Survey of blood pump diaphragm damage in the NIPRO-ventricular assist device

  • Original Article
  • Published:
Journal of Artificial Organs Aims and scope Submit manuscript

Abstract

We surveyed the incidence of blood pump diaphragm damage (rupture or crack) in the NIPRO-ventricular assist device (VAD). In the cases in which rupture or suspected blood pump crack was detected, we disassembled the pumps to visually check the condition of the diaphragm after replacement or use. Of 366 blood pumps surveyed, diaphragm damage was observed in 2.7 %. The duration of use of the blood pumps with diaphragm damage was significantly longer than that of pumps without damage. The incidence of diaphragm damage increased with longer duration of use. On the basis of these findings, blood pump diaphragm damage in the NIPRO-VAD may be associated with duration of use. However, some blood pumps were used for prolonged periods without diaphragm damage. All blood pumps with damage had a crack in the diaphragm on the air chamber side near the diaphragm–housing (D–H) junction. Cracks were not found in any specific part of the diaphragm. In blood pumps with diaphragm rupture, the crack had a through-hole reaching the blood-contacting surface. Although we were unable to identify the causes of the cracks, it is suggested that when a crack appears in the diaphragm it will gradually expand and eventually lead to rupture. If a crack is detected in a blood pump, we advocate replacing the pump before it grows. When the NIPRO-VAD is used, it is necessary to keep in mind that blood pump diaphragm damage may occur.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6

Similar content being viewed by others

References

  1. Takano H, Taenaka Y, Noda H, Kinoshita M, Yagura A, Tatsumi E, Sekii H, Sasaki E, Umezu M, Nakatani T. Multi-institutional studies of the national cardiovascular center ventricular assist system: use in 92 patients. ASAIO Trans. 1989;35:541–4.

    Article  PubMed  CAS  Google Scholar 

  2. Takano H, Nakatani T, Taenaka Y. Clinical experience with ventricular assist systems in Japan. Ann Thorac Surg. 1993;55:250–6.

    Article  PubMed  CAS  Google Scholar 

  3. Kurihara C, Nishimura T, Nawata K, Kinoshita O, Hisagi M, Motomura N, Kyo S, Ono M. Successful bridge to recovery with VAD implantation for anthracycline-induced cardiomyopathy. J Artif Organs. 2011;14:249–52.

    Article  PubMed  Google Scholar 

  4. Kurihara C, Ono M, Nishimura T, Nawata K, Kinoshita O, Hisagi M, Motomura N, Kyo S. Prolonged biventricular assist device support as a bridge to heart transplantation. J Artif Organs. 2011;14:367–70.

    Article  PubMed  Google Scholar 

  5. Nishimura T, Kyo S. High-dose carvedilol therapy for mechanical circulatory assisted patients. J Artif Organs. 2010;13:88–91.

    Article  PubMed  CAS  Google Scholar 

  6. Ono M. Cardiomyopathy and myocarditis; surgical treatment including ventricular assist device and heart transplantation. Kyobu Geka. 2011;64:666–72 (in Japanese).

    Google Scholar 

  7. Samuels LE, Holmes EC, Garwood P, Ferdinand F. Initial experience with the Abiomed AB5000 ventricular assist device system. Ann Thorac Surg. 2005;80:309–12.

    Article  PubMed  Google Scholar 

  8. Anderson M, Smedira N, Samuels L, Madani M, Naka Y, Acker M, Hout M, Benali K. Use of the AB5000 ventricular assist device in cardiogenic shock after acute myocardial infarction. Ann Thorac Surg. 2010;90:706–12.

    Article  PubMed  Google Scholar 

  9. Engin C, Ayik F, Oguz E, Eygi B, Yagdi T, Karakula S, Ozbaran M. Ventricular assist device as a bridge to heart transplantation in adults. Transpl Proc. 2011;43:927–30.

    Article  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Koichi Kashiwa.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Kashiwa, K., Nishimura, T., Nakahata, A. et al. Survey of blood pump diaphragm damage in the NIPRO-ventricular assist device. J Artif Organs 15, 341–346 (2012). https://doi.org/10.1007/s10047-012-0657-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10047-012-0657-1

Keywords

Navigation