Clin Orthop Relat Res. 1993;295:135–141.
Gruen TA, McNeice GM, Amstutz HC. “Modes of failure” of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res. 1979;141:17–27.
Hamadouche M, Boutin P, Daussange J, Bolander ME, Sedel L. Alumina-on-alumina total hip arthroplasty: a minimum 18.5 year follow-up study. J Bone Joint Surg Am. 2002;84:69–77.
Haraguchi K, Sugano N, Nishii T, Miki H, Oka K, Yoshikawa H. Phase transformation of a zirconia ceramic head after total hip arthroplasty. J Bone Joint Surg Br. 2001;83:996–1000.
Harris WH. Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg Am. 1969;51:737–755.
Hasegawa M, Sudo A, Hirata H, Uchida A. Ceramic acetabular liner fracture in total hip arthroplasty with a ceramic sandwich cup. J Arthroplasty. 2003;18:658–661.
Hasegawa M, Sudo A, Uchida A. Cobalt-chromium head wear following revision hip arthroplasty performed after ceramic fracture—a case report. Acta Orthop. 2006;77:833–835.
Hasegawa M, Sudo A, Uchida A. Alumina ceramic-on-ceramic total hip replacement with a layered acetabular component. J Bone Joint Surg Br. 2006;88:877–882.
Huo MH, Martin RP, Zatorski LE, Keggi KJ. Ceramic total hip replacements done without cement: an average 9-year follow-up study. Clin Orthop Relat Res. 1996;332:143–150.
Hwang SK, Oh JR, Her MS, Shim YJ, Cho TY, Kwon SM. Fracture-dissociation of ceramic liner. Orthopedics. 2008;31:804.
Inzerillo CV, Garino JP. Alternative bearing surfaces in total hip arthroplasty. J Southern Orthop Assoc. 2003;12:106–111.
Jazrawi LM, Kummer FJ, DiCesare PE. Alternative bearing surfaces for total joint arthroplasty. J Am Acad Orthop Surg. 1998;6:198–203.
Kaplan EL, Meier P. Nonparametric estimation from incomplete observations. J Am Statistical Assoc. 1958;53:457–481.
Keurentjes JC, Kuipers RM, Wever DJ, Schreurs BW. High incidence of squeaking in THAs with alumina ceramic-on-ceramic bearings. Clin Orthop Relat Res. 2008;466:1438–1443.
Kraay MJ, Thomas RD, Rimnac CM, Fitzgerald SJ, Goldberg VM. Zirconia versus Co-Cr femoral heads in total hip arthroplasty. Clin Orthop Relat Res. 2006;453:86–90.
Lusty PJ, Tai CC, Sew-Hoy RP, Walter WL, Walter WK, Zicat BA. Third-generation alumina-on-alumina ceramic bearings in cementless total hip arthroplasty. J Bone Joint Surg Am. 2007;89:2676–2683.
Masonis JL, Bourne RB, Ries MD, McCalden RW, Salehi A, Kelman DC. Zirconia femoral head fracture: a clinical and retrieval analysis. J Arthroplasty. 2004;19:898–905.
Masson B. Emergence of the alumina matrix composite in total hip arthroplasty. Int Orthop. 2009;33:359–363.
Mehmood S, Riyaz JH, Pandit H. Review of ceramic-on-ceramic total hip arthroplasty. J Surg Ortho Adv. 2008;17:45–50.
Morita Y, Nakata K, Kim YH, Sekino T, Niihara K, Ikeuchi K. Wear properties of alumina/zirconia composite ceramics for joint prostheses measured with an end-face apparatus. Biomed Mater Eng. 2004;14:263–270.
Morlock M, Nassutt R, Janssen R, Willmann G, Honl M. Mismatched wear couple zirconium oxide and aluminum oxide in total hip arthroplasty. J Arthroplasty. 2001;16:1071–1074.
Nevelos J, Ingham E, Doyle C, Streicher R, Nevelos A, Walter W, Fisher J, Eng D. Microseparation of the centers of alumina-alumina artificial hip joints during simulator testing produces clinically relevant wear rates and patterns. J Arthroplasty. 2000;15:793–795.
Nizard R, Pourreyron D, Raould A, Hannouche D, Sedel L. Alumina-on-alumina hip arthroplasty in patients younger than 30 years old. Clin Orthop Relat Res. 2008;466:317–323.
Pulliam IT, Trousdale RT. Fracture of a ceramic femoral head after a revision operation. J Bone Joint Surg Am. 1997;79:118–121.
Ravasi F, Sansone V. Five-year follow-up with a ceramic sandwich cup in total hip replacement. Arch Orthop Trauma Surg. 2002;122:350–353.
Restrepo C, Parvizi J, Kurtz SM, Sharkey PF, Hozack WJ, Rothman RH. The noisy ceramic hip: is component malposition the cause? J Arthroplasty. 2008;23:643–649.
Rhoads DP, Baker KC, Israel R, Greene PW. Fracture of an alumina femoral head used in ceramic-on-ceramic total hip arthroplasty. J Arthroplasty. 2008;23:1239.e25–e30.
Rodríguez JA, Gonzalez DelaValle A, McCook N. Squeaking in total hip replacement: a cause for concern. Orthopedics. 2008;31:874, 877–878.
Simon JA, Dayan AJ, Enrique E, Stuchin SA, Di Cesare PE. Catastrophic failure of the acetabular component in a ceramic-polyethylene bearing total hip arthroplasty. J Arthroplasty. 1998;13:108–113.
Stewart TD, Tipper JL, Streicher RM, Ingham E, Fisher J. Long-term wear of HIPed alumina on alumina bearings for THR under microseparation conditions. J Mater Sci Mater Med. 2001;12:1053–1056.
Streicher RM, Semlitsch M, Schoen R, Weber H, Rieker C. Metal-on-metal articulation for artificial hip joints: laboratory study and clinical results. Proc Inst Mech Eng. 1996;210:223–232.
Sychterz CJ, Engh CA, Swope SW, McNulty DE, Engh CA. Analysis of prosthetic femoral heads retrieved at autopsy. Clin Orthop Relat Res. 1999;358:223–234.
Tateiwa T, Clarke IC, Williams PA, Garino J, Manaka M, Shishido T, Yamamoto K, Imakiire A. Ceramic total hip arthroplasty in the United States: safety and risk issues revisited. Am J Orthop. 2008;37:E26–E31.
Taylor S, Manley MT, Sutton K. The role of stripe wear in causing acoustic emissions from alumina ceramic-on-ceramic bearings. J Arthroplasty. 2007;22(suppl 3):47–51.
Tradonsky S, Postak PD, Froimson AI, Greenwald AS. A comparison of the disassociation strength of modular acetabular components. Clin Orthop Relat Res. 1993;296:154–160.
Urban JA, Garvin KL, Boese CK, Bryson L, Pedersen DR, Callaghan JJ, Miller RK. Ceramic-on-polyethylene bearing surfaces in total hip arthroplasty: seventeen to twenty-one year results. J Bone Joint Surg Am. 2001;83:1688–1694.
Walter WL, Insley GM, Walter WK, Tuke MA. Edge loading in third generation alumina ceramic-on-ceramic bearings: stripe wear. J Arthroplasty. 2004;19:402–413.
Walter WL, O’Toole GC, Walter WK, Ellis A, Zicat BA. Squeaking in ceramic-on-ceramic hips: the importance of acetabular orientation. J Arthroplasty. 2007;22:496–503.
Walter WL, Waters TS, Gillies M, Donohoo S, Kurtz SM, Ranawat AS, Hozack WJ, Tuke MA. Squeaking hips. J Bone Joint Surg Am. 2008;90 (suppl 4):102–111.
Williams VG 2nd, Whiteside LA, White SE, McCarthy DS. Fixation of ultrahigh-molecular-weight polyethylene liners to metal-backed acetabular cups. J Arthroplasty. 1997;12:25–31.
Willmann G (2000) Ceramic femoral head retrieval data. Clin Orthop Relat Res. 379:22–28.
Yamamoto K, Shishido T, Tateiwa T, Katori Y, Masaoka T, Imakiire A, Clarke IC. Failure of ceramic THR with liner dissociation—a case report. Acta Orthop Scand. 2004;75:500–502.
Yang CC, Kim RH, Dennis DA. The squeaking hip: a cause for concern—Disagrees. Orthopedics. 2007;30:739–742.
Yoo JJ, Kim YM, Yoon KS, Koo KH, Song WS, Kim HJ. Alumina-on-alumina total hip arthroplasty: a five-year minimum follow-up study. J Bone Joint Surg Am. 2005;87:530–535.
Acknowledgments
We thank Tom Donaldson, MD, and Koseki Kubo, MD, PhD, for their assistance in analysis of the delta head fracture. We also thank Tawnya L. Tucker, MT, for her assistance with the collection of data in this study.
Author information
Authors and Affiliations
Corresponding author
Additional information
One or more of the authors (AVL, KRB) receive royalties and institutional research support from and have consulting agreements with Biomet, Inc (Warsaw, IN). One author (KRB) has consulting agreements with Synvasive Technology, Inc (El Dorado Hills, CA) and Salient Surgical Technologies (Portsmouth, NH) and owns stock in Angiotech Pharmaceuticals, Inc (Vancouver, BC). One author (AVL) receives royalties from Innomed, Inc (Savannah, GA).
Each author certifies that his or her institution has approved the human protocol for this investigation, that all investigations were conducted in conformity with ethical principles of research, and that informed consent for participation in the study was obtained.
This work was performed at Joint Implant Surgeons, Inc.
About this article
Cite this article
Lombardi, A.V., Berend, K.R., Seng, B.E. et al. Delta Ceramic-on-Alumina Ceramic Articulation in Primary THA: Prospective, Randomized FDA-IDE Study and Retrieval Analysis. Clin Orthop Relat Res 468, 367–374 (2010). https://doi.org/10.1007/s11999-009-1143-9
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11999-009-1143-9