Abstract

Purpose: Luminal diameters measured in vivo by calibrated-catheter angiography and by intravascular ultrasound were correlated with those obtained from pressure-fixed histologic cross-sections to determine the accuracy of both methods.

Methods: Angiographic and endosonographic diameter measurements were performed in the center of stents placed in the iliac arteries of 10 miniature pigs and were compared with luminal and stent diameters in postmortem, pressure-fixed, histologic cross-sections from identical locations.

Results: Compared with histologic diameters, magnification-corrected angiographic measurements still magnified vascular luminal diameters by 0.7 ± 0.71 mm (r= 0.41, Pearson; p < 0.003, Wilcoxon, matched pairs), whereas intravascular ultrasound measurements proved to be almost identical to the histologic lumina (r= 0.95, Pearson; p> 0.5, Wilcoxon, matched pairs). Similarly, stent diameters correlated well between endosonographic and histologic measurements (r= 0.91; p= 0.002), and less well between angiographic and histologic diameters (r= 0.62; p= 0.002).

Conclusion: Since calibrated angiography still overestimates vascular lumina, endosonography is the preferred technique for accurate in vivo measurements.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Additional information

Received: 0/00/00/Accepted: 0/00/00

Rights and permissions

Reprints and permissions

About this article

Cite this article

Froelich, J., Hoppe, M., Nahrstedt, C. et al. The Precise Determination of Vascular Lumen and Stent Diameters: Correlation Among Calibrated Angiography, Intravascular Ultrasound, and Pressure-Fixed Specimens. Cardiovasc Intervent Radiol 20, 452–456 (1997). https://doi.org/10.1007/s002709900192

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s002709900192

Navigation